The impact of extra-column band broadening on the chromatographic efficiency of 5 cm long narrow-bore very efficient columns

被引:83
作者
Fekete, Szabolcs [1 ]
Fekete, Jeno [1 ]
机构
[1] Budapest Univ Technol & Econ, Dept Inorgan & Analyt Chem, H-1111 Budapest, Hungary
关键词
Column efficiency; Extra-column band-broadening; Core-shell particles; Sub-2 mu m particles; Monolithic columns; UHPLC; PERFORMANCE LIQUID-CHROMATOGRAPHY; PARTICLES; PRESSURE; SEPARATION; SUPPORTS;
D O I
10.1016/j.chroma.2011.06.045
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Small columns packed with core-shell and sub-2 mu m totally porous particles and monolith columns are very popular to conduct fast and efficient chromatographic separations. In order to carry out fast separations, short (2-5 cm) and narrow-bore (2-2.1 mm) columns are used to decrease the analyte retention volume. Beside the column efficiency, another significant issue is the extra-column band-spreading. The extra-column dispersion of a given LC system can dramatically decrease the performance of a small very efficient column. The aim of this study was to compare the extracolumn peak variance contribution of several commercially available LC systems. The efficiency loss of three different type 5 cm long narrow bore, very efficient columns (monolith, sub-2 mu m fully porous and sub-2 mu m core-shell packing) as a function of extra-column peak variance, and as a function of flow rate and also kinetic plots (analysis time versus apparent column efficiency) are presented. (C) 2011 Elsevier B.V. All rights reserved.
引用
收藏
页码:5286 / 5291
页数:6
相关论文
共 19 条
[1]   HIGH-SPEED HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY OF PEPTIDES AND PROTEINS [J].
CHEN, H ;
HORVATH, C .
JOURNAL OF CHROMATOGRAPHY A, 1995, 705 (01) :3-20
[2]   Geometry-independent plate height representation methods for the direct comparison of the kinetic performance of LC supports with a different size or morphology [J].
Desmet, G ;
Clicq, D ;
Gzil, P .
ANALYTICAL CHEMISTRY, 2005, 77 (13) :4058-4070
[3]   Characteristics of superficially-porous silica particles for fast HPLC:: Some performance comparisons with sub-2-μm particles [J].
DeStefano, J. J. ;
Langlois, T. J. ;
Kirkland, J. J. .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 2008, 46 (03) :254-260
[4]   Efficiency of the new sub-2 μm core-shell (Kinetex™) column in practice, applied for small and large molecule separation [J].
Fekete, Szabolcs ;
Ganzler, Katalin ;
Fekete, Jeno .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2011, 54 (03) :482-490
[5]  
Felinger A., 1998, DATA ANAL SIGNAL PRO
[6]   Effects of extra-column band spreading, liquid chromatography system operating pressure, and column temperature on the performance of sub-2-μm porous particles [J].
Fountain, Kenneth J. ;
Neue, Uwe D. ;
Grumbach, Eric S. ;
Diehl, Diane M. .
JOURNAL OF CHROMATOGRAPHY A, 2009, 1216 (32) :5979-5988
[7]   Influence of the errors made in the measurement of the extra-column volume on the accuracies of estimates of the column efficiency and the mass transfer kinetics parameters [J].
Gritti, Fabrice ;
Felinger, Attila ;
Guiochon, Georges .
JOURNAL OF CHROMATOGRAPHY A, 2006, 1136 (01) :57-72
[8]   On the extra-column band-broadening contributions of modern, very high pressure liquid chromatographs using 2 1 mm I D columns packed with sub-2 μm particles [J].
Gritti, Fabrice ;
Guiochon, Georges .
JOURNAL OF CHROMATOGRAPHY A, 2010, 1217 (49) :7677-7689
[9]   Achieving the full performance of highly efficient columns by optimizing conventional benchmark high-performance liquid chromatography instruments [J].
Gritti, Fabrice ;
Sanchez, Carl A. ;
Farkas, Tivadar ;
Guiochon, Georges .
JOURNAL OF CHROMATOGRAPHY A, 2010, 1217 (18) :3000-3012
[10]   Performance of columns packed with the new shell particles, Kinetex-C18 [J].
Gritti, Fabrice ;
Leonardis, Irene ;
Shock, David ;
Stevenson, Paul ;
Shalliker, Andrew ;
Guiochon, Georges .
JOURNAL OF CHROMATOGRAPHY A, 2010, 1217 (10) :1589-1603