Force and depth of mechanical chest compressions and their relation to chest height and gender in an out-of-hospital setting

被引:22
作者
Beesems, Stefanie G. [1 ]
Hardig, Bjarne Madsen [2 ]
Nilsson, Anders [2 ]
Koster, Rudolph W. [1 ]
机构
[1] Univ Amsterdam, Acad Med Ctr, Dept Cardiol, NL-1105 AZ Amsterdam, Netherlands
[2] Physiocontrol Jolife AB, Lund, Sweden
关键词
Cardiopulmonary resuscitation; Mechanical chest compression; Chest height; Chest compression force; Chest compression depth; EUROPEAN RESUSCITATION COUNCIL; AMERICAN-HEART-ASSOCIATION; BASIC LIFE-SUPPORT; CARDIOPULMONARY-RESUSCITATION; CARDIAC-ARREST; FEEDBACK; QUALITY; GUIDELINES; PERSONNEL; TRIAL;
D O I
10.1016/j.resuscitation.2015.03.020
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Introduction: The LUCAS 2 device stores technical data that documents the chest compression process. We analyzed chest wall dimensions and mechanics stored during chest compressions on humans using data gathered with the LUCAS 2 device. Methods: Data from LUCAS 2 devices used in out-of-hospital cardiac arrest were downloaded with dedicated proprietary software and matched to the corresponding patient data. Cases were included only if the suction cup was placed correctly, if it was not realigned during the first 5 min of chest compressions, and if no other anomaly in device use was noted. Trauma cases were excluded. Results: Ninety-five patients were included. All patients received manual cardiopulmonary resuscitation prior to the application of the device. The mean (SD) chest height was 232 (25) mm for males and 209 (26) mm for females (P < 0.001). The mean (min-max) compression depth in patients with chest height > 185 mm was 53 (50-55) mm, corresponding with 19-28% of the chest diameter. The mean force required to achieve the compression depth of 53 mm ranged between 219 and 568 N. No correlation was found between chest height and force to reach 53 mm depth (females: R-2 = 0.001, males: R-2 = 0.007). Conclusion: There was a large variation of the required force to achieve a compression depth of 53 mm. No correlation was seen between chest height and maximum force required to compress the chest 53 mm. (C) 2015 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:67 / 72
页数:6
相关论文
共 19 条
[11]   The quality of chest compressions by trained personnel: The effect of feedback, via the CPREzy, in a randomized controlled trial using a manikin model [J].
Noordergraaf, GJ ;
Drinkwaard, BWPM ;
van Berkom, PFJ ;
van Hemert, HP ;
Venema, A ;
Scheffer, GJ ;
Noordergraaf, A .
RESUSCITATION, 2006, 69 (02) :241-252
[12]   Chest compressions by ambulance personnel on chests with variable stiffness:: Abilities and attitudes [J].
Odegaard, Silje ;
Kramer-Johansen, Jo ;
Bromley, Allan ;
Myklebust, Helge ;
Nysaether, Jon ;
Wik, Lars ;
Steen, Petter Andreas .
RESUSCITATION, 2007, 74 (01) :127-134
[13]   THE EFFECT OF APPLIED CHEST COMPRESSION FORCE ON SYSTEMIC ARTERIAL-PRESSURE AND END-TIDAL CARBON-DIOXIDE CONCENTRATION DURING CPR IN HUMAN-BEINGS [J].
ORNATO, JP ;
LEVINE, RL ;
YOUNG, DS ;
RACHT, EM ;
GARNETT, AR ;
GONZALEZ, ER .
ANNALS OF EMERGENCY MEDICINE, 1989, 18 (07) :732-737
[14]   Compression feedback devices over estimate chest compression depth when performed on a bed [J].
Perkins, Gavin D. ;
Kocierz, Laura ;
Smith, Samuel C. L. ;
McCulloch, Robert A. ;
Davies, Robin P. .
RESUSCITATION, 2009, 80 (01) :79-82
[15]   Radiological assessment of the adult chest: Implications for chest compressions [J].
Pickard, Amelia ;
Darby, Michael ;
Soar, Jasmeet .
RESUSCITATION, 2006, 71 (03) :387-390
[16]   What Is the Optimal Chest Compression Depth During Out-of-Hospital Cardiac Arrest Resuscitation of Adult Patients? [J].
Stiell, Ian G. ;
Brown, Siobhan P. ;
Nichol, Graham ;
Cheskes, Sheldon ;
Vaillancourt, Christian ;
Callaway, Clifton W. ;
Morrison, Laurie J. ;
Christenson, James ;
Aufderheide, Tom P. ;
Davis, Daniel P. ;
Free, Cliff ;
Hostler, Dave ;
Stouffer, John A. ;
Idris, Ahamed H. .
CIRCULATION, 2014, 130 (22) :1962-+
[17]   Compression force-depth relationship during out-of-hospital cardiopulmonary resuscitation [J].
Tomlinson, A. E. ;
Nysaether, J. ;
Kramer-Johansen, J. ;
Steen, P. A. ;
Dorph, E. .
RESUSCITATION, 2007, 72 (03) :364-370
[18]   ELASTIC PROPERTIES OF THE HUMAN CHEST DURING CARDIOPULMONARY RESUSCITATION [J].
TSITLIK, JE ;
WEISFELDT, ML ;
CHANDRA, N ;
EFFRON, MB ;
HALPERIN, HR ;
LEVIN, HR .
CRITICAL CARE MEDICINE, 1983, 11 (09) :685-692
[19]   Quality of cardiopulmonary resuscitation during out-of-hospital cardiac arrest [J].
Wik, L ;
Kramer-Johansen, J ;
Myklebust, H ;
Sorebo, H ;
Svensson, L ;
Fellows, B ;
Steen, PA .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2005, 293 (03) :299-304