Assessment of Damage on a Dual Mobility Acetabular System

被引:50
作者
D'Apuzzo, Michele R. [1 ]
Koch, Chelsea N. [2 ]
Esposito, Christina I. [2 ]
Elpers, Marcella E. [2 ]
Wright, Timothy M. [2 ]
Westrich, Geoffrey H. [1 ]
机构
[1] Hosp Special Surg, Adult Reconstruct & Joint Replacement Div, Dept Orthoped Surg, 535 E 70th St, New York, NY 10021 USA
[2] Hosp Special Surg, Dept Biomech, 535 East 70th St, New York, NY 10021 USA
关键词
total hip arthroplasty; retrieval analysis; bearing surfaces; polyethylene wear; dual mobility; TOTAL HIP-ARTHROPLASTY; INTRAPROSTHETIC DISLOCATION; RETRIEVAL ANALYSIS; FOLLOW-UP; WEAR; CUPS; REVISION; LINERS; THA;
D O I
10.1016/j.arth.2016.01.039
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Dual mobility designs were introduced to increase stability and reduce the risk of dislocation, both being common reasons for surgical revision after total hip arthroplasty. The in vivo behavior of dual mobility constructs remains unclear, and to our knowledge, no data have been published describing in vivo surface damage to the polyethylene bearing surfaces. Methods: We used surface damage assessed on the inner and outer polyethylene bearing surfaces in 33 short-term retrieved dual mobility liners as evidence of relative motion at the 2 bearings. A lever out test was performed to determine the force required for dislocation of the cobalt-chromium femoral head from the polyethylene liner. Results: Both bearings showed damage; however, the inner polyethylene bearings had higher damage scores, lower prevalence of remaining machining marks, and higher incidence of concentric wear, all consistent with more motion at the inner polyethylene bearing. The inner polyethylene bearings also had a higher occurrence of embedded titanium debris. The damage sustained in vivo was insufficient to lead to intraprosthetic dislocation in any of the retrieved components. Lever out tests of 12 retrievals had a mean dislocation load of 261 +/- 52 N, which was unrelated to the length of implantation. Conclusion: Our short-term retrieval data of 33 highly cross-linked polyethylene dual mobility components suggest that although motion occurs at both bearing articulations, the motion of the femoral head against the inner polyethylene bearing dominates. Although damage was not severe enough to lead to intraprosthetic dislocation, failure may occur long term and should be assessed in future studies. (C) 2016 Elsevier Inc. All rights reserved.
引用
收藏
页码:1828 / 1835
页数:8
相关论文
共 19 条
[1]   Dual mobility retentive acetabular liners and wear: Surface analysis of 40 retrieved polyethylene implants [J].
Adam, P. ;
Farizon, F. ;
Fessy, M. -H. .
ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2014, 100 (01) :85-91
[2]   Early Intraprosthetic Dislocation in a Revision Dual-mobility Hip Prosthesis [J].
Banka, Trevor R. ;
Ast, Michael P. ;
Parks, Michael L. .
ORTHOPEDICS, 2014, 37 (04) :E395-E397
[3]   MR Imaging of Adverse Local Tissue Reactions around Rejuvenate Modular Dual-Taper Stems [J].
Burge, Alissa J. ;
Gold, Stephanie L. ;
Lurie, Brett ;
Nawabi, Danyal H. ;
Fields, Kara G. ;
Koff, Matthew F. ;
Westrich, Geoffrey ;
Potter, Hollis G. .
RADIOLOGY, 2015, 277 (01) :142-150
[4]   Low Rate of Dislocation of Dual-mobility Cups in Primary Total Hip Arthroplasty [J].
Combes, Antoine ;
Migaud, Henri ;
Girard, Julien ;
Duhamel, Alain ;
Fessy, Michel Henri .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2013, 471 (12) :3891-3900
[5]   Results with a cementless alumina-coated cup with dual mobility - A twelve-year follow-up study [J].
Farizon, F ;
de Lavison, R ;
Azoulai, JJ ;
Bousquet, G .
INTERNATIONAL ORTHOPAEDICS, 1998, 22 (04) :219-224
[6]   In vivo and ex vivo measurement of polyethylene wear in total hip arthroplasty [J].
Goldvasser, Dov ;
Hansen, Viktor J. ;
Noz, Marilyn E. ;
Maguire, Gerald Q., Jr. ;
Zeleznik, Michael P. ;
Olivecrona, Henrik ;
Bragdon, Charles R. ;
Weidenhielm, Lars ;
Malchau, Henrik .
ACTA ORTHOPAEDICA, 2014, 85 (03) :271-275
[7]   Biomechanical concept and clinical outcome of dual mobility cups [J].
Grazioli, Aron ;
Ek, Eugene Teow Hin ;
Ruediger, Hannes Andreas .
INTERNATIONAL ORTHOPAEDICS, 2012, 36 (12) :2411-2418
[8]   Dual-mobility cups for revision due to instability are associated with a low rate of re-revisions due to dislocation 228 patients from the Swedish Hip Arthroplasty Register [J].
Hailer, Nils P. ;
Weiss, Rudiger J. ;
Stark, Andre ;
Karrholm, Johan .
ACTA ORTHOPAEDICA, 2012, 83 (06) :566-571
[9]   RETRIEVAL ANALYSIS OF TOTAL KNEE PROSTHESES - A METHOD AND ITS APPLICATION TO 48 TOTAL CONDYLAR PROSTHESES [J].
HOOD, RW ;
WRIGHT, TM ;
BURSTEIN, AH .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1983, 17 (05) :829-842
[10]   Computerized Range of Motion Analysis Following Dual Mobility Total Hip Arthroplasty, Traditional Total Hip Arthroplasty, and Hip Resurfacing [J].
Klingenstein, Gregory G. ;
Yeager, Alyssa M. ;
Lipman, Joseph D. ;
Westrich, Geoffrey H. .
JOURNAL OF ARTHROPLASTY, 2013, 28 (07) :1173-1176