Biomechanical study of injectable hollow pedicle screws for PMMA augmentation in severely osteoporotic lumbar vertebrae: effect of PMMA distribution and volume on screw stability

被引:14
作者
Liu, Da [1 ]
Sheng, Jun [1 ]
Wu, Hong-hua [1 ]
Kang, Xia [1 ]
Xie, Qing-yun [1 ]
Luo, Yang [2 ]
Zhou, Jiang-jun [3 ]
Zheng, Wei [1 ]
机构
[1] Chengdu Mil Gen Hosp, Dept Orthopaed, Chengdu, Sichuan, Peoples R China
[2] Chengdu Mil Gen Hosp, Dept Anesthesiol, Chengdu, Sichuan, Peoples R China
[3] 184 Hosp Nanjing Mil Reg, Dept Orthopaed, Yingtan, Jiangxi, Peoples R China
基金
中国国家自然科学基金;
关键词
severe osteoporosis; injectable hollow pedicle screw; polymethylmethacrylate volume; polymethylmethacrylate distribution; maximum axial pullout strength; lumbar vertebrae; CEMENT AUGMENTATION; PULLOUT STRENGTH; BURST FRACTURES; FIXATION; POLYMETHYLMETHACRYLATE; SPINE; STABILIZATION; REVISION;
D O I
10.3171/2018.4.SPINE171225
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
OBJECTIVE The purpose of this study was to compare stability of injectable hollow pedicle screws with different numbers of holes using different volumes of polymethylmethacrylate (PMMA) in severely osteoporotic lumbar vertebrae and analyze the relationship between screw stability and distribution and volume of PMMA. METHODS Forty-eight severely osteoporotic cadaveric lumbar vertebrae were randomly divided into 3 groups-groups A, B, and C (16 vertebrae per group). The screws used in group A had 4 holes (2 pairs of holes, with the second hole of each pair placed 180 degrees further along the thread than the first). The screws used in group B had 6 holes (3 pairs of holes, placed with the same 180 degrees difference in position). Unmodified conventional screws were used in group C. Each group was randomly divided into subgroups 0, 1, 2, and 3, with different volumes of PMMA used in each subgroup. Type A and B pedicle screws were directly inserted into the vertebrae in groups A and B, respectively, and then different volumes of PMMA were injected through the screws into the vertebrae in subgroups 0, 1, 2, and 3. The pilot hole was filled with different volumes of PMMA followed by insertion of screws in groups C0, C1, C2, and C3. Distributions of PMMA were evaluated radiographically, and axial pull-out tests were performed to measure the maximum axial pullout strength (F-max ). RESULTS Radiographic examination revealed that PMMA surrounded the anterior third of the screws in the vertebral bodies (VBs) in groups A1, A2, and A3; the middle third of screws in the junction area of the vertebral body (VB) and pedicle in groups B1, B2, and B3; and the full length of screws evenly in both VB and pedicle in groups C1, C2, and C3. In addition, in groups A3 and B3, PMMA from each of the screws (left and right) was in contact with PMMA from the other screw and the PMMA was closer to the posterior wall and pedicle than in groups A1, A2, B1, and B2. One instance of PMMA leakage was found (in group B3). Two-way analysis of variance revealed that 2 factors-distribution and volume of PMMA-significantly influenced F-max (p < 0.05) but that they were not significantly correlated (p = 0.078). The F-max values in groups in which screws were augmented with PMMA were significantly better than those in groups in which no PMMA was used (p < 0.05). CONCLUSIONS PMMA can significantly improve stability of different injectable pedicle screws in severely osteoporotic lumbar vertebrae, and screw stability is significantly correlated with distribution and volume of PMMA. The closer the PMMA is to the pedicle and the greater the quantity of injected PMMA used, the greater the pedicle screw stability is. Injection of 3.0 mL PMMA through screws with 4 holes (2 pair of holes, with the screws in each pair placed on opposite sides of the screw) produces optimal stability in severely osteoporotic lumbar vertebrae.
