Use of the bougie in simulated difficult intubation. 2. Comparison of single-use bougie with multiple-use bougie

被引:40
|
作者
Marfin, AG
Pandit, JJ [1 ]
Hames, KC
Popat, MT
Yentis, SM
机构
[1] John Radcliffe Hosp, Nuffield Dept Anaesthesia, Oxford OX3 9DU, England
[2] Chelsea & Westminster Hosp, Magill Dept Anaesthet, London, England
关键词
Intubation; tracheal; difficult; Equipment; multiple-use and single-use tracheal tube introducers;
D O I
10.1046/j.1365-2044.2003.03289.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
We studied the success rates for tracheal intubation in 32 healthy, anaesthetised patients during simulated grade IIIa laryngoscopy, randomised to either the multiple-use or the single-use bougie. Success rates (primary end-point) and times taken (secondary end-point) to achieve tracheal intubation were recorded. The multiple-use bougie was more successful than the single-use one (15/16 successful intubations vs. 9/16; p=0.03). With either device, median [range] total tracheal intubation times for successful attempts were < 54 [24-84] s and there were no clinically important differences between these times. We conclude that the multiple-use bougie is a more reliable aid to tracheal intubation than the single-use introducer in grade IIIa laryngoscopy.
引用
收藏
页码:852 / 855
页数:4
相关论文
共 21 条
  • [1] A comparison of simulated difficult intubation with multiple-use and single-use bougies in a manikin
    Annamaneni, R
    Hodzovic, I
    Wilkes, AR
    Latto, IP
    ANAESTHESIA, 2003, 58 (01) : 45 - 49
  • [2] Simulated difficult intubation - Comparison of the gum elastic bougie and the stylet
    Gataure, PS
    Vaughan, RS
    Latto, IP
    ANAESTHESIA, 1996, 51 (10) : 935 - 938
  • [3] Use of the endotracheal Bougie introducer for difficult intubations
    Phelan, MP
    AMERICAN JOURNAL OF EMERGENCY MEDICINE, 2004, 22 (06) : 479 - 482
  • [4] To shape or not to shape ... Simulated bougie-assisted difficult intubation in a manikin
    Hodzovic, I
    Wilkes, AR
    Latto, IP
    ANAESTHESIA, 2003, 58 (08) : 791 - 797
  • [5] Comparison of a Disposable Bougie versus a Newly Designed Malleable Bougie in the Intubation of a Difficult Manikin Airway
    Boedeker, Ben H.
    Bernhagen, Mary
    Miller, David J.
    Murray, W. Bosseau
    MEDICINE MEETS VIRTUAL REALITY 18, 2011, 163 : 65 - 67
  • [6] Survey of the use of the gum elastic bougie in clinical practice
    Latto, IP
    Stacey, M
    Mecklenburgh, J
    Vaughan, RS
    ANAESTHESIA, 2002, 57 (04) : 379 - 384
  • [7] Evaluation of Frova, single-use intubation introducer, in a manikin. Comparison with Eschmann multiple-use introducer and Portex single-use introducer
    Hodzovic, I
    Latto, IP
    Wilkes, AR
    Hall, JE
    Mapleson, WW
    ANAESTHESIA, 2004, 59 (08) : 811 - 816
  • [8] Iatrogenic oesophageal perforation following the use of a bougie
    Sievert, K.
    Fischler, M.
    ANNALES FRANCAISES D ANESTHESIE ET DE REANIMATION, 2010, 29 (10): : 737 - 738
  • [9] Managing a critical airway with an unloaded single-use flexible bronchoscope as bougie: The SUBB-approach
    Bergqvist, Joel
    Aghajanzadeh, Susan
    Tulinius, Arnar
    Hammarfjord, Oscar
    OTOLARYNGOLOGY CASE REPORTS, 2023, 27
  • [10] Randomized evaluation of the performance of single-use laryngoscopes in simulated easy and difficult intubation
    Twigg, SJ
    McCormick, B
    Cook, TM
    BRITISH JOURNAL OF ANAESTHESIA, 2003, 90 (01) : 8 - 13