Increasing bioanalytical throughput using pcSFC-MS/MS: 10 minutes per 96-well plate

被引:43
作者
Hoke, SH
Tomlinson, JA
Bolden, RD
Morand, KL
Pinkston, JD
Wehmeyer, KR
机构
[1] Procter & Gamble Co, Hlth Care Res Ctr, Mason, OH 45040 USA
[2] Procter & Gamble Co, Miami Valley Labs, Cincinnati, OH 45253 USA
关键词
D O I
10.1021/ac0014820
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The utility of packed-column supercritical, subcritical, and enhanced fluidity liquid chromatographies (pcSFC) for high-throughput applications has increased during the past few years. In contrast to traditional reversed-phase liquid chromatography, the addition of a volatile component to the mobile phase, such as CO2, produces a lower mobile-phase viscosity. This allows the use of higher flow rates which can translate into faster analysis times. In addition, the resulting mobile phase is considerably more volatile than the aqueous-based mobile phases that are typically used with LC-MS, allowing the entire effluent to be directed into the MS interface. High-throughput bioanalytical quantitation using pcSFC-MS/MS for pharmacokinetics applications is demonstrated in this report using dextromethorphan as a model compound. Plasma samples were prepared by automated liquid/liquid extraction in the 96-well format prior to pcSFC-MS/MS analysis. Three days of validation data are provided along with study sample data from a patient dosed with commercially available Vicks 44, Using pcSFC and MS/MS, dextromethorphan was quantified in 96-well plates at a rate of similar to 10 min/plate with average intraday accuracy of 9% or better. Daily relative standard deviations (RSDs) were less than 10% for the 2.21 and 14.8 ng/mL quality control (QC) samples, while the RSDs were less than 15% at the 0.554 ng/mL QC level.
引用
收藏
页码:3083 / 3088
页数:6
相关论文
共 25 条
[1]  
ANTON K, 1998, CHROMATOGRAPHIC SCI, V75
[2]  
BAILLIE TA, 1908, MASS SPECTROMETRY BI
[3]   Development and application of packed-column supercritical fluid chromatography pneumatically assisted electrospray mass spectrometry [J].
Baker, TR ;
Pinkston, JD .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1998, 9 (05) :498-509
[4]  
Bayliss MK, 2000, RAPID COMMUN MASS SP, V14, P2039, DOI 10.1002/1097-0231(20001115)14:21<2039::AID-RCM130>3.0.CO
[5]  
2-2
[6]  
Berger T.A., 1995, PACKED COLUMN SFC RS
[7]   Pressure-regulating fluid interface and phase behavior considerations in the coupling of packed-column supercritical fluid chromatography with low-pressure detectors [J].
Chester, TL ;
Pinkston, JD .
JOURNAL OF CHROMATOGRAPHY A, 1998, 807 (02) :265-273
[8]   DETERMINATION OF DEXTROMETHORPHAN AND METABOLITES IN HUMAN-PLASMA AND URINE BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY WITH FLUORESCENCE DETECTION [J].
EAST, T ;
DYE, D .
JOURNAL OF CHROMATOGRAPHY, 1985, 338 (01) :99-112
[9]  
Eichhold TH, 1997, J MASS SPECTROM, V32, P1205
[10]   Use of short high-performance liquid chromatography columns and tandem-mass spectrometry for the rapid analysis of a prostaglandin analog, fluprostenol, in rat plasma [J].
Eichhold, TH ;
Kuhlenbeck, DL ;
Baker, TR ;
Stella, ME ;
Amburgey, JS ;
deLong, MA ;
Hartke, JR ;
Cruze, CA ;
Pierce, SA ;
Wehmeyer, KR .
JOURNAL OF CHROMATOGRAPHY B, 2000, 741 (02) :213-220