Constrained predictive control of thermal therapies for minimum-time delivery of thermal dose

被引:8
作者
Arora, Dhiraj [1 ]
Skliar, Mikhail
Cooley, Daniel
Roemer, Robert B.
机构
[1] GE Global Res, Moscow 123098, Russia
[2] Univ Utah, Dept Chem Engn, Salt Lake City, UT 84112 USA
[3] Univ Utah, Dept Mech Engn, Salt Lake City, UT 84112 USA
[4] George Mason Univ, Arlington, VA USA
关键词
focused ultrasound; minimum-time control; thermal dose control; thermal therapies;
D O I
10.1109/TCST.2007.899680
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
A method for time-optimal, direct control of thermal dose in. thermal therapies is developed and experimentally validated using a focused ultrasound transducer and a phantom patient. State constraint on the maximum allowable temperature in a selected spatial location is imposed to prevent damage to critical normal tissues. A saturation constraint on the ultrasound power is imposed to reflect hardware limitations. It is shown that to achieve the minimum time treatment it is necessary to control the therapy with either saturated ultrasound power or active normal-tissue temperature constraints. The special cases for which the necessary condition are also sufficient for time optimality are also established. The model-based treatment control system is then designed that ensures that the necessary condition for time optimal treatment, is satisfied throughout the treatment. During validation experiments, the ultrasound specific absorption rate and thermal response models of the phantom, needed for the operation of the designed treatment control system, were identified using temperature measurements. The performance of the treatment control system during the experiments demonstrates that the proposed approach is effective at delivering the desired thermal dose in a near-minimum time without violating safety constraints imposed in healthy tissues.
引用
收藏
页码:1030 / 1037
页数:8
相关论文
共 18 条
[1]   MR thermometry-based feedback control of efficacy and safety in minimum-time thermal therapies:: Phantom and in-vivo evaluations [J].
Arora, D ;
Cooley, D ;
Perry, T ;
Guo, JY ;
Richardson, A ;
Moellmer, J ;
Hadley, R ;
Parker, D ;
Skliar, M ;
Roemer, RB .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2006, 22 (01) :29-42
[2]   Control of thermal therapies with moving power deposition field [J].
Arora, D ;
Minor, MA ;
Skliar, M ;
Roemer, RB .
PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (05) :1201-1219
[3]   Direct thermal dose control of constrained focused ultrasound treatments: phantom and in vivo evaluation [J].
Arora, D ;
Cooley, D ;
Perry, T ;
Skliar, M ;
Roemer, RB .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (08) :1919-1935
[4]   Minimum-time thermal dose control of thermal therapies [J].
Arora, D ;
Skliar, M ;
Roemer, RB .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2005, 52 (02) :191-200
[5]  
Arora D, 2003, P AMER CONTR CONF, P1200
[6]   Model-predictive control of hyperthermia treatments [J].
Arora, D ;
Skliar, M ;
Roemer, RB .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2002, 49 (07) :629-639
[7]   RTOG QUALITY ASSURANCE GUIDELINES FOR CLINICAL-TRIALS USING HYPERTHERMIA [J].
DEWHIRST, MW ;
PHILLIPS, TL ;
SAMULSKI, TV ;
STAUFFER, P ;
SHRIVASTAVA, P ;
PALIWAL, B ;
PAJAK, T ;
GILLIM, M ;
SAPOZINK, M ;
MYERSON, R ;
WATERMAN, FM ;
SAPARETO, SA ;
CORRY, P ;
CETAS, TC ;
LEEPER, DB ;
FESSENDEN, P ;
KAPP, D ;
OLESON, JR ;
EMAMI, B .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1990, 18 (05) :1249-1259
[8]   Tumor ablation with radio-frequency energy [J].
Gazelle, GS ;
Goldberg, SN ;
Solbiati, L ;
Livraghi, T .
RADIOLOGY, 2000, 217 (03) :633-646
[9]   The feasibility of MRI feedback control for intracavitary phased array hyperthermia treatments [J].
Hutchinson, E ;
Dahleh, M ;
Hynynen, K .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1998, 14 (01) :39-56
[10]   The safety and feasibility of extracorporeal high-intensity focused ultrasound (HIFU) for the treatment of liver and kidney tumours in a Western population [J].
Illing, RO ;
Kennedy, JE ;
Wu, F ;
ter Haar, GR ;
Protheroe, AS ;
Friend, PJ ;
Gleeson, FV ;
Cranston, DW ;
Phillips, RR ;
Middleton, MR .
BRITISH JOURNAL OF CANCER, 2005, 93 (08) :890-895