External audits of electron beams using mailed TLD dosimetry: preliminary results

被引:14
作者
Gomola, I
Van Dam, J
Isern-Verdum, J
Verstraete, J
Reymen, R
Dutreix, A
Davis, B
Huyskens, D
机构
[1] Univ Hosp Gasthuisberg, Sect Radiat Phys, Dept Oncol, B-3000 Louvain, Belgium
[2] Hosp Santa Creu & Sant Pau, E-08025 Barcelona, Spain
[3] Univ Zurich Hosp, CH-8091 Zurich, Switzerland
关键词
mailed dosimetry; external audits; electron beams;
D O I
10.1016/S0167-8140(00)00265-6
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Aim: A feasibility study has been performed to investigate the possibility of using mailed thermoluminescence dosimetry (TLD) for external audits of clinical electron beams in Europe. Methods: In the frame of the EC Network Project for Quality Assurance in Radiotherapy, instruction sheets and mailing procedures have been defined for mailed TLD dosimetry using the dedicated holder developed by a panel of experts of the International Atomic Energy Agency (IAEA). Three hundred and thirty electron beam set-ups have been checked in the reference centres and some local centres of the EC Network Project and in addition through the centres participating to the EORTC Radiotherapy Group trial 22922. Results: The mean ratio of measured dose to stated dose is 0.2% and the standard deviation of measured dose to stated dose is 3.2%. In seven beam set-ups, deviations greater than 10% were observed (max. 66%), showing the usefulness of these checks. Conclusion: The results of this feasibility study (instruction sheets, mailing procedures, holder) are presently endorsed by the EQUAL-ESTRO structure in order to offer in the future to all ESTRO members the possibility to request external audits of clinical electron beams. (C) 2001 Elsevier Science Ireland Ltd. All rights reserved.
引用
收藏
页码:163 / 168
页数:6
相关论文
共 19 条
[1]  
BARTHE J, 1990, RADIAT PROT DOSIM, V34, P261
[2]   A EUROPEAN QUALITY ASSURANCE NETWORK FOR RADIOTHERAPY - DOSE MEASUREMENT PROCEDURE [J].
DERREUMAUX, S ;
CHAVAUDRA, J ;
BRIDIER, A ;
ROSSETTI, V ;
DUTREIX, A .
PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (07) :1191-1208
[3]   QUALITY-CONTROL OF RADIOTHERAPY CENTERS IN EUROPE - BEAM CALIBRATION [J].
DUTREIX, A ;
DERREUMAUX, S ;
CHAVAUDRA, J ;
VANDERSCHUEREN, E .
RADIOTHERAPY AND ONCOLOGY, 1994, 32 (03) :256-264
[4]  
DUTREIX A, 1994, TECDOC734 IAEA, P305
[5]   The ESTRO-QUALity assurance network (EQUAL) [J].
Ferreira, IH ;
Dutreix, A ;
Bridier, A ;
Chavaudra, J ;
Svensson, H .
RADIOTHERAPY AND ONCOLOGY, 2000, 55 (03) :273-284
[6]  
Hansen O., 1991, Comments on Nuclear and Particle Physics, V20, P1
[7]  
*INT AT EN AG, 1994, IAEA TECHN REP SER, V381
[8]  
*INT AT EN AG, 1996, IAEA TECHN REP SER, V277
[9]  
*INT AT EN AG, 1987, IAEA TECHN REP SER, V277
[10]   QUALITY ASSURANCE NETWORK IN CENTRAL-EUROPE - EXTERNAL AUDIT ON OUTPUT CALIBRATION FOR PHOTON BEAMS [J].
IZEWSKA, J ;
NOVOTNY, J ;
GWIAZDOWSKA, B ;
KINDLOVA, A ;
KONTRA, G ;
VANDAM, J ;
DUTREIX, A ;
VANDERSCHUEREN, E .
ACTA ONCOLOGICA, 1995, 34 (06) :829-838