Out-of-field dose and its constituent components for a 1.5 T MR-Linac

被引:1
作者
Yang, Bin [1 ]
Tang, Ka Keung [1 ]
Huang, Chen-Yu [1 ]
Geng, Hui [1 ]
Lam, Wai Wang [1 ]
Wong, Yeung Sum [1 ]
Tse, Mei Yan [1 ]
Lau, Ka Ki [1 ]
Cheung, Kin Yin [1 ]
Yu, Siu Ki [1 ]
机构
[1] Hong Kong Sanat & Hosp, Med Phys Dept, 2 Village Rd, Hong Kong, Peoples R China
关键词
out-of-field dose; peripheral dose; scatter; leakage; MR-Linac; IMRT; RADIATION-THERAPY; MAGNETIC-FIELD; RADIOTHERAPY; ACCELERATOR; PHOTON; ELECTRON; DISTRIBUTIONS; IRRADIATION; SCANNER; BEAMS;
D O I
10.1088/1361-6560/ac3346
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This study aims to quantify the relative contributions of phantom scatter, collimator scatter and head leakage to the out-of-field doses (OFDs) of both static fields and clinical intensity-modulated radiation therapy (IMRT) treatments in a 1.5 T MR-Linac. The OFDs of static fields were measured at increasing distances from the field edge in an MR-conditional water phantom. Inline scans at depths of dmax (14 mm), 50 and 100 mm were performed for static fields of 5 x 5, 10 x 10 and 15 x 15 cm(2) under three different conditions: full scatter, with phantom scatter prevented, and head leakage only. Crossline scans at isocenter and offset positions were performed in full scatter condition. EBT3 radiochromic films were placed at 100 mm depth of solid water phantom to measure the OFD of clinical IMRT plans. All water tank data were normalized to Dmax of a 10 x 10 cm(2) field and the film results were presented as a fraction of the target mean dose. The OFD in the inline direction varied from 3.5% (15 x 15 cm(2), 100 mm depth, 50 mm distance) to 0.014% (5 x 5 cm(2), dmax, 400 mm distance). For all static fields, the collimator scatter was higher than the phantom scatter and head leakage at a distance of 100-400 mm. Head leakage remained the smallest among the three components, except at long distances (>375 mm) with small field size. Compared to the inline scans, the crossline scans at the isocenter showed higher doses at distances longer than 80 mm. All crossline profiles at longitudinal offset positions showed a cone shape with laterally shifted maxima. The OFD of IMRT deliveries varied with different target size. For prostate stereotactic body radiation therapy (SBRT) treatment, the OFD decreased from 2% to 0.03% at a distance of 50-500 mm. The OFDs have been measured for a 1.5 T MR-Linac. The presented dosimetric data are valuable for radiation safety assessments on patients treated with the MR-Linac, such as evaluating carcinogenic risk and radiation exposure to cardiac implantable electronic devices.
引用
收藏
页数:11
相关论文
共 33 条
[1]   Monte Carlo study of in-field and out-of-field dose distributions from a linear accelerator operating with and without a flattening-filter [J].
Almberg, S. S. ;
Frengen, J. ;
Lindmo, T. .
MEDICAL PHYSICS, 2012, 39 (08) :5194-5203
[2]   The effect of magnetic field strength on the response of Gafchromic EBT-3 film [J].
Billas, Ilias ;
Bouchard, Hugo ;
Oelfke, Uwe ;
Duane, Simon .
PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (06)
[3]   A new Monte Carlo model of a Cyberknife® system for the precise determination of out-of-field doses [J].
Colnot, J. ;
Barraux, V ;
Loiseau, C. ;
Berejny, P. ;
Batalla, A. ;
Gschwind, R. ;
Huet, C. .
PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (19)
[4]   Technical Report: Evaluation of peripheral dose for flattening filter free photon beams [J].
Covington, E. L. ;
Ritter, T. A. ;
Moran, J. M. ;
Owrangi, A. M. ;
Prisciandaro, J. I. .
MEDICAL PHYSICS, 2016, 43 (08) :4789-4796
[5]   Spiraling contaminant electrons increase doses to surfaces outside the photon beam of an MRI-linac with a perpendicular magnetic field [J].
Hackett, S. L. ;
van Asselen, B. ;
Wolthaus, J. W. H. ;
Bluemink, J. J. ;
Ishakoglu, K. ;
Kok, J. ;
Lagendijk, J. J. W. ;
Raaymakers, B. W. .
PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (09)
[6]   Radiation Dose-Response for Risk of Myocardial Infarction in Breast Cancer Survivors [J].
Jacobse, Judy N. ;
Duane, Frances K. ;
Boekel, Naomi B. ;
Schaapveld, Michael ;
Hauptmann, Michael ;
Hooning, Maartje J. ;
Seynaeve, Caroline M. ;
Baaijens, Margreet H. A. ;
Gietema, Jourik A. ;
Darby, Sarah C. ;
van Leeuwen, Flora E. ;
Aleman, Berthe M. P. ;
Taylor, Carolyn W. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2019, 103 (03) :595-604
[7]   Flattening filter free beams in SBRT and IMRT: Dosimetric assessment of peripheral doses [J].
Kragl, Gabriele ;
Baier, Franziska ;
Lutz, Steffen ;
Albrich, David ;
Dalaryd, Marten ;
Kroupa, Bernhard ;
Wiezorek, Tilo ;
Knoos, Tommy ;
Georg, Dietmar .
ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2011, 21 (02) :91-101
[8]   Out-of-field photon and neutron dose equivalents from step-and-shoot intensity-modulated radiation therapy [J].
Kry, SF ;
Salehpour, M ;
Followill, DS ;
Stovall, M ;
Kuban, DA ;
White, RA ;
Rosen, II .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 62 (04) :1204-1216
[9]   The calculated risk of fatal secondary malignancies from intensity-modulated radiation therapy [J].
Kry, SF ;
Salehpour, M ;
Followill, DS ;
Stovall, M ;
Kuban, DA ;
White, RA ;
Rosen, II .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 62 (04) :1195-1203
[10]   AAPM TG 158: Measurement and calculation of doses outside the treated volume from external-beam radiation therapy [J].
Kry, Stephen F. ;
Bednarz, Bryan ;
Howell, Rebecca M. ;
Dauer, Larry ;
Followill, David ;
Klein, Eric ;
Paganetti, Harald ;
Wang, Brian ;
Wuu, Cheng-Shie ;
Xu, X. George .
MEDICAL PHYSICS, 2017, 44 (10) :e391-e429