Agreement and Repeatability of Central and Peripheral Refraction by One Novel Multispectral-Based Refractor

被引:22
作者
Lu, Weicong [1 ]
Ji, Rongyuan [1 ]
Ding, Wenzhi [1 ]
Tian, Yuyin [1 ]
Long, Keli [1 ]
Guo, Zhen [1 ]
Leng, Lin [1 ]
机构
[1] Shandong First Med Univ, Qingdao Eye Hosp, Inst Eye,State Key Lab Cultivat Base, Shandong Prov Key Lab Ophthalmol, Qingdao, Peoples R China
关键词
agreement; repeatability; refraction; myopia; ophthalmology; MYOPIA; AUTOREFRACTION; PREVALENCE; ACCURACY; FOVEAL; SHAPE;
D O I
10.3389/fmed.2021.777685
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Purpose: To evaluate the repeatability of a multispectral-based refractor in central and peripheral refraction measurement, and to assess the agreement of such measurements with objective refraction (OR) and subjective refraction (SR) in patients with myopia.Methods: A total of 60 subjects were recruited in this prospective research. Patients were divided into three groups according to the refractive error. Next, the central and peripheral refraction parameters were measured using multispectral refractive tomography (MRT) before and after cycloplegia. In addition, OR and SR measurements were also performed. The intraobserver repeatability was analyzed using within-subject standard deviation (Sw), test-retest repeatability (TRT), and intraclass correlation coefficient (ICC). Agreement was evaluated using Bland-Altman plot and 95% limits of agreement (LoA).Results: The ICC value of central and peripheral refraction were all higher than 0.97 with or without cycloplegia. The peripheral refraction in the nasal, temporal, superior, and inferior quadrants was slightly worse than other parameters, with the largest error interval being 1.43 D. The 95% LoA of the central refraction and OR or SR ranged from -0.89 to 0.88 D and -1.24 to 1.16 D without cycloplegia, respectively, and from -0.80 to 0.42 D and -1.39 to -0.84 D under cycloplegia, respectively.Conclusions: The novel multispectral refraction topography demonstrated good repeatability in central and peripheral refraction. However, the refraction in the nasal, temporal, superior, and inferior quadrants were not as good as that of central and circle peripheral refraction.
引用
收藏
页数:12
相关论文
共 26 条
[1]   Repeatability and validity of the PowerRefractor and the Nidek AR600-A in an adult population with healthy eyes [J].
Allen, PM ;
Radhakrishnan, H ;
O'Leary, DJ .
OPTOMETRY AND VISION SCIENCE, 2003, 80 (03) :245-251
[2]   Myopia [J].
Baird, Paul N. ;
Saw, Seang-Mei ;
Lanca, Carla ;
Guggenheim, Jeremy A. ;
Smith III, Earl L. ;
Zhou, Xiangtian ;
Matsui, Kyoko-Ohno ;
Wu, Pei-Chang ;
Sankaridurg, Padmaja ;
Chia, Audrey ;
Rosman, Mohamad ;
Lamoureux, Ecosse L. ;
Man, Ryan ;
He, Mingguang .
NATURE REVIEWS DISEASE PRIMERS, 2020, 6 (01)
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]   The Effect of Spectacle Lenses Containing Peripheral Defocus on Refractive Error and Horizontal Eye Shape in the Guinea Pig [J].
Bowrey, Hannah E. ;
Zeng, Guang ;
Tse, Dennis Y. ;
Leotta, Amelia J. ;
Wu, Yi ;
To, Chi-Ho ;
Wildsoet, Christine F. ;
McFadden, Sally A. .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2017, 58 (05) :2705-2714
[5]   Comparison of Two Wavefront Autorefractors: Binocular Open-Field versus Monocular Closed-Field [J].
Carracedo, Gonzalo ;
Carpena-Torres, Carlos ;
Batres, Laura ;
Serramito, Maria ;
Gonzalez-Bergaz, Anahi .
JOURNAL OF OPHTHALMOLOGY, 2020, 2020
[6]   Diagnostic accuracy and variability of autorefraction by the Tracey Visual Function Analyzer and the Shin-Nippon NVision-K 5001 in relation to subjective refraction [J].
Cleary, G. ;
Spalton, D. J. ;
Patel, P. M. ;
Lin, P. -F. ;
Marshall, J. .
OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 2009, 29 (02) :173-181
[7]   Repeatability and accuracy of automated refraction: A comparison of the Nikon NRK-8000, the Nidek AR-1000, and subjective refraction [J].
Elliott, M ;
Simpson, T ;
Richter, D ;
Fonn, D .
OPTOMETRY AND VISION SCIENCE, 1997, 74 (06) :434-438
[8]   Lateral Pupil Alignment Tolerance in Peripheral Refractometry [J].
Fedtke, Cathleen ;
Ehrmann, Klaus ;
Ho, Arthur ;
Holden, Brien A. .
OPTOMETRY AND VISION SCIENCE, 2011, 88 (05) :E570-E579
[9]   Repeatability of ARK-30 in a pediatric population [J].
Hernandez-Moreno, Laura ;
Vallelado-Alvarez, Ana ;
Martin, Raul .
INDIAN JOURNAL OF OPHTHALMOLOGY, 2018, 66 (09) :1262-1267
[10]   Global Prevalence of Myopia and High Myopia and Temporal Trends from 2000 through 2050 [J].
Holden, Brien A. ;
Fricke, Timothy R. ;
Wilson, David A. ;
Jong, Monica ;
Naidoo, Kovin S. ;
Sankaridurg, Padmaja ;
Wong, Tien Y. ;
Naduvilath, Thomas J. ;
Resnikoff, Serge .
OPHTHALMOLOGY, 2016, 123 (05) :1036-1042