Validation of a computerised logMAR visual acuity measurement system (COMPlog): comparison with ETDRS and the electronic ETDRS testing algorithm in adults and amblyopic children

被引:54
作者
Laidlaw, D. A. H. [1 ,2 ]
Tailor, V. [2 ]
Shah, N. [1 ]
Atamian, S. [1 ]
Harcourt, C. [2 ]
机构
[1] St Thomas Hosp, Vitreo Retinal Unit, London SE1 7EH, England
[2] EEMU Maidstone Gen Hosp, Maidstone, Kent, England
关键词
D O I
10.1136/bjo.2007.121715
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Background/Aim: The COMPlog clinical visual acuity measuring system is being developed for both routine and research use. This study aimed to validate its performance in amblyopic children and both normal and diseased adults against the gold standard ETDRS chart and the E-ETDRS computerised acuity measurement algorithm. Method: Timed test and retest fully interpolated five letters per line logMAR visual acuity measurements were taken for 70 adults and 59 amblyopic children using the ETDRS chart and the COMPlog visual acuity measurement system. 39 of the adults also underwent computerised acuity testing using the E-ETDRS testing algorithm. The tested adults included normals as well as subjects with a range of ocular diseases. The methods of Bland and Altman were employed with test - retest variability (TRV) expressed as 95% confidence limits for agreement. Results: No significant bias was observed between the gold standard ETDRS acuity measurements and those taken with either COMPlog or E-ETDRS. TRVs of +/- 0.12 logMAR and +/- 0.10 logMAR were respectively found for COMPlog measurements in the amblyopic children and adult groups compared with +/- 0.12 logMAR for the ETDRS chart in both groups. The TRV of the E-ETDRS system was slightly greater at +/- 0.16 logMAR. Median testing times for COMPlog and ETDRS were 95 and 85 s and 66 and 56 s respectively in the paediatric and adult groups and 120 s for the E-ETDRS measurements on adults. Discussion: COMPlog measurements agree well with and are similarly reliable to the gold standard ETDRS chart with comparable test times. E-ETDRS algorithm measurements took approximately twice as long.
引用
收藏
页码:241 / 244
页数:4
相关论文
共 17 条
[1]  
ARDITI A, 1993, INVEST OPHTH VIS SCI, V34, P120
[2]  
BAILEY IL, 1991, INVEST OPHTH VIS SCI, V32, P422
[3]  
BAILEY IL, 1976, AM J OPTOM PHYS OPT, V53, P740
[4]   A computerized method of visual acuity testing: Adaptation of the early treatment of diabetic retinopathy study testing protocol [J].
Beck, RW ;
Moke, PS ;
Turpin, AH ;
Ferris, FL ;
Sangiovanni, JP ;
Johnson, CA ;
Birch, EE ;
Chandler, DL ;
Cox, TA ;
Blair, RC ;
Kraker, RT .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2003, 135 (02) :194-205
[5]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[6]   Reliability of the electronic early treatment diabetic retinopathy study testing protocol in children 7 to <13 years old [J].
Cotter, SA ;
Chu, RH ;
Chandler, DL ;
Beck, RW ;
Holmes, JM ;
Rice, ML ;
Hertle, RW ;
Birch, EE ;
Moke, PS .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2003, 136 (04) :655-661
[7]   THE USE OF ACCURATE VISUAL-ACUITY MEASUREMENTS IN CLINICAL ANTI-CATARACT FORMULATION TRIALS [J].
ELLIOTT, DB ;
SHERIDAN, M .
OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 1988, 8 (04) :397-401
[8]   NEW VISUAL-ACUITY CHARTS FOR CLINICAL RESEARCH [J].
FERRIS, FL ;
KASSOFF, A ;
BRESNICK, GH ;
BAILEY, I .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1982, 94 (01) :91-96
[9]   Development of a clinically feasible logMAR alternative to the Snellen chart: performance of the ''compact reduced logMAR'' visual acuity chart in amblyopic children [J].
Laidlaw, DAH ;
Abbott, A ;
Rosser, DA .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2003, 87 (10) :1232-1234
[10]   VALIDITY AND RELIABILITY OF VISUAL-ACUITY MEASUREMENTS [J].
LOVIEKITCHIN, JE .
OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 1988, 8 (04) :363-370