New Intramedullary Locking Nail for Olecranon Fracture Fixation-An In Vitro Biomechanical Comparison With Tension Band Wiring

被引:42
作者
Nowak, Tobias E. [1 ]
Burkhart, Klaus J. [1 ]
Mueller, Lars P. [1 ]
Mattyasovszky, Stefan G. [1 ]
Andres, Torsten [1 ]
Sternstein, Werner [1 ]
Rommens, Pol M. [1 ]
机构
[1] Johannes Gutenberg Univ Mainz, Dept Trauma Surg & Orthopaed Surg, D-55131 Mainz, Germany
来源
JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE | 2010年 / 69卷 / 05期
关键词
Olecranon fracture; Osteotomy; Intramedullary nailing; Biomechanical analysis; Tension band wiring; ADULTS; ULNA;
D O I
10.1097/TA.0b013e3181c9af9b
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Background: The aim of this study was to determine the difference in displacement of a newly designed intramedullary olecranon fracture fixation device compared with multifilament tension band wiring after 4 cycles and 300 cycles of dynamic continuous loading. Methods: In eight pairs of fresh-frozen cadaver ulnae, oblique olecranon fractures were created and stabilized using either newly designed intramedullary olecranon nail or multifilament tension band wiring. The specimens were then subjected to continuous dynamic loading (from 25 N to 200 N) using matched pairs of cadaveric upper extremities. The Wilcoxon test was used to determine statistical differences of the displacement in the fracture gap. Results: After 4 cycles and 300 cycles, the displacement in the fracture model was significantly higher in the tension band wiring group than in the intramedullary nailing group. Conclusions: The newly designed interlocking nailing system showed higher stability in comparison with multifilament tension band wiring after continuous dynamic loading.
引用
收藏
页码:E56 / E61
页数:6
相关论文
共 21 条
[1]   Morphologic evaluation of the ulna [J].
Akpinar, F ;
Aydinlioglu, A ;
Tosun, N ;
Tuncay, I .
ACTA ORTHOPAEDICA SCANDINAVICA, 2003, 74 (04) :415-419
[2]  
AN KN, 2000, ELBOW ITS DISORDERS, P43
[3]  
Davenport S R, 1988, J Orthop Trauma, V2, P146
[4]   The olecranon sled - a new device for fixation of fractures of the olecranon - A mechanical comparison of two fixation methods in cadaver elbows [J].
Dieterich, Jan ;
Kummer, Frederick J. ;
Ceder, Leif .
ACTA ORTHOPAEDICA, 2006, 77 (03) :440-444
[5]   METHODS OF FIXATION OF OLECRANON FRACTURES - AN EXPERIMENTAL MECHANICAL STUDY [J].
FYFE, IS ;
MOSSAD, MM ;
HOLDSWORTH, BJ .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1985, 67 (03) :367-372
[6]  
Hak D J, 2000, J Am Acad Orthop Surg, V8, P266
[7]   ELBOW FUNCTION FOLLOWING TENSION BAND FIXATION OF DISPLACED FRACTURES OF THE OLECRANON [J].
HOLDSWORTH, BJ ;
MOSSAD, MM .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 1984, 16 (03) :182-187
[8]  
JUPITER JB, 1992, ORTHOPEDICS, V15, P825
[9]  
KNIGHT R A, 1957, Instr Course Lect, V14, P123
[10]   Distal femoral fixation: A biomechanical comparison of the standard condylar buttress plate, a locked buttress plate, and the 95-degree blade plate [J].
Koval, KJ ;
Hoehl, JJ ;
Kummer, FJ ;
Simon, JA .
JOURNAL OF ORTHOPAEDIC TRAUMA, 1997, 11 (07) :521-524