引用
收藏
页码:639 / 646
页数:8
相关论文
共 25 条
[1]   Assessment of different screw augmentation techniques and screw designs in osteoporotic spines [J].
Becker, S. ;
Chavanne, A. ;
Spitaler, R. ;
Kropik, K. ;
Aigner, N. ;
Ogon, M. ;
Redl, H. .
EUROPEAN SPINE JOURNAL, 2008, 17 (11) :1462-1469
[2]   Revision characteristics of cement-augmented, cannulated-fenestrated pedicle screws in the osteoporotic vertebral body: a biomechanical in vitro investigation Technical note [J].
Blattert, Thomas R. ;
Glasmacher, Stefan ;
Riesner, Hans-Joachim ;
Josten, Christoph .
JOURNAL OF NEUROSURGERY-SPINE, 2009, 11 (01) :23-27
[3]   Revision of Cannulated and Perforated Cement-Augmented Pedicle Screws A Biomechanical Study in Human Cadavers [J].
Bullmann, Viola ;
Schmoelz, Werner ;
Richter, Marcus ;
Grathwohl, Corinna ;
Schulte, Tobias L. .
SPINE, 2010, 35 (19) :E932-E939
[4]   Polymethylmethacrylate augmentation of pedicle screw for osteoporotic spinal surgery [J].
Chang, Ming-Chau ;
Liu, Chien-Lin ;
Chen, Tain-Hsiung .
SPINE, 2008, 33 (10) :E317-E324
[5]   Biomechanical analysis of different types of pedicle screw augmentation: A cadaveric and synthetic bone sample study of instrumented vertebral specimens [J].
Chao, Kuo-Hua ;
Lai, Yu-Shu ;
Chen, Wen-Chuan ;
Chang, Chia-Ming ;
McClean, Colin J. ;
Fan, Chang-Yuan ;
Chang, Chia-Hao ;
Lin, Leou-Chyr ;
Cheng, Cheng-Kung .
MEDICAL ENGINEERING & PHYSICS, 2013, 35 (10) :1506-1512
[6]   Pullout strength for cannulated pedicle screws with bone cement augmentation in severely osteoporotic bone: Influences of radial hole and pilot hole tapping [J].
Chen, Lih-Huei ;
Tai, Ching-Lung ;
Lai, Po-Liang ;
Lee, De-Mei ;
Tsai, Tsung-Tin ;
Fu, Tsai-Sheng ;
Niu, Chi-Chien ;
Chen, Wen-Jer .
CLINICAL BIOMECHANICS, 2009, 24 (08) :613-618
[7]   Biomechanical study of expandable pedicle screw fixation in severe osteoporotic bone comparing with conventional and cement-augmented pedicle screws [J].
Chen, Yi-Long ;
Chen, Wen-Chuan ;
Chou, Chi-Wei ;
Chen, Jou-Wen ;
Chang, Chia-Ming ;
Lai, Yu-Shu ;
Cheng, Cheng-Kung ;
Wang, Shih-Tien .
MEDICAL ENGINEERING & PHYSICS, 2014, 36 (11) :1416-1420
[8]   Pedicle Screw Design and Cement Augmentation in Osteoporotic Vertebrae Effects of Fenestrations and Cement Viscosity on Fixation and Extraction [J].
Choma, Theodore J. ;
Pfeiffer, Ferris M. ;
Swope, Ryan W. ;
Hirner, Jesse P. .
SPINE, 2012, 37 (26) :E1628-E1632
[9]  
Cook Stephen D, 2004, Spine J, V4, P402, DOI 10.1016/j.spinee.2003.11.010
[10]   Complications of thoracic pedicle screws in scoliosis treatment [J].
Di Silvestre, Mario ;
Parisini, Patrizio ;
Lolli, Francesco ;
Bakaloudis, Georgios .
SPINE, 2007, 32 (15) :1655-1661