EZ-IO® intraosseous device implementation in German Helicopter Emergency Medical Service

被引:38
作者
Helm, Matthias [1 ]
Haunstein, Benedikt [1 ]
Schlechtriemen, Thomas [3 ]
Ruppert, Matthias [2 ]
Lampl, Lorenz [1 ]
Gaessler, Michael [2 ]
机构
[1] Armed Forces Med Ctr Ulm, Dept Anaesthesiol & Intens Care Med, Sect Emergency Med HEMS Christoph 22, D-8970 Ulm, Germany
[2] ADAC Luftrettung, Dept Med, Munich, Germany
[3] ADAC Luftrettung, Med Qual Management, Munich, Germany
关键词
Emergency medicine; EZ-IO (R); Helicopter Emergency Medical Service; Intraosseous vascular access; Vascular access; ACCESS; INFUSION;
D O I
10.1016/j.resuscitation.2014.12.015
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Background: Intraosseous access (IO) is a rapid and safe alternative when peripheral venous access is difficult. Our aim was to summarize the first three years experience with the use of a semi-automatic IO device (EZ-IO (R)) in German Helicopter Emergency Medical Service (HEMS). Methods: Included were all patients during study period (January 2009-December 2011) requiring an IO access performed by HEMS team. Outcome variables were IO rate, IO insertion success rates, site of IO access, type of EZ-IO (R) needle set used, strategy of vascular access, procedure related problems and operator's satisfaction. Results: IO rate was 0.3% (348/120.923). Overall success rate was 99.6% with a first attempt success rate of 85.9%; there was only one failure (0.4%). There were three insertion sites: proximal tibia (87.2%), distal tibia (7.5%) and proximal humerus (5.3%). Within total study group IO was predominantly the second-line strategy (39% vs. 61%, p < 0.001), but in children < 7 years, in trauma cases and in cardiac arrest IO was more often first-line strategy (64% vs. 28%, p < 0.001; 48% vs. 34%, p < 0.032; 50% vs. 29%, p < 0.002 respectively). Patients with IO access were significantly younger (41.7 +/- 28.7 vs. 56.5 +/- 24.4 years; p < 0.001), more often male (63.2% vs. 57.7%; p = 0.037), included more trauma cases (37.3% vs. 30.0%; p = 0.003) and more often patients with a NACA-Score >= 5 rating (77.0% vs. 18.6%; p < 0.001). Patients who required IO access generally presented with more severely compromised vital signs associated with the need for more invasive resuscitation actions such as intubation, chest drains, CPR and defibrillation. In 93% EZ-IO (R) needle set handling was rated "good". Problems were reported in 1.6% (needle dislocation 0.8%, needle bending 0.4% and parafusion 0.4%). Conclusions: The IO route was generally used in the most critically ill of patients. Our relatively low rate of usage would indicate that this would be compatible with the recommendations of established guidelines. The EZ-IO (R) intraosseous device proved feasible with a high success rate in adult and pediatric emergency patients in HEMS. (C) 2014 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:43 / 47
页数:5
相关论文
共 26 条
  • [1] Bernhard M, 2010, ANASTH INTENSIVMED, V51, pS615
  • [2] European Resuscitation Council Guidelines for Resuscitation 2010 Section 6 Paediatric life support
    Biarent, Dominique
    Bingham, Robert
    Eich, Christoph
    Lopez-Herce, Jesus
    Maconochie, Ian
    Rodriguez-Nunez, Antonio
    Rajka, Thomas
    Zideman, David
    [J]. RESUSCITATION, 2010, 81 (10) : 1364 - 1388
  • [3] Comparison of two intraosseous infusion systems for adult emergency medical use
    Brenner, Thorsten
    Bernhard, Michael
    Helm, Matthias
    Doll, Sara
    Voelkl, Alfred
    Ganion, Nicole
    Friedmann, Claudia
    Sikinger, Marcus
    Knapp, Juergen
    Martin, Eike
    Gries, Andre
    [J]. RESUSCITATION, 2008, 78 (03) : 314 - 319
  • [4] Committee on Trauma American College of Surgeons, 2008, ATLS ADV TRAUM SUPP
  • [5] Intraosseous Devices for Intravascular Access in Adult Trauma Patients
    Day, Michael W.
    [J]. CRITICAL CARE NURSE, 2011, 31 (02) : 76 - 88
  • [6] The role of intraosseous vascular access in the out-of-hospital environment (resource document to NAEMSP position statement)
    Fowler, Raymond
    Gallagher, John V.
    Isaacs, S. Marshal
    Ossman, Eric
    Pepe, Paul
    Wayne, Marvin
    [J]. PREHOSPITAL EMERGENCY CARE, 2007, 11 (01) : 63 - 66
  • [7] Consecutive field trials using two different intraosseous devices
    Frascone, Ralph J.
    Jensen, Joe P.
    Kaye, Kory
    Salzman, Joshua G.
    [J]. PREHOSPITAL EMERGENCY CARE, 2007, 11 (02) : 164 - 171
  • [8] Efficacy and safety of the EZ-IO™ intraosseous device: Out-of-hospital implementation of a management algorithm for difficult vascular access
    Gazin, Nicolas
    Auger, Harold
    Jabre, Patricia
    Jaulin, Christine
    Lecarpentier, Eric
    Bertrand, Catherine
    Margenet, Alain
    Combes, Xavier
    [J]. RESUSCITATION, 2011, 82 (01) : 126 - 129
  • [9] Helm M, 2011, ANAESTHESIST, V60, P1119, DOI 10.1007/s00101-011-1937-5
  • [10] Kellner P, 2011, NOTFALL RETTUNGSMED, V14, P379, DOI 10.1007/s10049-010-1381-0