Optimization of temporal sampling for 82rubidium PET myocardial blood flow quantification

被引:20
作者
Lee, Benjamin C. [1 ]
Moody, Jonathan B. [1 ]
Weinberg, Richard L. [2 ]
Corbett, James R. [1 ,3 ]
Ficaro, Edward P. [1 ,3 ]
Murthy, Venkatesh L. [3 ]
机构
[1] INVIA Med Imaging Solut, 3025 Boardwalk St,Suite 200, Ann Arbor, MI 48108 USA
[2] Univ Michigan, Dept Internal Med, Div Cardiovasc Med, Ann Arbor, MI 48109 USA
[3] Univ Michigan, Dept Radiol, Div Nucl Med, Ann Arbor, MI 48109 USA
关键词
Myocardial perfusion imaging: PET; myocardial blood flow; myocardial flow reserve; Rubidium-82; temporal sampling; POSITRON-EMISSION-TOMOGRAPHY; CORONARY-ARTERY-DISEASE; DYNAMIC PET; RESERVE; PRECISION;
D O I
10.1007/s12350-017-0899-7
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background. Suboptimal temporal sampling of left ventricular (LV) blood pool and tissue time-activity curves (TACs) may introduce bias and increased variability in estimates of myocardial blood flow (MBF) and flow reserve (MFR) from dynamic PET myocardial perfusion images. We aimed to optimize temporal sampling for estimation of MBF and MFR. Methods. Twenty-four normal volunteers and 32 patients underwent dynamic stress/rest rubidium-82 chloride (Rb-82) PET imaging. Fine temporal sampling was used to estimate the full width at half maximum (FWHM) of the LV blood pool TAC. Fourier analysis was used to determine the longest sampling interval, T-S, as a function of FWHM, which preserved the information content of the blood phase. Dynamic datasets were reconstructed with frame durations varying from 2 to 20 seconds over the first 2 minutes for the blood phase and 30 to 120 seconds for the tissue phase. The LV blood pool and tissue TACs were sampled using regions of interest (ROI) and fit to a compartment model for quantification of MBF and MFR. The effects of temporal sampling on MBF and MFR were evaluated using clinical data and simulations. Results. T-S increased linearly with input function FWHM (R = 0.93). Increasing the blood phase frame duration from 5 to 15 seconds resulted in MBF and MFR biases of 6-12% and increased variability of 14-24%. Frame durations <5 seconds had biases of less than 5% for both MBF and MFR values. Increasing the tissue phase frame durations from 30 to 120 seconds resulted in <5% biases. Conclusions. A two-phase framing of dynamic Rb-82 PET images with frame durations of 5 seconds (blood phase) and 120 seconds (tissue phase) optimally samples the blood pool TAC for modern 3D PET systems.
引用
收藏
页码:1517 / 1529
页数:13
相关论文
共 14 条
[1]   Short-term repeatability of resting myocardial blood flow measurements using rubidium-82 PET imaging [J].
Efseaff, Matthew ;
Klein, Ran ;
Ziadi, Maria C. ;
Beanlands, Rob S. ;
deKemp, Robert A. .
JOURNAL OF NUCLEAR CARDIOLOGY, 2012, 19 (05) :997-1006
[2]   QUANTITATION OF MYOCARDIAL BLOOD-FLOW WITH (H2O)-O-15 AND POSITRON EMISSION TOMOGRAPHY - ASSESSMENT AND ERROR ANALYSIS OF A MATHEMATICAL APPROACH [J].
HERRERO, P ;
MARKHAM, J ;
BERGMANN, SR .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1989, 13 (05) :862-873
[3]   Precision-controlled elution of a 82Sr/82Rb generator for cardiac perfusion imaging with positron emission tomography [J].
Klein, R. ;
Adler, A. ;
Beanlands, R. S. ;
deKemp, R. A. .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (03) :659-673
[4]   Quantification of myocardial blood flow and flow reserve: Technical aspects [J].
Klein, Ran ;
Beanlands, Rob S. B. ;
deKemp, Robert A. .
JOURNAL OF NUCLEAR CARDIOLOGY, 2010, 17 (04) :555-570
[5]   Optimized dynamic framing for PET-based myocardial blood flow estimation [J].
Kolthammer, Jeffrey A. ;
Muzic, Raymond F. .
PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (16) :5783-5801
[6]   Quantification of myocardial blood flow with 82Rb dynamic PET imaging [J].
Lortie, Mireille ;
Beanlands, Rob S. B. ;
Yoshinaga, Keiichiro ;
Klein, Ran ;
DaSilva, Jean N. ;
dekemp, Robert A. .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2007, 34 (11) :1765-1774
[7]   DYNAMIC PET DATA-ANALYSIS [J].
MAZOYER, BM ;
HUESMAN, RH ;
BUDINGER, TF ;
KNITTEL, BL .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1986, 10 (04) :645-653
[8]   Precision and accuracy of clinical quantification of myocardial blood flow by dynamic PET: A technical perspective [J].
Moody, Jonathan B. ;
Lee, Benjamin C. ;
Corbett, James R. ;
Ficaro, Edward P. ;
Murthy, Venkatesh L. .
JOURNAL OF NUCLEAR CARDIOLOGY, 2015, 22 (05) :935-951
[9]   Improved Cardiac Risk Assessment With Noninvasive Measures of Coronary Flow Reserve [J].
Murthy, Venkatesh L. ;
Naya, Masanao ;
Foster, Courtney R. ;
Hainer, Jon ;
Gaber, Mariya ;
Di Carli, Gilda ;
Blankstein, Ron ;
Dorbala, Sharmila ;
Sitek, Arkadiusz ;
Pencina, Michael J. ;
Di Carli, Marcelo F. .
CIRCULATION, 2011, 124 (20) :2215-2224
[10]   Preserved Coronary Flow Reserve Effectively Excludes High-Risk Coronary Artery Disease on Angiography [J].
Naya, Masanao ;
Murthy, Venkatesh L. ;
Taqueti, Viviany R. ;
Foster, Courtney R. ;
Klein, Josh ;
Garber, Mariya ;
Dorbala, Sharmila ;
Hainer, Jon ;
Blankstein, Ron ;
Resnic, Frederick ;
Di Carli, Marcelo F. .
JOURNAL OF NUCLEAR MEDICINE, 2014, 55 (02) :248-255