A single-stage membrane-based interface has been developed for real-time mass spectrometric monitoring of the starting materials and products of a highly concentrated pharmaceutical process reaction mixture. The interface was directly connected to the atmospheric pressure chemical ionisation source of a quadrupole mass spectrometer (APCI-MS). Dilution of the concentrated reaction mixture was achieved in a single-step by using the interface with a precision of 3.2% for replicate samples. The combination of the membrane inlet with APCI-MS is demonstrated for the Michael addition reaction of phenylethylamine and acrylonitrile in ethanol using a hydrophobic polyvinylidene fluoride microporous membrane. The reaction was monitored throughout its course, allowing the endpoint to be determined based on the relative concentrations of the reaction precursors and products. The device required minimal analyst intervention, reducing sample preparation and handling prior to on-line real-time MS analysis. (c) 2005 Elsevier B.V. All rights reserved.
引用
收藏
页码:133 / 140
页数:8
相关论文
共 30 条
[1]  
BEEBE KR, 1993, ANAL CHEM, V65, pR199
[2]   PROCESS ANALYTICAL-CHEMISTRY [J].
BLASER, WW ;
BREDEWEG, RA ;
HARNER, RS ;
LAPACK, MA ;
LEUGERS, A ;
MARTIN, DP ;
PELL, RJ ;
WORKMAN, J ;
WRIGHT, LG .
ANALYTICAL CHEMISTRY, 1995, 67 (12) :R47-R70
[3]  
Bohatka S, 1997, RAPID COMMUN MASS SP, V11, P656, DOI 10.1002/(SICI)1097-0231(199704)11:6<656::AID-RCM835>3.0.CO
[4]  
2-W
[5]   Monitoring organic reactions with on-line atmospheric pressure ionization mass spectrometry: the hydrolysis of isatin [J].
Brum, J ;
Dell'Orco, P ;
Lapka, S ;
Muske, K ;
Sisko, J .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2001, 15 (17) :1548-1553
[6]   On-line biodegradation monitoring of nitrogen-containing compounds by membrane inlet mass spectrometry [J].
Creaser, CS ;
Lamarca, DG ;
dos Santos, LMF ;
LoBiundo, G ;
New, AP .
JOURNAL OF CHEMICAL TECHNOLOGY AND BIOTECHNOLOGY, 2003, 78 (11) :1193-1200
[7]   Reversed-phase membrane inlet mass spectrometry applied to the real-time monitoring of low molecular weight alcohols in chloroform [J].
Creaser, CS ;
Lamarca, DG ;
Brum, J ;
Werner, C ;
New, AP ;
dos Santos, LMF .
ANALYTICAL CHEMISTRY, 2002, 74 (01) :300-304
[8]  
Creaser CS, 1998, ANAL COMMUN, V35, p9H
[9]   Monitoring process-scale reactions using API mass spectrometry [J].
Dell'Orco, P ;
Brum, J ;
Matsuoka, R ;
Badlani, M ;
Muske, K .
ANALYTICAL CHEMISTRY, 1999, 71 (22) :5165-5170
[10]   Real-time mass spectrometry [J].
Hoffmann, T .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2003, 375 (01) :36-37