T1 fast acquisition relaxation mapping (T1-FARM):: An optimized reconstruction

被引:16
作者
Chen, ZQ [1 ]
Prato, FS [1 ]
McKenzie, C [1 ]
机构
[1] St Josephs Hlth Ctr, Lawson Res Inst, Dept Nucl Med & Magnet Resonance, London, ON N6A 4V2, Canada
关键词
MRI; non-Fourier k-space reconstruction; parametric reconstruction; (T-1) map;
D O I
10.1109/42.700728
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
Maps of spin lattice relaxation time (T-1) can be reconstructed directly from magnetic resonance imaging (MRI) k-space data measured with very short data acquisition times, e,g,, a data set for a 128 x 128 T-1 map can be acquired in less than 3 s using gradients with 10-T/m/s slew rate, In principle, this approach could be extended to quantitate other MRI parameters but current use is limited by the lack of precise, accurate and fast reconstruction. Using theoretical calculations, computer simulations, and experiments we have optimized a parametric reconstruction method using a Leverberg-Marquardt (L-M) algorithm and compared it to the quasi-Newton method originally used. We have found significant improvement using the L-M method provided T-1 is solved for directly without linearization, Reconstruction time was reduced by a factor of 60. Computer simulations show that the method has acceptable accuracy even in signals with 5% noise. Optimization included the investigation of the signal-to-noise (S/N) of each k-space data point and its impact on relative error of the reconstruction. This result indicates that rectangular k-space data could be collected for further reduction of data acquisition times. Determination of T-1 maps by direct parametric reconstruction of L-space data appears feasible and may stimulate further application of quantitative MRI.
引用
收藏
页码:155 / 160
页数:6
相关论文
共 12 条
  • [1] FAST AND PRECISE T1 IMAGING USING A TOMROP SEQUENCE
    BRIX, G
    SCHAD, LR
    DEIMLING, M
    LORENZ, WJ
    [J]. MAGNETIC RESONANCE IMAGING, 1990, 8 (04) : 351 - 356
  • [2] CHEN Z, 1996, P SMR, P1557
  • [3] JACOBY SL, 1972, ITERATIVE METHOD NON
  • [4] Levenberg K., 1944, Q APPL MATH, V2, P164, DOI [10.1090/QAM/10666, DOI 10.1090/QAM/10666]
  • [5] Look D. C., 1970, Review of Scientific Instruments, V41, P250, DOI 10.1063/1.1684482
  • [6] AN ALGORITHM FOR LEAST-SQUARES ESTIMATION OF NONLINEAR PARAMETERS
    MARQUARDT, DW
    [J]. JOURNAL OF THE SOCIETY FOR INDUSTRIAL AND APPLIED MATHEMATICS, 1963, 11 (02): : 431 - 441
  • [7] MCKENZIE CA, 1995, P SMR, P519
  • [8] Scales LE., 1985, INTRO NONLINEAR OPTI
  • [9] TECHNIQUES FOR THE MEASUREMENT OF THE LOCAL MYOCARDIAL EXTRACTION EFFICIENCY FOR INERT DIFFUSIBLE CONTRAST AGENTS SUCH AS GADOPENTATE DIMEGLUMINE
    TONG, CY
    PRATO, FS
    WISENBERG, G
    LEE, TY
    CARROLL, E
    SANDLER, D
    WILLS, J
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1993, 30 (03) : 332 - 336
  • [10] A NOVEL FAST T1-MAPPING METHOD
    TONG, CY