Calculation of Radiofrequency Electromagnetic Fields and Their Effects in MRI of Human Subjects

被引:115
作者
Collins, Christopher M. [1 ]
Wang, Zhangwei [2 ]
机构
[1] Penn State Univ, Dept Radiol, Hershey, PA USA
[2] GE Healthcare, Aurora, OH USA
关键词
MRI; electromagnetics; radiofrequency; signal; noise; SAR; safety; SIGNAL-TO-NOISE; STATIC MAGNETIC-FIELD; ABSORPTION RATE; HUMAN HEAD; HUMAN-BODY; POWER DEPOSITION; B-1; DISTRIBUTION; FDTD-METHOD; SAR; ARRAY;
D O I
10.1002/mrm.22845
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Radiofrequency magnetic fields are critical to nuclear excitation and signal reception in magnetic resonance imaging. The interactions between these fields and human tissues in anatomical geometries results in a variety of effects regarding image integrity and safety of the human subject. In recent decades, numerical methods of calculation have been used increasingly to understand the effects of these interactions and aid in engineering better, faster, and safer equipment and methods. As magnetic resonance imaging techniques and technology have evolved through the years, so to have the requirements for meaningful interpretation of calculation results. Here, we review the basic physics of radiofrequency electromagnetics in magnetic resonance imaging and discuss a variety of ways radiofrequency field calculations are used in magnetic resonance imaging in engineering and safety assurance from simple systems and sequences through advanced methods of development for the future. Magn Reson Med 65:1470-1482, 2011. (C) 2011 Wiley-Liss, Inc.
引用
收藏
页码:1470 / 1482
页数:13
相关论文
共 107 条
[51]  
2-Q
[52]  
Hu B., 2010, P INT SOC MAG RESON, V18, P1481
[53]   Effect of RF coil excitation on field inhomogeneity at ultra high fields: A field optimized TEM resonator [J].
Ibrahim, TS ;
Lee, R ;
Baertlein, BA ;
Abduljalil, AM ;
Zhu, H ;
Robitaille, PML .
MAGNETIC RESONANCE IMAGING, 2001, 19 (10) :1339-1347
[54]   Computational analysis of the high pass birdcage resonator: finite difference time domain simulations for high-field MRI [J].
Ibrahim, TS ;
Lee, R ;
Baertlein, BA ;
Yu, Y ;
Robitaille, PML .
MAGNETIC RESONANCE IMAGING, 2000, 18 (07) :835-843
[55]  
Ibrahim TS., 2007, Proceedings of the Joint Annual Meeting ISMRM-ESMRMB, Berlin, May 19-25, 2007, P1013
[56]   Transmit SENSE [J].
Katscher, U ;
Börnert, P ;
Leussler, C ;
van den Brink, JS .
MAGNETIC RESONANCE IN MEDICINE, 2003, 49 (01) :144-150
[57]  
Katscher U., 2006, P 14 SCI M INT SOC M, P3035
[58]   ELECTROMAGNETIC-FIELDS OF SURFACE COIL INVIVO NMR AT HIGH-FREQUENCIES [J].
KELTNER, JR ;
CARLSON, JW ;
ROOS, MS ;
WONG, STS ;
WONG, TL ;
BUDINGER, TF .
MAGNETIC RESONANCE IN MEDICINE, 1991, 22 (02) :467-480
[59]   Fast MRI coil analysis based on 3-D electromagnetic and RF circuit co-simulation [J].
Kozlov, Mikhail ;
Turner, Robert .
JOURNAL OF MAGNETIC RESONANCE, 2009, 200 (01) :147-152
[60]  
LATTANZI R, 2007, JOINT ANN M ISMRM ES, P675