Theoretical and experimental sensitivity to X-rays of single and polycrystalline HgI2 compared with different single-crystal detectors

被引:44
作者
Schieber, M [1 ]
Hermon, H
Zuck, A
Vilensky, A
Melekhov, L
Shatunovsky, R
Meerson, E
Saado, H
机构
[1] Hebrew Univ Jerusalem, Grad Sch Appl Sci, IL-91904 Jerusalem, Israel
[2] Real Time Radiog Readout, IL-91487 Jerusalem, Israel
关键词
imaging; X-ray; polycrystalline; mercuric iodide; current mode detector;
D O I
10.1016/S0168-9002(00)00849-4
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The response for medical X-ray generator of single and polycrystalline HgI2 to 65 kVp X-rays has been compared to the response of single-crystal CdZnTe and PbI2 detectors using the current integration mode. The dark current density expressed in pA/mm(2) and sensitivity expressed in muC/R x cm(2) are given for these detectors for several thicknesses. The polycrystalline HgI2 thick film detectors were fabricated using the hot wall physical vapor-phase deposition (PVD) method. (C) 2001 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:41 / 46
页数:6
相关论文
共 18 条
[1]  
HENDEE WR, 1984, MED RAD PHYSICS
[2]  
*NAT I STAND TECH, NAT I STAND TECH USA
[3]   CRYSTAL-GROWTH AND APPLICATIONS OF MERCURIC IODIDE [J].
SCHIEBER, M ;
ROTH, M ;
SCHNEPPLE, WF .
JOURNAL OF CRYSTAL GROWTH, 1983, 65 (1-3) :353-364
[4]   Radiation-hard polycrystalline mercuric iodide semiconductor particle counters [J].
Schieber, M ;
Zuck, A ;
Melekhov, L ;
Nissenbaum, J ;
Turchetta, R ;
Dulinski, W ;
Husson, D ;
Riester, JL .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1998, 410 (01) :107-110
[5]   Radiological X-ray response of polycrystalline mercuric iodide detectors [J].
Schieber, M ;
Hermon, H ;
Street, RA ;
Ready, SE ;
Zuck, A ;
Vilensky, A ;
Melekhov, L ;
Shatunovsky, R ;
Meerson, E ;
Saado, Y .
MEDICAL IMAGING 2000: PHYSICS OF MEDICAL IMAGING, 2000, 3977 :48-55
[6]   Towards imaging with polycrystalline mercuric iodide semiconductor detectors [J].
Schieber, M ;
Zuck, A ;
Braiman, M ;
Melekhov, L ;
Nissenbaum, J ;
Turchetta, R ;
Dulinski, W ;
Husson, D ;
Riester, JL ;
Schlesinger, TE ;
Toney, J ;
Sanguinetti, S ;
Montalti, M ;
Guzzi, M .
SEMICONDUCTORS FOR ROOM-TEMPERATURE RADIATION DETECTOR APPLICATIONS II, 1997, 487 :329-337
[7]   Evaluation of mercuric iodide ceramic semiconductor detectors [J].
Schieber, M ;
Zuck, A ;
Braiman, M ;
Nissenbaum, J ;
Turchetta, R ;
Dulinski, W ;
Husson, D ;
Riester, JL .
NUCLEAR PHYSICS B, 1998, :321-329
[8]   High flux X-ray response of composite mercuric iodide detectors [J].
Schieber, M ;
Zuck, A ;
Melekhov, L ;
Shatunovsky, R ;
Hermon, H ;
Turchetta, R .
HARD X-RAY, GAMMA-RAY, AND NEUTRON DETECTOR PHYSICS, 1999, 3768 :296-309
[9]   Novel mercuric iodide polycrystalline nuclear particle counters [J].
Schieber, M ;
Zuck, A ;
Braiman, M ;
Nissenbaum, J ;
Turchetta, R ;
Dulinski, W ;
Husson, D ;
Riester, JL .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1997, 44 (06) :2571-2575
[10]   Polycrystalline mercuric iodide detectors [J].
Schieber, M ;
Zuck, A ;
Braiman, M ;
Melekhov, L ;
Nissenbaum, J ;
Turchetta, R ;
Dulinski, W ;
Husson, D ;
Riester, JL .
HARD X-RAY AND GAMMA-RAY DETECTOR PHYSICS, OPTICS, AND APPLICATIONS, 1997, 3115 :183-187