An adaptive spot placement method on Cartesian grid for pencil beam scanning proton therapy

被引:8
作者
Lin, Bowen [1 ]
Fu, Shujun [1 ]
Lin, Yuting [2 ]
Rotondo, Ronny L. [2 ]
Huang, Weizhang [3 ]
Li, Harold H. [2 ]
Chen, Ronald C. [2 ]
Gao, Hao [2 ]
机构
[1] Shandong Univ, Sch Math, Jinan, Peoples R China
[2] Univ Kansas, Med Ctr, Dept Radiat Oncol, Kansas City, KS 66103 USA
[3] Univ Kansas, Dept Math, Lawrence, KS 66045 USA
关键词
proton therapy; treatment planning; inverse optimization; OPTIMIZATION; IMPROVEMENT; IMPACT; SIZE;
D O I
10.1088/1361-6560/ac3b65
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Pencil beam scanning proton radiotherapy (RT) offers flexible proton spot placement near treatment targets for delivering tumoricidal radiation dose to tumor targets while sparing organs-at-risk. Currently the spot placement is mostly based on a non-adaptive sampling (NS) strategy on a Cartesian grid. However, the spot density or spacing during NS is a constant for the Cartesian grid that is independent of the geometry of tumor targets, and thus can be suboptimal in terms of plan quality (e.g. target dose conformality) and delivery efficiency (e.g. number of spots). This work develops an adaptive sampling (AS) spot placement method on the Cartesian grid that fully accounts for the geometry of tumor targets. Compared with NS, AS places (1) a relatively fine grid of spots at the boundary of tumor targets to account for the geometry of tumor targets and treatment uncertainties (setup and range uncertainty) for improving dose conformality, and (2) a relatively coarse grid of spots in the interior of tumor targets to reduce the number of spots for improving delivery efficiency and robustness to the minimum-minitor-unit (MMU) constraint. The results demonstrate that (1) AS achieved comparable plan quality with NS for regular MMU and substantially improved plan quality from NS for large MMU, using merely about 10% of spots from NS, where AS was derived from the same Cartesian grid as NS; (2) on the other hand, with similar number of spots, AS had better plan quality than NS consistently for regular and large MMU.
引用
收藏
页数:18
相关论文
共 26 条
[1]  
Albertini Francesca., 2009, Biomedical Mathematics: Promising Directions in Imaging, Therapy Planning, and Inverse Problems, V1, P1
[2]   Distributed optimization and statistical learning via the alternating direction method of multipliers [J].
Boyd S. ;
Parikh N. ;
Chu E. ;
Peleato B. ;
Eckstein J. .
Foundations and Trends in Machine Learning, 2010, 3 (01) :1-122
[3]   Incorporating deliverable monitor unit constraints into spot intensity optimization in intensity-modulated proton therapy treatment planning [J].
Cao, Wenhua ;
Lim, Gino ;
Li, Xiaoqiang ;
Li, Yupeng ;
Zhu, X. Ronald ;
Zhang, Xiaodong .
PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (15) :5113-5125
[4]  
Freitag LA, 1997, INT J NUMER METH ENG, V40, P3979, DOI 10.1002/(SICI)1097-0207(19971115)40:21<3979::AID-NME251>3.0.CO
[5]  
2-9
[6]   Simultaneous dose and dose rate optimization (SDDRO) of the FLASH effect for pencil-beam-scanning proton therapy [J].
Gao, Hao ;
Liu, Jiulong ;
Lin, Yuting ;
Gan, Gregory N. ;
Pratx, Guillem ;
Wang, Fen ;
Langen, Katja ;
Bradley, Jeffrey D. ;
Rotondo, Ronny L. ;
Li, Harold H. ;
Chen, Ronald C. .
MEDICAL PHYSICS, 2022, 49 (03) :2014-2025
[7]   Simultaneous dose and dose rate optimization (SDDRO) for FLASH proton therapy [J].
Gao, Hao ;
Lin, Bowen ;
Lin, Yuting ;
Fu, Shujun ;
Langen, Katja ;
Liu, Tian ;
Bradley, Jeffery .
MEDICAL PHYSICS, 2020, 47 (12) :6388-6395
[8]   Technical Note: Plan-delivery-time constrained inverse optimization method with minimum-MU-per-energy-layer (MMPEL) for efficient pencil beam scanning proton therapy [J].
Gao, Hao ;
Clasie, Benjamin ;
McDonald, Mark ;
Langen, Katja M. ;
Liu, Tian ;
Lin, Yuting .
MEDICAL PHYSICS, 2020, 47 (09) :3892-3897
[9]   Minimum MU optimization (MMO): an inverse optimization approach for the PBS minimum MU constraint [J].
Gao, Hao ;
Clasie, Benjamin ;
Liu, Tian ;
Lin, Yuting .
PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (12)
[10]   Hybrid proton-photon inverse optimization with uniformity-regularized proton and photon target dose [J].
Gao, Hao .
PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (10)