Tunable external cavity quantum cascade laser for the simultaneous determination of glucose and lactate in aqueous phase

被引:56
作者
Brandstetter, Markus [1 ]
Genner, Andreas [1 ]
Anic, Kresimir [1 ]
Lendl, Bernhard [1 ]
机构
[1] Vienna Univ Technol, Inst Chem Technol & Analyt, A-1060 Vienna, Austria
关键词
BLOOD-GLUCOSE; TRANSMISSION SPECTROMETRY; WHOLE-BLOOD; SPECTROSCOPY; CONSTITUENTS; SYSTEMS;
D O I
10.1039/c0an00532k
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A room temperature operated pulsed external-cavity (EC) quantum cascade laser (QCL) was used for mid-infrared (mid-IR) transmission measurements of glucose and lactate in aqueous solution. The high spectral power density of the EC-QCL (ranging from 1-350 mW) over a wide tuning range (1030 1230 cm(-1)) allowed transmission measurements through optical paths of 130 mu m and more. This is a significant improvement in terms of robustness of the measurement setup, especially when samples containing cells or other particles, as is the case for biofluids, are to be analyzed. The broad tuning range furthermore permitted multi-analyte detection based on multivariate calibrations. Promising results on the simultaneous determination of glucose (c = 0-800 mg dL(-1)) and sodium-lactate (c = 0-224 mg dL(-1)) in aqueous solutions in the presence of the interferents maltose and xylose are reported. A partial least squares (PLS) calibration model was calculated which was able to predict the glucose concentration with a root mean square error of prediction (RMSEP) of 9.4 mg dL (1), as proved by external validation. Due to their small size and room temperature operation, EC-QCLs offer an attractive alternative regarding the way mid-IR measurements are carried out. This may be of special importance for new reagent-free bedside monitoring systems.
引用
收藏
页码:3260 / 3265
页数:6
相关论文
共 23 条
[1]   EVALUATING CLINICAL ACCURACY OF SYSTEMS FOR SELF-MONITORING OF BLOOD-GLUCOSE [J].
CLARKE, WL ;
COX, D ;
GONDERFREDERICK, LA ;
CARTER, W ;
POHL, SL .
DIABETES CARE, 1987, 10 (05) :622-628
[2]   Quantum cascade lasers in chemical physics [J].
Curl, Robert F. ;
Capasso, Federico ;
Gmachl, Claire ;
Kosterev, Anatoliy A. ;
McManus, Barry ;
Lewicki, Rafal ;
Pusharsky, Michael ;
Wysocki, Gerard ;
Tittel, Frank K. .
CHEMICAL PHYSICS LETTERS, 2010, 487 (1-3) :1-18
[3]   Applications of FT-IR spectrometry to plasma contents analysis and monitoring [J].
Déléris, G ;
Petibois, C .
VIBRATIONAL SPECTROSCOPY, 2003, 32 (01) :129-136
[4]   QUANTUM CASCADE LASER [J].
FAIST, J ;
CAPASSO, F ;
SIVCO, DL ;
SIRTORI, C ;
HUTCHINSON, AL ;
CHO, AY .
SCIENCE, 1994, 264 (5158) :553-556
[5]   EFFECTIVE THICKNESS OF BULK MATERIALS AND OF THIN FILMS FOR INTERNAL REFLECTION SPECTROSCOPY [J].
HARRICK, NJ ;
DUPRE, FK .
APPLIED OPTICS, 1966, 5 (11) :1739-&
[6]   Towards reagent-free blood glucose monitoring using micro-dialysis and infrared transmission spectrometry [J].
Heise, H. Michael ;
Damm, Uwe ;
Vogt, Oliver ;
Kondepati, Venkata R. .
VIBRATIONAL SPECTROSCOPY, 2006, 42 (01) :124-129
[7]   MULTIVARIATE DETERMINATION OF GLUCOSE IN WHOLE-BLOOD BY ATTENUATED TOTAL REFLECTION INFRARED-SPECTROSCOPY [J].
HEISE, HM ;
MARBACH, R ;
JANATSCH, G ;
KRUSEJARRES, JD .
ANALYTICAL CHEMISTRY, 1989, 61 (18) :2009-2015
[8]  
HEISE HM, 2002, HDB VIBRATIONAL SPEC, V5, P3280
[9]   Clinical chemistry without reagents?: An infrared spectroscopic technique for determination of clinically relevant constituents of body fluids [J].
Hosafci, Gamze ;
Klein, Oliver ;
Oremek, Gerhard ;
Maentele, Werner .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2007, 387 (05) :1815-1822
[10]   External cavity quantum cascade laser [J].
Hugi, Andreas ;
Maulini, Richard ;
Faist, Jerome .
SEMICONDUCTOR SCIENCE AND TECHNOLOGY, 2010, 25 (08)