Comprehensive Echocardiographic Assessment of the Hemodynamic Parameters of 285 Tricuspid Valve Bioprostheses Early after Implantation

被引:26
作者
Blauwet, Lori A. [1 ]
Danielson, Gordon K. [2 ]
Burkhart, Harold M. [2 ]
Dearani, Joseph A. [2 ]
Malouf, Joseph F. [1 ]
Connolly, Heidi M. [1 ]
Hodge, David O. [3 ]
Herges, Regina M. [3 ]
Miller, Fletcher A., Jr. [1 ]
机构
[1] Mayo Clin, Div Cardiovasc Dis, Rochester, MN 55905 USA
[2] Mayo Clin, Div Cardiovasc Surg, Rochester, MN 55905 USA
[3] Mayo Clin, Div Biomed Informat & Biostat, Rochester, MN 55905 USA
关键词
Doppler; Echocardiography; Prosthesis; Tricuspid; VELOCITY INTEGRAL RATIO; PROSTHESIS PERFORMANCE INDEX; PRESSURE HALF-TIME; DOPPLER-ECHOCARDIOGRAPHY; MECHANICAL PROSTHESES; CONTINUITY EQUATION; MITRAL PROSTHESES; STENOSIS; POSITION; RECOMMENDATIONS;
D O I
10.1016/j.echo.2010.08.007
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Doppler-derived hemodynamic data for normal tricuspid valve bioprostheses are limited. Methods: A comprehensive retrospective Doppler echocardiographic assessment of 285 normal Carpentier-Edwards Duraflex, Medtronic Mosaic, St. Jude Medical Biocor, Carpentier-Edwards Perimount, and Medtronic Hancock II tricuspid valve bioprostheses was performed early after implantation. All the important Doppler-derived hemodynamic variables reported to date for mitral valve prostheses were used. Mean values for hemodynamic variables were obtained by averaging measurements of five and nine consecutive cardiac cycles. Results: No clinically significant difference was found in the mean values obtained for the Doppler parameters when measurements were averaged from five or nine consecutive cardiac cycles. The mean value for the mean gradient was 5.2 mm Hg. Regardless of valve type and body surface area, pressure half-time was <200 msec for all 76 patients in whom it could be measured. Mean gradient <9 mm Hg, E velocity <2.1 m/sec, time-velocity integral of the tricuspid valve prosthesis <66 cm, and ratio of the time-velocity integral of the tricuspid valve prosthesis to the time-velocity integral of the left ventricular outflow tract <3.3 were recorded in 254 of the 285 patients (89%). Conclusions: This study establishes parameters for Doppler-derived hemodynamic data for various types of normal tricuspid valve bioprostheses. These threshold values (mean + 2 standard deviations) are specific, but not sensitive, for identifying tricuspid valve bioprosthesis dysfunction. Prostheses with hemodynamic values that are higher than these threshold values are likely dysfunctional, but in select cases, tricuspid valve bioprosthesis dysfunction may be present even when hemodynamic values are lower than these thresholds. (J Am Soc Echocardiogr 2010;23:1045-59.)
引用
收藏
页码:1045 / U72
页数:17
相关论文
共 29 条
[1]  
Aoyagi S, 1993, J Heart Valve Dis, V2, P279
[2]  
Aoyagi Shigeaki, 2003, Asian Cardiovasc Thorac Ann, V11, P193
[3]   Echocardiographic Assessment of Valve Stenosis: EAE/ASE Recommendations for Clinical Practice [J].
Baumgartner, Helmut ;
Hung, Judy ;
Bermejo, Javier ;
Chambers, John B. ;
Evangelista, Arturo ;
Griffin, Brian P. ;
Iung, Bernard ;
Otto, Catherine M. ;
Pellikka, Patricia A. ;
Quinones, Miguel .
JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 2009, 22 (01) :1-23
[4]   DOPPLER-ECHOCARDIOGRAPHIC ASSESSMENT WITH THE CONTINUITY EQUATION OF ST-JUDE MEDICAL MECHANICAL PROSTHESES IN THE MITRAL-VALVE POSITION [J].
BITAR, JN ;
LECHIN, ME ;
SALAZAR, G ;
ZOGHBI, WA .
AMERICAN JOURNAL OF CARDIOLOGY, 1995, 76 (04) :287-293
[5]   Doppler echocardiography of 79 normal CarboMedics mitral prostheses: A comprehensive assessment including time-velocity integral ratio and prosthesis performance index [J].
Blauwet, Lori A. ;
Malouf, Joseph F. ;
Connolly, Heidi M. ;
Hodge, David O. ;
Herges, Regina M. ;
Sundt, Thoralf M., III ;
Miller, Fletcher A., Jr. .
JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 2007, 20 (10) :1125-1130
[6]   Doppler Echocardiography of 240 Normal Carpentier-Edwards Duraflex Porcine Mitral Bioprostheses: A Comprehensive Assessment Including Time Velocity Integral Ratio and Prosthesis Performance Index [J].
Blauwet, Lori A. ;
Malouf, Joseph F. ;
Connolly, Heidi M. ;
Hodge, David O. ;
Evans, Katie N. ;
Herges, Regina M. ;
Sundt, Thoralf M., III ;
Miller, Fletcher A., Jr. .
JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 2009, 22 (04) :388-393
[7]  
CARABELLO BA, 1986, CARDIAC CATHETERIZAT, P143
[8]   MODIFIED ORIFICE EQUATION FOR CALCULATION OF MITRAL-VALVE AREA [J].
COHEN, MV ;
GORLIN, R .
AMERICAN HEART JOURNAL, 1972, 84 (06) :839-&
[9]   DOPPLER HEMODYNAMIC PROFILES OF 82 CLINICALLY AND ECHOCARDIOGRAPHICALLY NORMAL TRICUSPID-VALVE PROSTHESES [J].
CONNOLLY, HM ;
MILLER, FA ;
TAYLOR, CL ;
NAESSENS, JM ;
SEWARD, JB ;
TAJIK, AJ .
CIRCULATION, 1993, 88 (06) :2722-2727
[10]   CLINICAL AND HEMODYNAMIC ASSESSMENT OF THE HANCOCK-II BIOPROSTHESIS [J].
DAVID, TE ;
ARMSTRONG, S ;
SUN, Z .
ANNALS OF THORACIC SURGERY, 1992, 54 (04) :661-668