Preliminary investigation of direct sea-water analysis by inductively coupled plasma mass spectrometry using a mixed-gas plasma, flow injection and external calibration

被引:22
作者
Holliday, AE [1 ]
Beauchemin, D [1 ]
机构
[1] Queens Univ, Dept Chem, Kingston, ON K7L 3N6, Canada
关键词
D O I
10.1039/b303184e
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A method is described that allows the direct analysis of sea-water by inductively coupled plasma mass spectrometry (ICP-MS) using a simple external calibration without internal standardization. A 50 muL aliquot of sample is introduced by flow injection and diluted 2-fold with deionized distilled water on-line to an ICP-MS instrument. A 4% nitrogen - 96% argon plasma, operated at a 2 mm greater sampling depth than optimal for sensitivity, provides robust operating conditions that greatly reduce non-spectroscopic interferences (i.e. matrix effects) while still maintaining sufficient sensitivity for analyte quantitation. This simple approach allowed the accurate determination of Mo in a sea-water certified reference material (NASS-5) using a 150 ms dwell time on a second-generation ICP-MS instrument. The determination of other elements should be possible with more recent and more sensitive ICP-MS instruments. Although this mixed-gas plasma also reduced the ArCl spectroscopic interference on As, the residual interference prevented accurate quantitation of As in sea-water. However, the approach may still be suitable for the determination of As ( and other elements that suffer from spectroscopic interferences arising from the sample matrix) in samples where the troublesome matrix elements are less concentrated.
引用
收藏
页码:1109 / 1112
页数:4
相关论文
共 25 条
[1]   Inductively coupled plasma mass spectrometry [J].
Beauchemin, D .
ANALYTICAL CHEMISTRY, 2002, 74 (12) :2873-2893
[2]   ONLINE STANDARD ADDITION METHOD WITH ICPMS USING FLOW-INJECTION [J].
BEAUCHEMIN, D .
ANALYTICAL CHEMISTRY, 1995, 67 (09) :1553-1557
[3]   STUDY OF THE EFFECTS OF CONCOMITANT ELEMENTS IN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
BEAUCHEMIN, D ;
MCLAREN, JW ;
BERMAN, SS .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1987, 42 (03) :467-490
[4]   On-line isotope dilution analysis with ICPMS using reverse flow injection [J].
Beauchemin, D ;
Specht, AA .
ANALYTICAL CHEMISTRY, 1997, 69 (16) :3183-3187
[5]  
BEAUCHEMIN D, 2000, DISCRETE SAMPLE INTR, P278
[6]  
BEAUCHEMIN D, 2000, DISCRETE SAMPLE INTR, P97
[7]   INDUCTIVELY COUPLED PLASMA AS AN ION-SOURCE FOR HIGH-RESOLUTION MASS-SPECTROMETRY [J].
BRADSHAW, N ;
HALL, EFH ;
SANDERSON, NE .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1989, 4 (08) :801-803
[8]   Computer simulation of argon-nitrogen and argon-oxygen inductively coupled plasmas [J].
Cai, MX ;
Haydar, DA ;
Montaser, A ;
Mostaghimi, J .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1997, 52 (03) :369-386
[9]   INTERFERENCES IN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY - A REVIEW [J].
EVANS, EH ;
GIGLIO, JJ .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1993, 8 (01) :1-18
[10]   OPTIMIZATION STRATEGIES FOR THE REDUCTION OF NONSPECTROSCOPIC INTERFERENCES IN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
EVANS, EH ;
CARUSO, JA .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1992, 47 (08) :1001-1012