The dynamic vacuum orthosis:: a functional and economical benefit?

被引:8
作者
Franke, Jochen [1 ]
Goldhahn, Sabine [2 ]
Audige, Laurent [2 ]
Kohler, Henry [1 ]
Wentzensen, Andreas [1 ]
机构
[1] BG Trauma Ctr Ludwigshafen, D-67071 Ludwigshafen, Germany
[2] AO Clin Invest & Documentat, CH-7270 Davos, Switzerland
关键词
D O I
10.1007/s00264-006-0306-5
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
In a prospective, randomised study, 27 patients with internally fixed ankle fractures were treated post-operatively for a period of six weeks by application of either a new dynamic vacuum orthosis with permitted mobilisation to 10 degrees-0 degrees-10 degrees at the ankle joint or a synthetic cast. Full weight bearing was allowed in both groups after two weeks. The cast group was prescribed four weeks of physiotherapy following six weeks of immobilisation. After ten weeks, the Olerud and Molander score showed a significant difference in favour of the orthosis. Early functional outcomes were significantly better for this group after six weeks and ten weeks. Patients in the orthosis group who were in formal employment returned to work 24 days earlier than those in the control group. Treatment of the orthosis patients took up three to four times less working time for the medical personnel. A saving of 38 euros on directly ascertainable costs could be evaluated. Rehabilitation of ankle fractures with a dynamic vacuum orthosis leads to better early functional results and greater patient satisfaction. The orthosis fulfils the conditions for early return to work. Treatment with an orthosis not only reduces working time for medical personnel but economises on expenditure for treatment and rehabilitation.
引用
收藏
页码:153 / 158
页数:6
相关论文
共 26 条
[1]   EARLY MOBILIZATION OF OPERATED ON ANKLE FRACTURES - PROSPECTIVE, CONTROLLED-STUDY OF 40 BIMALLEOLAR CASES [J].
AHL, T ;
DALEN, N ;
LUNDBERG, A ;
BYLUND, C .
ACTA ORTHOPAEDICA SCANDINAVICA, 1993, 64 (01) :95-99
[2]  
AKESON WH, 1987, CLIN ORTHOP RELAT R, P28
[3]  
Augat P, 1996, CLIN ORTHOP RELAT R, P194
[4]  
*BASIS I TUEV PROD, 2005, 08942401 BASIS I TUE
[5]  
*BASIS I TUEV PROD, 2005, 08962402 BASIS I TUE
[6]  
CIMINO W, 1991, CLIN ORTHOP RELAT R, P152
[7]   Conservative functional treatment of ankle fractures [J].
Dietrich, A ;
Lill, H ;
Engel, T ;
Schönfelder, M ;
Josten, C .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2002, 122 (03) :165-168
[8]   Functional outcome of surgery for fractures of the ankle - A prospective, randomised comparison of management in a cast or a functional brace [J].
Egol, KA ;
Dolan, R ;
Koval, KJ .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2000, 82B (02) :246-249
[9]  
Goodship AE, 1998, CLIN ORTHOP RELAT R, pS105
[10]   Site-specific bone measurements in patients with ankle fracture [J].
Ingle, BM ;
Eastell, R .
OSTEOPOROSIS INTERNATIONAL, 2002, 13 (04) :342-347