CT angiography of various superficial femoral artery stents: An in vitro phantom study

被引:10
作者
Kaempf, Michael [1 ]
Ketelsen, Dominik [1 ]
Syha, Roland [1 ]
Sixt, Sebastian [2 ]
Mangold, Stefanie [1 ]
Thomas, Christoph [1 ]
Claussen, Claus D. [1 ]
Heuschmid, Martin [1 ]
Brechtel, Klaus [1 ]
机构
[1] Univ Tubingen, Dept Diagnost & Intervent Radiol, D-72076 Tubingen, Germany
[2] Heart Ctr Bad Krozingen, Dept Angiol, Bad Krozingen, Germany
关键词
CT angiography; Stent lumen visualization; Kernel; Slice thickness; Nitinol stents; MULTIDETECTOR CORONARY CT; LOWER-EXTREMITY; NITINOL STENTS; DISEASE; ANGIOPLASTY; PROLIFERATION; MANAGEMENT;
D O I
10.1016/j.ejrad.2011.04.014
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To evaluate CT reconstruction parameters to improve stent lumen visualization in vitro. Material and methods: 12 latest superficial femoral artery (SFA) stents were placed in a vessel phantom (diameter 4.7 mm, intravascular attenuation 250 HU, extravascular density 50 HU). Stents were imaged with a 128-slice scanner (SOMATOM Definition Flash, Siemens, Germany) with standard parameters: 120 kV, 200 mAs, collimation 128 mm x 0.6 mm. Different reconstruction parameters were evaluated: B26f, B30f, B45f, B46f and B60f kernel; slice thickness of 0.6, 2.0 and 5.0 mm. To measure visualization characteristics, stent lumen diameter and intraluminal attenuation were assessed. Results: Best stent lumen visualization could be obtained using the B46f kernel (p < 0.001). The visible stent lumen ranged from 66.4% to 83.3% with a mean diameter of 77.7 +/- 4.6%. Nitinol stents showed a significant improved lumen visibility compared to the cobalt-chromium stent (p = 0.02). The most realistic lumen attenuation was achieved using the B46f kernel with a mean attenuation of 259.3 +/- 8.9 HU. The visible lumen diameter in protocols with 5 mm slice thickness was significantly lower (70.0 +/- 4.9%) compared to thinner slices (p < 0.001). Conclusion: CTA of SFA stents should be reconstructed with a slice thickness of 2.0 mm and a B46f kernel to achieve best image quality and to become more sensitive to exclude instent restenosis. (C) 2011 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:1584 / 1588
页数:5
相关论文
共 50 条
  • [41] Carotid artery stents on CT angiography: In vitro comparison of different stent designs and sizes using 16-, 64-and 320-row CT scanners
    Lettau, Michael
    Kotter, Elmar
    Bendszus, Martin
    Haehnel, Stefan
    [J]. JOURNAL OF NEURORADIOLOGY, 2014, 41 (04) : 259 - 268
  • [42] 64-slice multidetector coronary CT angiography: in vitro evaluation of 68 different stents
    David Maintz
    Harald Seifarth
    Rainer Raupach
    Thomas Flohr
    Michael Rink
    Torsten Sommer
    Murat Özgün
    Walter Heindel
    Roman Fischbach
    [J]. European Radiology, 2006, 16 : 818 - 826
  • [43] 64-slice multidetector coronary CT angiography:: in vitro evaluation of 68 different stents
    Maintz, D
    Seifarth, H
    Raupach, R
    Flohr, T
    Rink, M
    Sommer, T
    Özgün, M
    Heindel, W
    Fischbach, R
    [J]. EUROPEAN RADIOLOGY, 2006, 16 (04) : 818 - 826
  • [44] Dual-source CT Angiography in Aortic Stent Grafting: An in vitro Aorta Phantom Study of Image Noise and Radiation Dose
    Sun, Zhonghua
    Ng, Curtise
    [J]. ACADEMIC RADIOLOGY, 2010, 17 (07) : 884 - 893
  • [45] Carotid Artery Stents: In Vitro Comparison of Different Stent Designs and Sizes Using CT Angiography and Contrast-Enhanced MR Angiography at 15T and 3T
    Lettau, M.
    Sauer, A.
    Heiland, S.
    Rohde, S.
    Bendszus, M.
    Haehnel, S.
    [J]. AMERICAN JOURNAL OF NEURORADIOLOGY, 2009, 30 (10) : 1993 - 1997
  • [46] Successful carbon dioxide angiography guided endovascular thrombectomy of the superficial femoral artery in a young patient with critical limb ischemia
    Giusca, Sorin
    Eisele, Tom
    Raupp, Dorothea
    Eisenbach, Christoph
    Korosoglou, Grigorios
    [J]. HEART AND VESSELS, 2017, 32 (03) : 359 - 363
  • [47] Treatment of In-Stent Restenosis in the Superficial Femoral Artery: At the Crossroads of Using Either Covered Stents or Paclitaxel-Based Modalities
    Truijers, Maarten
    Reijnen, Michel M. P. J.
    [J]. JOURNAL OF ENDOVASCULAR THERAPY, 2015, 22 (01) : 11 - 13
  • [48] Performance of ePTFE-covered endograft in patients with occlusive disease of the superficial femoral artery: a three-year clinical follow-up study
    Fritschy, W. M.
    Kruse, R. R.
    Frakking, T. G.
    van Geloven, A. A. W.
    Blomme, A. M.
    [J]. JOURNAL OF CARDIOVASCULAR SURGERY, 2010, 51 (06) : 783 - 790
  • [49] Retrograde popliteal and pedal access in management of superficial femoral artery occlusion after failed antegrade approach: a comparative study
    Ismail, Osama
    [J]. EGYPTIAN JOURNAL OF SURGERY, 2020, 39 (02) : 466 - 475
  • [50] Incidental Intracranial Aneurysms in Patients With Internal Carotid Artery Stenosis A CT Angiography Study and a Metaanalysis
    Heman, Laura M.
    Jongen, Lisa M.
    van der Worp, H. Bart
    Rinkel, Gabriel J. E.
    Hendrikse, Jeroen
    [J]. STROKE, 2009, 40 (04) : 1341 - 1346