Costoclavicular pinching: a complication of long-term central venous catheters - A report of three cases

被引:8
作者
Nostdahl, T [1 ]
Waagsbo, NA [1 ]
机构
[1] Telemark Cent Hosp, Dept Anaesthesia, N-3906 Porsgrunn, Norway
关键词
central venous catheter; vascular port; material fatigue; leakage; pinch-off;
D O I
10.1111/j.1399-6576.1998.tb05337.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
In the course of 1994-1995 we found three nearly identical cases involving fracture of polyurethane catheters from subcutaneous implantable vascular ports. This resulted in complications: leakage of drugs and infusions and thereby soft tissue damage. The leakage occurred in the costoclavicular space in all instances. We assume that the catheters have been subjected to repeated pinching, resulting in material fatigue. In order to eliminate the risk of this serious complication, the subclavian route should be avoided and preference given to the approach via the internal jugular vein.
引用
收藏
页码:872 / 875
页数:4
相关论文
共 12 条
  • [1] AITKEN DR, 1984, AM J SURG, V148, P63
  • [2] AITKENHEAD AR, 1990, GXB ANAESTHESIA, P3
  • [3] BACH F, 1991, CANCER-AM CANCER SOC, V68, P538, DOI 10.1002/1097-0142(19910801)68:3<538::AID-CNCR2820680315>3.0.CO
  • [4] 2-W
  • [5] COMPLICATIONS OF INDWELLING VENOUS CATHETERS - WITH PARTICULAR REFERENCE TO CATHETER EMBOLUS
    DOERING, RB
    STEMMER, EA
    CONNOLLY, JE
    [J]. AMERICAN JOURNAL OF SURGERY, 1967, 114 (02) : 259 - &
  • [6] EVALUATION OF CURRENT TECHNIQUES FOR NON-SURGICAL REMOVAL OF INTRA-VASCULAR IATROGENIC FOREIGN-BODIES
    FISHER, RG
    FERREYRO, R
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1978, 130 (03) : 541 - 548
  • [7] ANATOMICAL AND MORPHOLOGICAL EVALUATION OF PACEMAKER LEAD COMPRESSION
    JACOBS, DM
    FINK, AS
    MILLER, RP
    ANDERSON, WR
    MCVENES, RD
    LESSAR, JF
    COBIAN, KE
    STAFFANSON, DB
    UPTON, JE
    BUBRICK, MP
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (03): : 434 - 444
  • [8] ANATOMICAL MECHANISMS EXPLAINING DAMAGE TO PACEMAKER LEADS, DEFIBRILLATOR LEADS, AND FAILURE OF CENTRAL VENOUS CATHETERS ADJACENT TO THE STERNOCLAVICULAR JOINT
    MAGNEY, JE
    FLYNN, DM
    PARSONS, JA
    STAPLIN, DH
    CHINPURCELL, MV
    MILSTEIN, S
    HUNTER, DW
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (03): : 445 - 457
  • [9] MOGIL RA, 1967, ARCH SURG-CHICAGO, V95, P320
  • [10] Nace C S, 1993, Oncol Nurs Forum, V20, P1227