Quantitative analysis of ablation technique predicts arrhythmia recurrence following atrial fibrillation ablation

被引:9
作者
Jankelson, Lior [1 ]
Dai, Matthew [1 ]
Bernstein, Scott [1 ]
Park, David [1 ]
Holmes, Douglas [1 ]
Aizer, Anthony [1 ]
Chinitz, Larry [1 ]
Barbhaiya, Chirag [1 ]
机构
[1] NYU, Sch Med, Leon H Charney Div Cardiol, Langone Med Ctr, New York, NY 10016 USA
关键词
PULMONARY VEIN ISOLATION; CONTACT FORCE; CATHETER ABLATION; RADIOFREQUENCY ABLATION; LESION SIZE; IMPEDANCE; IMPACT;
D O I
10.1016/j.ahj.2019.11.011
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Optimal ablation technique, including catheter-tissue contact during atrial fibrillation (AF) radiofrequency (RF) ablation, is associated with improved procedural outcomes. We used a custom developed software to analyze high-frequency catheter position data to study the interaction between catheter excursion during lesion placement, lesion-set sequentiality, and arrhythmia recurrence. Methods A total of 100 consecutive patients undergoing first-time RF ablation for paroxysmal AF were analyzed. Spatial positioning of the ablation catheter sampled at 60 Hz during RF application was extracted from the CARTO3 system (Biosense Webster Inc, USA) and analyzed using custom-developed MATLAB software to determine precise catheter spatial 3D excursion during RF ablation. The primary end point was freedom from atrial arrhythmia lasting longer than 30 seconds after a single ablation procedure. Results At 1 year, 86% of patients were free from recurrent arrhythmia. There was no significant difference in clinical, echocardiographic, or ablation characteristics between patients with and without recurrent arrhythmia. Analyzing 15,356,998 position data points revealed that lesion-set sequentiality and mean lesion catheter excursion were predictors of arrhythmia recurrence. Analyzing arrhythmia recurrence by mean single-lesion catheter excursion (excursion >2.81 mm) and by sequentiality (using 46% of lesions with interlesion distance >6 mm as cutoff) revealed significantly increased arrhythmia recurrence in the higher excursion group (23% vs 6%, P = .03) and in the less sequential group (24% vs 4%, P = .02). Conclusions Ablation lesion sequentiality measured by catheter interlesion distance and catheter stability measured by catheter excursion during lesion placement are potentially modifiable factors affecting arrhythmia recurrence after RF ablation for AF.
引用
收藏
页码:176 / 183
页数:8
相关论文
共 23 条
[1]  
CALKINS H, 2017, J ARRHYTHMIA, V2017, P369
[2]   Sites With Small Impedance Decrease During Catheter Ablation for Atrial Fibrillation Are Associated With Recovery of Pulmonary Vein Conduction [J].
Chinitz, Jason S. ;
Kapur, Sunil ;
Barbhaiya, Chirag ;
Kumar, Saurabh ;
John, Roy ;
Epstein, Laurence M. ;
Tedrow, Usha ;
Stevenson, William G. ;
Michaud, Gregory F. .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2016, 27 (12) :1390-1398
[3]   Optimization of catheter/tissue contact during pulmonary vein isolation: the impact of atrial rhythm [J].
De Ponti, Roberto ;
Marazzi, Raffaella ;
Doni, Lorenzo A. ;
Marazzato, Jacopo ;
Baratto, Claudia ;
Salerno-Uriarte, Jorge A. .
EUROPACE, 2018, 20 (02) :288-294
[4]   Remotely controlled steerable sheath improves result and procedural parameters of atrial fibrillation ablation with magnetic navigation [J].
Errahmouni, Abdelkarim ;
Latcu, Decebal Gabriel ;
Bun, Sok-Sithikun ;
Rijo, Nicolas ;
Dugourd, Celine ;
Saoudi, Nadir .
EUROPACE, 2015, 17 (07) :1045-1050
[5]   Electroanatomical mapping of the heart: Basic concepts and implications for the treatment of cardiac arrhythmias [J].
Gepstein, L ;
Evans, SJ .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (06) :1268-1278
[6]   High-frequency jet ventilation: Utility in posterior left atrial catheter ablation [J].
Goode, JS ;
Taylor, RL ;
Buffington, CW ;
Klain, MM ;
Schwartzman, D .
HEART RHYTHM, 2006, 3 (01) :13-19
[7]   OBSERVATIONS ON ELECTRODE-TISSUE INTERFACE TEMPERATURE AND EFFECT ON ELECTRICAL-IMPEDANCE DURING RADIOFREQUENCY ABLATION OF VENTRICULAR MYOCARDIUM [J].
HAINES, DE ;
VEROW, AF .
CIRCULATION, 1990, 82 (03) :1034-1038
[8]   Efforts to enhance catheter stability improve atrial fibrillation ablation outcome [J].
Hutchinson, Mathew D. ;
Garcia, Fermin C. ;
Mandel, Jeff E. ;
Elkassabany, Nabil ;
Zado, Erica S. ;
Riley, Michael P. ;
Cooper, Joshua M. ;
Bala, Rupa ;
Frankel, David S. ;
Lin, David ;
Supple, Gregory E. ;
Dixit, Sanjay ;
Gerstenfeld, Edward P. ;
Callans, David J. ;
Marchlinski, Francis E. .
HEART RHYTHM, 2013, 10 (03) :347-353
[9]  
Kumar Saurabh, 2015, J Atr Fibrillation, V8, P1189, DOI 10.4022/jafib.1189
[10]   Catheter-Tissue Contact Force Determines Atrial Electrogram Characteristics Before and Lesion Efficacy After Antral Pulmonary Vein Isolation in Humans [J].
Kumar, Saurabh ;
Chan, Martin ;
Lee, Justin ;
Wong, Michael C. G. ;
Yudi, Matias ;
Morton, Joseph B. ;
Spence, Steven J. ;
Halloran, Karen ;
Kistler, Peter M. ;
Kalman, Jonathan M. .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2014, 25 (02) :122-129