Contact atomic force microscopy using piezoresistive cantilevers in load force modulation mode

被引:7
|
作者
Biczysko, P. [1 ]
Dzierka, A. [1 ]
Jozwiak, G. [1 ]
Rudek, M. [1 ]
Gotszalk, T. [1 ]
Janus, P. [2 ]
Grabiec, P. [2 ]
Rangelow, I. W. [3 ]
机构
[1] Wroclaw Univ Technol, Fac Microsyst Elect & Photon, Wroclaw, Poland
[2] Inst Electr Mat Technol, Warsaw, Poland
[3] Ilmenau Univ Technol, Inst Micro & Nanoelect, Dept Micro & Nanoelectr Syst, Fac Elect Engn & Informat Technol, Gustav Kirchhoff Str 1, D-98693 Ilmenau, Germany
关键词
Atomic force microscopy; Piezoresistive probe; Drift free scanning; Load force modulation; PROBE; CALIBRATION; RESOLUTION; ACTUATOR;
D O I
10.1016/j.ultramic.2017.09.002
中图分类号
TH742 [显微镜];
学科分类号
摘要
Scanning probe microscopy (SPM) encompasses several techniques for imaging of the physical and chemical material properties at nanoscale. The scanning process is based on the detection of the deflection of the cantilever, which is caused by near field interactions, while the tip runs over the sample's surface. The variety of deflection detection methods including optical, piezoresistive, piezoelectric technologies has been developed and applied depending on the measurement mode and measurement environment. There are many advantages (compactness, vacuum compatibility, etc.) of the piezoresistive detection method, which makes it very attractive for almost all SPM experiments. Due to the technological limitations the stiffness of the piezoresistive beams is usually higher than the stiffness of the cantilever detected using optical methods. This is the basic constraint for the application of the piezoresistive beams in contact mode (CM) atomic force microscopy (AFM) investigations performed at low load forces (usually less than 20 nN). Drift of the deflection signal, which is related to thermal fluctuations of the measurement setup, causes that the microscope controller compensates the fluctuations instead of compensating the strength of tip-surface interactions. Therefore, it is quite difficult to keep near field interaction precisely at the setpoint level during the whole scanning process. This can lead to either damage of the cantilever's tip and material surface or loosing the contact with the investigated sample and making the measurement unreliable. For these reasons, load force modulation (LoFM) scanning mode, in which the interaction at the tip is precisely controlled at every point of the sample surface, is proposed to enable precise AFM surface investigations using the piezoresistive cantilevers. In this article we describe the developed measurement algorithm as well as proposed and introduced hardware and software solutions. The results of the experiments confirm strong reduction of the AFM entire setup drift. The results demonstrating contactless tip lateral movements are presented. It is common knowledge that such a scanning reduces tip wear. (C) 2017 Elsevier B.V. All rights reserved.
引用
收藏
页码:199 / 208
页数:10
相关论文
共 50 条
  • [41] Contact atomic force microscopy of biological tissues
    Dedkov, V. G.
    Dedkova, E. G.
    TECHNICAL PHYSICS LETTERS, 2010, 36 (02) : 130 - 132
  • [42] Bimodal frequency-modulated atomic force microscopy with small cantilevers
    Dietz, Christian
    Schulze, Marcus
    Voss, Agnieszka
    Riesch, Christian
    Stark, Robert W.
    NANOSCALE, 2015, 7 (05) : 1849 - 1856
  • [43] Automated setpoint adjustment for biological contact mode atomic force microscopy imaging
    Casuso, Ignacio
    Scheuring, Simon
    NANOTECHNOLOGY, 2010, 21 (03)
  • [44] Modelling and nanoscale force spectroscopy of frequency modulation atomic force microscopy
    Payam, Amir Farokh
    APPLIED MATHEMATICAL MODELLING, 2020, 79 (79) : 544 - 554
  • [45] CTE-matched Cantilevers for Improved Heated Atomic Force Microscopy and Passive-mode Scanning Thermal Microscopy
    Sharar, Darin J.
    Wilson, Adam A.
    PROCEEDINGS OF THE 17TH IEEE INTERSOCIETY CONFERENCE ON THERMAL AND THERMOMECHANICAL PHENOMENA IN ELECTRONIC SYSTEMS (ITHERM 2018), 2018, : 191 - 195
  • [46] Fourier transformed atomic force microscopy: tapping mode atomic force microscopy beyond the Hookian approximation
    Stark, RW
    Heckl, WM
    SURFACE SCIENCE, 2000, 457 (1-2) : 219 - 228
  • [47] Functionalization of atomic force microscopy cantilevers and tips by activated vapour silanization
    Daza, Rafael
    Colchero, Luis
    Corregidor, Daniel
    Elices, Manuel
    Guinea, Gustavo, V
    Javier Rojo, Francisco
    Perez-Rigueiro, Jose
    APPLIED SURFACE SCIENCE, 2019, 484 : 1141 - 1148
  • [48] Self-sensing atomic force microscopy cantilevers based on tunnel magnetoresistance sensors
    Tavassolizadeh, Ali
    Meier, Tobias
    Rott, Karsten
    Reiss, Guenter
    Quandt, Eckhard
    Hoelscher, Hendrik
    Meyners, Dirk
    APPLIED PHYSICS LETTERS, 2013, 102 (15)
  • [49] Monitoring of glass derivatization with pulsed force mode atomic force microscopy
    Ebner, A
    Kienberger, F
    Stroh, CM
    Gruber, HJ
    Hinterdorfer, P
    MICROSCOPY RESEARCH AND TECHNIQUE, 2004, 65 (4-5) : 246 - 251
  • [50] Thin film piezoelectric response characterisation using atomic force microscopy with standard contact mode imaging
    Sriram, S.
    Bhaskaran, M.
    Short, K. T.
    Matthews, G. I.
    Holland, A. S.
    MICRON, 2009, 40 (01) : 109 - 113