Simulations of the Stent Artifacts in Magnetic Resonance Imaging

被引:10
作者
Guo, Yan [1 ]
Jiang, Xiaohua [1 ]
机构
[1] Tsinghua Univ, Dept Elect Engn, Beijing 100084, Peoples R China
关键词
Artifacts; finite-element methods; MRI; stents; SUSCEPTIBILITY;
D O I
10.1109/TMAG.2011.2176473
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This paper presents an approach to simulate the stent artifacts in magnetic resonance imaging (MRI) based on electromagnetic (EM) analysis. The metallic stent disturbs the static magnetic field (B0) and the radiofrequency (RF) field during the MRI procedure, which consequently results in susceptibility artifacts and RF shielding artifacts in MR image. Both the static and RF field distributions with the metallic stent in a uniform imaging phantom are calculated using the commercial finite element method (FEM) software JMAG 10.0. The images with stent artifacts are simulated by an MRI simulator, according to the calculated field distributions. The susceptibility artifacts and RF shielding artifacts are investigated at several conditions respectively. It can be concluded that the stent made of Ni-Ti alloy with a small magnetic susceptibility is promising to obtain the inside information with small susceptibility artifacts, and the stent with the open-cell design causes smaller RF shielding artifacts in MRI than the close-cell design.
引用
收藏
页码:659 / 662
页数:4
相关论文
共 12 条
[1]   Improved lumen visualization in metallic vascular implants by reducing RF artifacts [J].
Bartels, LW ;
Bakker, CJG ;
Viergever, MA .
MAGNETIC RESONANCE IN MEDICINE, 2002, 47 (01) :171-180
[2]   Reduction of Artifact of Metallic Implant in Magnetic Resonance Imaging by Coating of Diamagnetic Material [J].
Gao, Yanhui ;
Muramatsu, Kazuhiro ;
Kushibe, Atsumichi ;
Yamazaki, Keita ;
Chiba, Akihiko ;
Yamamoto, Toru .
IEEE TRANSACTIONS ON MAGNETICS, 2009, 45 (10) :4837-4840
[3]   Physics of magnetic resonance imaging: from spin to pixel [J].
Gossuin, Yves ;
Hocq, Aline ;
Gillis, Pierre ;
Vuong, Quoc Lam .
JOURNAL OF PHYSICS D-APPLIED PHYSICS, 2010, 43 (21)
[4]   Coronary arterial stents: Safety and artifacts during MR imaging [J].
Hug, J ;
Nagel, E ;
Bornstedt, A ;
Schnackenburg, B ;
Oswald, H ;
Fleck, E .
RADIOLOGY, 2000, 216 (03) :781-787
[5]   Magnetic Resonance Imaging Near Metal Implants [J].
Koch, K. M. ;
Hargreaves, B. A. ;
Pauly, K. Butts ;
Chen, W. ;
Gold, G. E. ;
King, K. F. .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2010, 32 (04) :773-787
[6]   In vitro comparison of different carotid artery stents: a pixel-by-pixel analysis using CT angiography and contrast-enhanced MR angiography at 1.5 and 3 T [J].
Lettau, Michael ;
Sauer, Annett ;
Heiland, Sabine ;
Rohde, Stefan ;
Reinhardt, Julia ;
Bendszus, Martin ;
Haehnel, Stefan .
NEURORADIOLOGY, 2010, 52 (09) :823-830
[7]  
Liang Z.-P., 2000, PRINCIPLES MAGNETIC, P76
[8]   Coronary stents: A materials perspective [J].
Mani, Gopinath ;
Feldman, Marc D. ;
Patel, Devang ;
Agrawal, C. Mauli .
BIOMATERIALS, 2007, 28 (09) :1689-1710
[9]   A LINEAR CLASS OF LARGE-TIP-ANGLE SELECTIVE EXCITATION PULSES [J].
PAULY, J ;
NISHIMURA, D ;
MACOVSKI, A .
JOURNAL OF MAGNETIC RESONANCE, 1989, 82 (03) :571-587
[10]   Magnetic resonance imaging of stents: Quantitative in vitro examination at 3 Tesla [J].
Reinhardt, Julia ;
Nguyen-Trong, Thien-Hoa ;
Haehnel, Stefan ;
Bellemann, Matthias E. ;
Heiland, Sabine .
ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2009, 19 (04) :278-287