Deconvolution analysis improves real-time OSL of BeO ceramic

被引:5
作者
Madden, Levi [1 ]
Lukas, Erin [2 ,3 ]
Santos, Alexandre [2 ,3 ,4 ]
Ganija, Miftar [2 ,3 ,5 ]
Veitch, Peter [2 ,3 ]
Rosenfeld, Anatoly [1 ,6 ]
Li, Enbang [1 ]
机构
[1] Univ Wollongong, Ctr Med Radiat Phys, Wollongong, NSW 2522, Australia
[2] Univ Adelaide, Sch Phys Sci, Adelaide, SA 5005, Australia
[3] Univ Adelaide, Inst Photon & Adv Sensing, Adelaide, SA 5005, Australia
[4] Royal Adelaide Hosp, Dept Radiat Oncol, Adelaide, SA 5000, Australia
[5] Def Sci & Technol, Edinburgh, SA 5111, Australia
[6] Univ Wollongong, Illawarra Med & Hlth Res Inst, Wollongong, NSW 2522, Australia
关键词
Optically stimulated luminescence; in vivo dosimetry; Radiotherapy; BeO ceramic; Fibre-coupled luminescence dosimetry; OPTICALLY STIMULATED LUMINESCENCE; DOSIMETRY;
D O I
10.1016/j.radmeas.2021.106680
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
A novel method is presented to correct the non-linearity of real-time optically stimulated luminescence (rtOSL), investigated for a fibre-coupled BeO dosimetry system. The presented method applies novel kinetic modelling techniques to model the time-resolved optically stimulated luminescence, and least squares deconvolution to transform the measured rtOSL into the absorbed dose in BeO. This correction, termed the deconvolution correction, is compared against the Delta rtOSL and exponential correction methods. These rtOSL correction methods multiply the measured rtOSL by a correction function. In the..rtOSL correction, this correction function is theoretically derived, whereas the exponential correction approximates the correction function as an exponential growth with respect to time. The deconvolution correction method had a mean dose difference of 1.5%, improving upon the 4.4% and 4.1% achieved by the Delta rtOSL and exponential correction methods, respectively. These prior methods correct the measured rtOSL through multiplication with a sensitivity correction function. The deconvolution corrected responses were characterised by a 1.9% standard deviation with the expected doses, and an average root mean squared error (RMSE) of 2.0% between the corrected rtOSL signal and the modelled time-dependent accumulated dose. In comparison, the Delta rtOSL had a standard deviation of 6.4% and an average RMSE of 7.3% between the corrected rtOSL signal and the modelled time-dependent accumulated dose, while the exponential correction had a standard deviation of 5.4% and a 4.6% average RMSE between the corrected rtOSL and the time-dependent accumulated dose. To determine whether the deconvolution was feasible for real-time correction of measured rtOSL, a computational simulation of the real-time measurement and correction was performed. The maximum computation time for the deconvolution correction was 18 ms, demonstrating that the deconvolution correction is feasible for real-time correction of the rtOSL.
引用
收藏
页数:8
相关论文
共 22 条
[1]  
Abramowitz M., 1974, Handbook of Mathematical Functions, with Formulas, Graphs, and Mathematical Tables, DOI DOI 10.5555/1953048.2021068
[2]   Fiber-coupled Luminescence Dosimetry in Therapeutic and Diagnostic Radiology [J].
Andersen, Claus E. .
CONCEPTS AND TRENDS IN MEDICAL RADIATION DOSIMETRY, 2011, 1345 :100-119
[3]   An algorithmic approach to single-probe Cherenkov removal in pulsed x-ray beams [J].
Archer, James ;
Madden, Levi ;
Li, Enbang ;
Wilkinson, Dean ;
Rosenfeld, Anatoly B. .
MEDICAL PHYSICS, 2019, 46 (04) :1833-1839
[4]   Review of plastic and liquid scintillation dosimetry for photon, electron, and proton therapy [J].
Beaulieu, Luc ;
Beddar, Sam .
PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (20) :R305-R343
[5]   Blind deconvolution estimation by multi-exponential models and alternated least squares approximations: Free-form and sparse approach [J].
Campos-Delgado, Daniel U. ;
Gutierrez-Navarro, Omar ;
Salinas-Martinez, Ricardo ;
Duran, Elvis ;
Mejia-Rodriguez, Aldo R. ;
Velazquez-Duran, Miguel J. ;
Jo, Javier A. .
PLOS ONE, 2021, 16 (03)
[6]   The analysis of time-resolved optically stimulated luminescence: II. Computer simulations and experimental results [J].
Chithambo, M. L. .
JOURNAL OF PHYSICS D-APPLIED PHYSICS, 2007, 40 (07) :1880-1889
[7]   The analysis of time-resolved optically stimulated luminescence: I. Theoretical considerations [J].
Chithambo, M. L. .
JOURNAL OF PHYSICS D-APPLIED PHYSICS, 2007, 40 (07) :1874-1879
[8]   Fluorescence lifetime imaging microscopy: fundamentals and advances in instrumentation, analysis, and applications [J].
Datta, Rupsa ;
Heaster, Tiffany M. ;
Sharick, Joe T. ;
Gillette, Amani A. ;
Skala, Melissa C. .
JOURNAL OF BIOMEDICAL OPTICS, 2020, 25 (07)
[9]   A fiber-dosimetry method based on OSL from Al2O3:C for radiotherapy applications [J].
Gaza, R ;
McKeever, SWS ;
Akselrod, MS ;
Akselrod, A ;
Underwood, T ;
Yoder, C ;
Andersen, CE ;
Aznar, MC ;
Marckmann, CJ ;
Botter-Jensen, L .
RADIATION MEASUREMENTS, 2004, 38 (4-6) :809-812
[10]   Performance of a near-real-time KBr:Eu dosimetry system under computed tomography x-rays [J].
Klein, David ;
Peakheart, David W. ;
McKeever, Stephen W. S. .
RADIATION MEASUREMENTS, 2010, 45 (3-6) :663-667