Radiation Exposure to the Surgeon During Percutaneous Pedicle Screw Placement

被引:102
作者
Mroz, Thomas E. [1 ,2 ,3 ,4 ]
Abdullah, Kalil G. [2 ,3 ]
Steinmetz, Michael P. [2 ,3 ]
Klineberg, Eric O. [5 ]
Lieberman, Isador H.
机构
[1] Cleveland Clin, Neurol Inst, Ctr Spine Hlth, Cleveland, OH 44195 USA
[2] Cleveland Clin, Lerner Coll Med, Cleveland, OH 44195 USA
[3] Cleveland Clin, Dept Neurol Surg, Cleveland, OH 44195 USA
[4] Cleveland Clin, Dept Orthopaed Surg, Cleveland, OH 44195 USA
[5] Univ Calif Davis, Dept Orthoped Surg, Sacramento, CA 95817 USA
来源
JOURNAL OF SPINAL DISORDERS & TECHNIQUES | 2011年 / 24卷 / 04期
关键词
radiation exposure; percutaneous pedicle screws; minimally invasive surgery; TOTAL-BODY IRRADIATION; ATOMIC-BOMB SURVIVORS; CANCER-RISKS; IONIZING-RADIATION; SKIN INJURIES; FLUOROSCOPY; PATIENT; VERTEBROPLASTY;
D O I
10.1097/BSD.0b013e3181eed618
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design: In-vitro radiation exposure study. Objective: To determine the radiation exposure to the eyes, extremities, and deep tissue during percutaneous pedicle screw placement. Summary of Background Data: Image-guided minimally invasive spinal surgery is typically performed with the use of fluoroscopy, exposing the surgeon and patient to ionizing radiation. The radiation dose to the surgeon has not been reported and risk to the surgeon performing this procedure over the long term is uncertain. Methods: Percutaneous pedicle screws were placed in a cadaveric specimen from L2-S1 bilaterally using a cannulated pedicle screw system. Two fluoroscopes were used in the anteroposterior and lateral planes. The surgeon wore a thermolucent dosimeter ring on the right hand and badge over the left chest beneath the lead apron. Complete surgical time was recorded and a computed tomography scan was performed to assess screw placement. Radiation exposure was measured for total time of fluoroscopy use; average exposure per screw, surgical level, and dose to the eyes was calculated. This data was used to define the safety of percutaneous pedicle screw placement. Results: Total fluoroscope time for placement of 10 percutaneous pedicle screws was 4 minutes 56 seconds (29 s per screw). The protected dosimeter recorded less than the reportable dose. The ring dosimeter recorded 103 mREM, or 10.3 mREM per screw placed. All screws were within the bone confines with acceptable trajectory. Exposure to the eyes was 2.35 mREM per screw. Conclusions: On the basis of this data, percutaneous pedicle screw placement seems to be safe. A surgeon would exceed occupational exposure limit for the eyes and extremities by placing 4854 and 6396 screws percutaneously, respectively. Lead protected against radiation exposure during screw placement. The "hands-off" technique used in this study is recommended to minimize radiation exposure. Lead aprons, thyroid shields, and leaded glasses are recommended for this procedure.
引用
收藏
页码:264 / 267
页数:4
相关论文
共 27 条
[1]   Minimally Invasive Multilevel Percutaneous Correction and Fusion for Adult Lumbar Degenerative Scoliosis A Technique and Feasibility Study [J].
Anand, Neel ;
Baron, Eli M. ;
Thaiyananthan, Gowriharan ;
Khalsa, Kunwar ;
Goldstein, Theodore B. .
JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2008, 21 (07) :459-467
[2]  
[Anonymous], 2000, REPORT GEN ASSEMBLY
[3]  
[Anonymous], 1995, MED EFFECTS IONIZING
[4]   Cataractogenesis after total body irradiation [J].
Belkacemi, Y ;
Ozsahin, M ;
Pene, F ;
Rio, B ;
Laporte, JP ;
Leblond, V ;
Touboul, E ;
Schlienger, M ;
Gorin, NC ;
Laugier, A .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1996, 35 (01) :53-60
[5]   Cancer risks attributable to low doses of ionizing radiation: Assessing what we really know [J].
Brenner, DJ ;
Doll, R ;
Goodhead, DT ;
Hall, EJ ;
Land, CE ;
Little, JB ;
Lubin, JH ;
Preston, DL ;
Preston, RJ ;
Puskin, JS ;
Ron, E ;
Sachs, RK ;
Samet, JM ;
Setlow, RB ;
Zaider, M .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 2003, 100 (24) :13761-13766
[6]   Percutaneous pedicle screws significantly decrease muscle damage and operative time: surgical technique makes a difference! [J].
Dickerman, Rob D. ;
Reynolds, Ashley S. ;
Tackett, Jennifer ;
Winters, Karl ;
Alvarado, Cecilio .
EUROPEAN SPINE JOURNAL, 2008, 17 (10) :1398-1398
[7]   Patient and staff dosimetry in vertebroplasty [J].
Fitousi, Niki T. ;
Efstathopoulos, Efstathios P. ;
Delis, Harry B. ;
Kottou, Sofia ;
Kelekis, Alexis D. ;
Panayiotakis, George S. .
SPINE, 2006, 31 (23) :E884-E889
[8]  
FRITZSCHE AF, 1992, VOL KOLN
[9]   Cancer risks from diagnostic radiology [J].
Hall, E. J. ;
Brenner, D. J. .
BRITISH JOURNAL OF RADIOLOGY, 2008, 81 (965) :362-378
[10]   Radiation exposure to the surgeon during fluoroscopically assisted percutaneous vertebroplasty - A prospective study [J].
Harstall, R ;
Heini, PF ;
Mini, RL ;
Orler, R .
SPINE, 2005, 30 (16) :1893-1898