Force and depth of mechanical chest compressions and their relation to chest height and gender in an out-of-hospital setting

被引:22
作者
Beesems, Stefanie G. [1 ]
Hardig, Bjarne Madsen [2 ]
Nilsson, Anders [2 ]
Koster, Rudolph W. [1 ]
机构
[1] Univ Amsterdam, Acad Med Ctr, Dept Cardiol, NL-1105 AZ Amsterdam, Netherlands
[2] Physiocontrol Jolife AB, Lund, Sweden
关键词
Cardiopulmonary resuscitation; Mechanical chest compression; Chest height; Chest compression force; Chest compression depth; EUROPEAN RESUSCITATION COUNCIL; AMERICAN-HEART-ASSOCIATION; BASIC LIFE-SUPPORT; CARDIOPULMONARY-RESUSCITATION; CARDIAC-ARREST; FEEDBACK; QUALITY; GUIDELINES; PERSONNEL; TRIAL;
D O I
10.1016/j.resuscitation.2015.03.020
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Introduction: The LUCAS 2 device stores technical data that documents the chest compression process. We analyzed chest wall dimensions and mechanics stored during chest compressions on humans using data gathered with the LUCAS 2 device. Methods: Data from LUCAS 2 devices used in out-of-hospital cardiac arrest were downloaded with dedicated proprietary software and matched to the corresponding patient data. Cases were included only if the suction cup was placed correctly, if it was not realigned during the first 5 min of chest compressions, and if no other anomaly in device use was noted. Trauma cases were excluded. Results: Ninety-five patients were included. All patients received manual cardiopulmonary resuscitation prior to the application of the device. The mean (SD) chest height was 232 (25) mm for males and 209 (26) mm for females (P < 0.001). The mean (min-max) compression depth in patients with chest height > 185 mm was 53 (50-55) mm, corresponding with 19-28% of the chest diameter. The mean force required to achieve the compression depth of 53 mm ranged between 219 and 568 N. No correlation was found between chest height and force to reach 53 mm depth (females: R-2 = 0.001, males: R-2 = 0.007). Conclusion: There was a large variation of the required force to achieve a compression depth of 53 mm. No correlation was seen between chest height and maximum force required to compress the chest 53 mm. (C) 2015 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:67 / 72
页数:6
相关论文
共 19 条
[1]   Quality of cardiopulmonary resuscitation during in-hospital cardiac arrest [J].
Abella, BS ;
Alvarado, JP ;
Myklebust, H ;
Edelson, DP ;
Barry, A ;
O'Hearn, N ;
Vanden Hoek, TL ;
Becker, LB .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2005, 293 (03) :305-310
[2]  
[Anonymous], 2011, INSTRUCTIONS USE SOF
[3]   CPREzy™ improves performance of external chest compressions in simulated cardiac arrest [J].
Beckers, Stefan K. ;
Skorning, Max H. ;
Fries, Michael ;
Bickenbach, Johannes ;
Beuerlein, Stephan ;
Derwall, Matthias ;
Kuhlen, Ralf ;
Rossaint, Rolf .
RESUSCITATION, 2007, 72 (01) :100-107
[4]   Impact of Onsite or Dispatched Automated External Defibrillator Use on Survival After Out-of-Hospital Cardiac Arrest [J].
Berdowski, Jocelyn ;
Blom, Marieke T. ;
Bardai, Abdennasser ;
Tan, Hanno L. ;
Tijssen, Jan G. P. ;
Koster, Rudolph W. .
CIRCULATION, 2011, 124 (20) :2225-2232
[5]   Part 5: Adult Basic Life Support 2010 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care [J].
Berg, Robert A. ;
Hemphill, Robin ;
Abella, Benjamin S. ;
Aufderheide, Tom P. ;
Cave, Diana M. ;
Hazinski, Mary Fran ;
Lerner, E. Brooke ;
Rea, Thomas D. ;
Sayre, Michael R. ;
Swor, Robert A. .
CIRCULATION, 2010, 122 (18) :S685-S705
[6]  
Geddes Leslie A, 2007, Cardiovasc Eng, V7, P47, DOI 10.1007/s10558-007-9029-5
[7]   STERNAL FORCE-DISPLACEMENT RELATIONSHIP DURING CARDIOPULMONARY-RESUSCITATION [J].
GRUBEN, KG ;
GUERCI, AD ;
HALPERIN, HR ;
POPEL, AS ;
TSITLIK, JE .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1993, 115 (02) :195-201
[8]   Effect of real-time feedback during cardiopulmonary resuscitation outside hospital: prospective, cluster-randomised trial [J].
Hostler, David ;
Everson-Stewart, Siobhan ;
Rea, Thomas D. ;
Stiell, Ian G. ;
Callaway, Clifton W. ;
Kudenchuk, Peter J. ;
Sears, Gena K. ;
Emerson, Scott S. ;
Nichol, Graham .
BRITISH MEDICAL JOURNAL, 2011, 342 :371
[10]   European Resuscitation Council Guidelines for Resuscitation 2010 Section 2 Adult basic life support and use of automated external defibrillators [J].
Koster, Rudolph W. ;
Baubin, Michael A. ;
Bossaert, Leo L. ;
Caballero, Antonio ;
Cassan, Pascal ;
Castren, Maaret ;
Granja, Cristina ;
Handley, Anthony J. ;
Monsieurs, Koenraad G. ;
Perkins, Gavin D. ;
Raffay, Violetta ;
Sandroni, Claudio .
RESUSCITATION, 2010, 81 (10) :1277-1292