Development of a reliable automatic speed control system for rotary blood pumps

被引:72
作者
Vollkron, M
Schima, H
Huber, L
Benkowski, R
Morello, G
Wieselthaler, G
机构
[1] Med Univ Vienna, Dept Biomed Res, A-1090 Vienna, Austria
[2] Med Univ Vienna, Dept Cardiothorac Surg, Vienna, Austria
[3] Med Univ Vienna, Ludwig Boltzmann Inst Cardiosurg Res, Vienna, Austria
[4] Med Univ Vienna, Dept Biomed Engn & Phys, Vienna, Austria
[5] MicroMed Technol Inc, Houston, TX USA
关键词
D O I
10.1016/j.healun.2005.02.004
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Axial blood pumps have been very successfully introduced into the arena of prolonged clinical support. However, they do not offer inherent load-responsive mechanisms for adjusting pumping performance to venous return and changes in physiologic requirements of the patient. To provide for these adjustments we developed an algorithm for demand-responsive pump control based on a reliable suction detection system. Methods: A PC-based system that analyzes pump performance based on available flow, heart rate and short-term performance history was developed. The physician defines levels of "desired flow" at rest and during exercise, depending on heart rate. In case this desired flow cannot be maintained due to limited venous return, the maximal available flow level is determined from an analysis of the actual pump data (flow, speed and power consumption). An expert system continuously checks the flow signal for any indication of suction. Periodic speed variations then adapt pump performance to the patient's condition. Results: First, stability and functionality were proven under various settings in vitro. The algorithms were then tested in 15 patients in intensive care, in the standard ward, and during bicycle exercise. The system reacted properly to demand changes, at exercise level, in response to coughing and at various Valsalva maneuvers. Suction could also be successfully prevented during severe arrhythmia and in patients with critical cardiac geometry. Exercise tests showed decreases in pulmonary arterial pressure (-22 +/- 9.9%) and pulmonary capillary wedge pressure (-42 +/- 18-54%), and an increase in pump flow (19 +/- 9.5%) and workload (8 +/- 6.1%), all when compared with constant-speed pumping. Conclusions: A closed-loop control system equipped with an expert system for reliable suction detection was developed that improves response to change in venous return for rotary. pump recipients. The system was robust, stable and safe under a wide range of everyday living conditions. J Heart Lung Transplant 2005;24:1878-85. Copyright (c) 2005 by the. International Society for Heart and Lung Transplantation.
引用
收藏
页码:1878 / 1885
页数:8
相关论文
共 18 条
[1]   Rotary blood pump flow spontaneously increases during exercise under constant pump speed: Results of a chronic study [J].
Akimoto, T ;
Yamazaki, K ;
Litwak, P ;
Litwak, KN ;
Tagusari, O ;
Mori, T ;
Antaki, JF ;
Kameneva, MV ;
Watach, MJ ;
Umezu, M ;
Tomioka, J ;
Kormos, RL ;
Koyanagi, H ;
Griffith, BP .
ARTIFICIAL ORGANS, 1999, 23 (08) :797-801
[2]   Physiologic control algorithms for rotary blood pumps using pressure sensor input [J].
Bullister, E ;
Reich, S ;
Sluetz, J .
ARTIFICIAL ORGANS, 2002, 26 (11) :931-938
[3]   Synergies between statistical data analysis and neural networks in the control of rotary blood pumps [J].
Dorffner, G ;
Stocklmayer, C ;
Schmidt, C ;
Schima, H .
CYBERNETICS AND SYSTEMS, 1997, 28 (03) :215-224
[4]   Computer simulation of sensorless fuzzy control of a rotary blood pump to assure normal physiology [J].
Fu, MH ;
Xu, LY .
ASAIO JOURNAL, 2000, 46 (03) :273-278
[5]   Controller for an axial flow blood pump [J].
Konishi, H ;
Antaki, JF ;
Amin, DV ;
Boston, JR ;
Kerrigan, JP ;
Mandarino, WA ;
Litwak, P ;
Yamazaki, K ;
Macha, M ;
Butler, KC ;
Borovetz, HS ;
Kormos, RL .
ARTIFICIAL ORGANS, 1996, 20 (06) :618-620
[6]  
MAAR H, 1993, THESIS TU VIENNA VIE
[7]  
Nakata K I, 2000, Ann Thorac Cardiovasc Surg, V6, P242
[8]   Clinical experience with the MicroMed DeBakey ventricular assist device [J].
Noon, GP ;
Morley, DL ;
Irwin, S ;
Abdelsayed, SV ;
Benkowski, RJ ;
Lynch, BE .
ANNALS OF THORACIC SURGERY, 2001, 71 (03) :S133-S138
[9]   Control strategy for rotary blood pumps [J].
Ohuchi, K ;
Kikugawa, D ;
Takahashi, K ;
Uemura, M ;
Nakamura, M ;
Murakami, T ;
Sakamoto, T ;
Takatani, S .
ARTIFICIAL ORGANS, 2001, 25 (05) :366-370
[10]  
SCHIMA H, 1992, ARTIF ORGANS, V16, P195