VALIDATION OF AN INDEXED RADIOTHERAPY HEAD POSITIONING DEVICE FOR USE IN DOGS AND CATS

被引:20
作者
Hansen, Katherine S. [1 ]
Theon, Alain P. [1 ]
Dieterich, Sonja [2 ]
Kent, Michael S. [1 ]
机构
[1] Univ Calif Davis, Dept Surg & Radiol Sci, VM Surg Radiol, Davis, CA 95616 USA
[2] Univ Calif Davis, Ctr Comprehens Canc, Radiat Oncol, Sacramento, CA USA
基金
欧盟地平线“2020”;
关键词
immobilization; indexed; positioning; radiation; veterinary; PATIENT IMMOBILIZATION DEVICE; BEAM COMPUTED-TOMOGRAPHY; DAILY SETUP VARIATIONS; RADIATION-THERAPY; HELICAL TOMOTHERAPY; CANINE HEAD; ACCURACY; SYSTEM; IMPACT;
D O I
10.1111/vru.12257
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Setup variability affects the appropriate delivery of radiation and informs the setup margin required to treat radiation patients. Twenty-four veterinary patients with head and neck cancers were enrolled in this prospective, cross-sectional study to determine the accuracy of an indexed board immobilization device for positioning. Couch position values were defined at the first treatment based on setup films. At subsequent treatments, patients were moved to the previously defined couch location, orthogonal films were acquired, table position was modified, and displacement was recorded. The mean systematic displacement, random displacement, overall displacement, and mean displacement values of the three-dimensional (3D) vector were calculated. Three hundred thirty-two pairs of orthogonal setup films were analyzed for displacement in cranial-caudal, lateral, and dorsal-ventral directions. The mean systematic displacements were 0.5, 0.8, and 0.5 mm, respectively. The mean random displacements were 1.0, 1.1, and 0.7 mm, respectively. The overall displacements were 1.1, 1.4, and 0.9 mm, respectively. The mean 3D vector value was 1.6 mm with a standard deviation of 1.2 mm. Ninety-five percent of the vectors were <3.6 mm. These values were compared to data obtained with a previously used immobilization device. A t-test was used to compare the two devices. The 3D vector, random displacement in all directions, and overall displacement in the cranial-caudal and dorsal-ventral directions were significantly smaller than displacements with the previous device. The precision and accuracy of the indexed board device was superior to the historical head and neck device.
引用
收藏
页码:448 / 455
页数:8
相关论文
共 30 条
[1]  
[Anonymous], 1999, PRESCRIBING RECORDIN
[2]  
[Anonymous], 2010, J ICRU, DOI DOI 10.1093/JICRUNDQ002
[3]   Comparison of two dead and neck immobilization systems [J].
Bentel, GC ;
Marks, LB ;
Hendren, K ;
Brizel, DM .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 38 (04) :867-873
[4]   Assessment of a radiotherapy patient immobilization device using single plane port radiographs and a remote computed tomography scanner [J].
Bley, CR ;
Blattmann, H ;
Roos, M ;
Sumova, A ;
Kaser-Hotz, B .
VETERINARY RADIOLOGY & ULTRASOUND, 2003, 44 (04) :470-475
[5]   EVALUATION OF A HEAD-REPOSITIONER AND Z-PLATE SYSTEM FOR IMPROVED ACCURACY OF DOSE DELIVERY [J].
Charney, Sarah C. ;
Lutz, Wendell R. ;
Klein, Mary K. ;
Jones, Pamela D. .
VETERINARY RADIOLOGY & ULTRASOUND, 2009, 50 (03) :323-329
[6]   DOSIMETRIC IMPACT OF DAILY SETUP VARIATIONS DURING TREATMENT OF CANINE NASAL TUMORS USING INTENSITY-MODULATED RADIATION THERAPY [J].
Deveau, Michael A. ;
Gutierrez, Alonso N. ;
Mackie, Thomas R. ;
Tome, Wolfgang A. ;
Forrest, Lisa J. .
VETERINARY RADIOLOGY & ULTRASOUND, 2010, 51 (01) :90-96
[7]   A comparative evaluation of two head and neck immobilization devices using electronic portal imaging [J].
Donato, K ;
Leszczynski, K ;
Fleming, K .
BRITISH JOURNAL OF RADIOLOGY, 2006, 79 (938) :158-161
[8]   TIME TREND OF PATIENT SETUP DEVIATIONS DURING PELVIC IRRADIATION USING ELECTRONIC PORTAL IMAGING [J].
ELGAYED, AAH ;
BEL, A ;
VIJLBRIEF, R ;
BARTELINK, H ;
LEBESQUE, JV .
RADIOTHERAPY AND ONCOLOGY, 1993, 26 (02) :162-171
[9]   A SURVEY OF VETERINARY RADIATION FACILITIES IN 2010 [J].
Farrelly, John ;
McEntee, Margaret C. .
VETERINARY RADIOLOGY & ULTRASOUND, 2014, 55 (06) :638-643
[10]   The utility of megavoltage computed tomography images from a helical tomotherapy system for setup verification purposes [J].
Forrest, LJ ;
Mackie, TR ;
Ruchala, K ;
Turek, M ;
Kapatoes, J ;
Jaradat, H ;
Hui, S ;
Balog, J ;
Vail, DM ;
Mehta, MP .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 60 (05) :1639-1644