Evaluation of the biodurability of polyurethane-covered stent using a flow phantom

被引:39
作者
Kim, DH
Kang, SG
Choi, JR
Byun, JN
Kim, YC
Ahn, YM
机构
[1] Chosun Univ Hosp, Dept Diagnost Radiol, Dong Gu, Gwangju 501140, South Korea
[2] Hansung Univ, Dept Clothing & Text, Seoul, South Korea
关键词
biliary stent; biodegradation; polyurethane membrane;
D O I
10.3348/kjr.2001.2.2.75
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: To evaluate the biodurability of the covering material in retrievable metallic stents covered with polycarbonate polyurethane. Materials and Methods: Using a peristaltic pump at a constant rate of 1 ml/min, bile was recirculated from a reservoir through a long tube containing four stents. Each of these was removed from the system every two weeks and a radial tensile strength test and scanning electron microscopy (SEM) were performed. Each stent, removed at 2, 4, 6 and 8 weeks, was compared with a control stent not exposed to bile juice. Results: Gross examination showed that stents were intact at 2 weeks, but at 4, 6 and 8 weeks cracks were observed, The size of these increased gradually in accordance with the duration of exposure, and at 8 weeks several large holes in the polyurethane membrane were evident. With regard to radial tensile strength, extension and peak load at break were 84.47% and 10.030 N/mm, 54.90% and 6.769 N/mm, 16.55% and 2.452 N/mm, 11.21% and 1.373 N/mm at 0, 2, 4 and 6 weeks, respectively. Scanning electron microscopy at 2 weeks revealed intermittent pitting and cracking, and examination at 4, 6 and 8 weeks showed that the size of these defects was gradually increasing. Conclusion: When the polyurethane membrane was exposed to bile, biodegradation was first observed at week two and increased gradually according to the duration of exposure.
引用
收藏
页码:75 / 79
页数:5
相关论文
共 14 条
[1]   EVALUATION OF POLYMER-COATED BALLOON-EXPANDABLE STENTS IN BILE-DUCTS [J].
ALVARADO, R ;
PALMAZ, JC ;
GARCIA, OJ ;
TIO, FO ;
REES, CR .
RADIOLOGY, 1989, 170 (03) :975-978
[2]   BENIGN BILIARY STRICTURES - TREATMENT WITH PERCUTANEOUS CHOLANGIOPLASTY [J].
CITRON, SJ ;
MARTIN, LG .
RADIOLOGY, 1991, 178 (02) :339-341
[3]   SELF-EXPANDING STAINLESS-STEEL BILIARY STENTS [J].
COONS, HG .
RADIOLOGY, 1989, 170 (03) :979-983
[4]   In vitro and in vivo biodurability of a compliant microporous vascular graft [J].
Edwards, A ;
Carson, RJ ;
Szycher, M ;
Bowald, S .
JOURNAL OF BIOMATERIALS APPLICATIONS, 1998, 13 (01) :23-45
[5]  
GORE RM, 2000, TXB GASTROINTESTINAL, P1349
[6]   METALLIC STENTS IN BENIGN BILIARY STRICTURES - 3-YEAR FOLLOW-UP [J].
MACCIONI, F ;
ROSSI, M ;
SALVATORI, FM ;
RICCI, P ;
BEZZI, M ;
ROSSI, P .
CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 1992, 15 (06) :360-366
[7]  
Mathur AB, 1997, J BIOMED MATER RES, V36, P246, DOI 10.1002/(SICI)1097-4636(199708)36:2<246::AID-JBM14>3.3.CO
[8]  
2-9
[9]   BILIARY STRICTURE DILATATION - MULTICENTER REVIEW OF CLINICAL MANAGEMENT IN 73 PATIENTS [J].
MUELLER, PR ;
VANSONNENBERG, E ;
FERRUCCI, JT ;
WEYMAN, PJ ;
BUTCH, RJ ;
MALT, RA ;
BURHENNE, HJ .
RADIOLOGY, 1986, 160 (01) :17-22
[10]   PERCUTANEOUS MANAGEMENT OF BENIGN BILIARY STRICTURES WITH BALLOON DILATION AND SELF-EXPANDING METALLIC STENTS [J].
ROSSI, P ;
SALVATORI, FM ;
BEZZI, M ;
MACCIONI, F ;
PORCARO, ML ;
RICCI, P .
CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 1990, 13 (04) :231-239