X-ray micro-CT with a displaced detector array: Application to helical cone-beam reconstruction

被引:16
作者
Liu, V
Lariviere, NR
Wang, G [1 ]
机构
[1] Univ Iowa, CT Micro CT Lab, Dept Radiol, Iowa City, IA 52242 USA
[2] Univ Iowa, Dept Biomed Engn, Iowa City, IA 52242 USA
关键词
micro-CT; helical/spiral scanning; cone-beam; field of view; displaced detector array;
D O I
10.1118/1.1609058
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In x-ray micro-CT applications, it is useful to increase the field of view by offsetting a two-dimensional (2D) detector array. In this technical note, we briefly review the methods for image reconstruction with an asymmetric 2D detector array, elaborate on the use of an associated weighting scheme in the case of helical/spiral cone-beam scanning, and perform a series of numerical tests to demonstrate helical cone-beam image reconstruction with such an arrangement. (C) 2003 American Association of Physicists in Medicine.
引用
收藏
页码:2758 / 2761
页数:4
相关论文
共 12 条
[1]   SIMULTANEOUS ALGEBRAIC RECONSTRUCTION TECHNIQUE (SART) - A SUPERIOR IMPLEMENTATION OF THE ART ALGORITHM [J].
ANDERSEN, AH ;
KAK, AC .
ULTRASONIC IMAGING, 1984, 6 (01) :81-94
[2]   Cone-beam CT from width-truncated projections [J].
Cho, PS ;
Rudd, AD ;
Johnson, RH .
COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 1996, 20 (01) :49-57
[3]   PRACTICAL CONE-BEAM ALGORITHM [J].
FELDKAMP, LA ;
DAVIS, LC ;
KRESS, JW .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1984, 1 (06) :612-619
[4]   Convergence of the simultaneous algebraic reconstruction technique (SART) [J].
Jiang, M ;
Wang, G .
IEEE TRANSACTIONS ON IMAGE PROCESSING, 2003, 12 (08) :957-961
[5]   Convergence studies on iterative algorithms for image reconstruction [J].
Jiang, M ;
Wang, G .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2003, 22 (05) :569-579
[6]  
LARIVIERE PJ, 1990, IEEE T NUCL SCI, V48, P1357
[7]  
MULLER M, 1994, APPL OPTICS, V33, P8255, DOI 10.1364/AO.33.008255
[8]  
Shepp L A, 1982, IEEE Trans Med Imaging, V1, P113, DOI 10.1109/TMI.1982.4307558
[9]  
SILVER MD, 1994, RADIOLOGY, V193, P227
[10]   DEBLURRING SUBJECT TO NONNEGATIVITY CONSTRAINTS [J].
SNYDER, DL ;
SCHULZ, TJ ;
OSULLIVAN, JA .
IEEE TRANSACTIONS ON SIGNAL PROCESSING, 1992, 40 (05) :1143-1150