Use of the bougie in simulated difficult intubation. 2. Comparison of single-use bougie with multiple-use bougie

被引:40
|
作者
Marfin, AG
Pandit, JJ [1 ]
Hames, KC
Popat, MT
Yentis, SM
机构
[1] John Radcliffe Hosp, Nuffield Dept Anaesthesia, Oxford OX3 9DU, England
[2] Chelsea & Westminster Hosp, Magill Dept Anaesthet, London, England
关键词
Intubation; tracheal; difficult; Equipment; multiple-use and single-use tracheal tube introducers;
D O I
10.1046/j.1365-2044.2003.03289.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
We studied the success rates for tracheal intubation in 32 healthy, anaesthetised patients during simulated grade IIIa laryngoscopy, randomised to either the multiple-use or the single-use bougie. Success rates (primary end-point) and times taken (secondary end-point) to achieve tracheal intubation were recorded. The multiple-use bougie was more successful than the single-use one (15/16 successful intubations vs. 9/16; p=0.03). With either device, median [range] total tracheal intubation times for successful attempts were < 54 [24-84] s and there were no clinically important differences between these times. We conclude that the multiple-use bougie is a more reliable aid to tracheal intubation than the single-use introducer in grade IIIa laryngoscopy.
引用
收藏
页码:852 / 855
页数:4