Rectangular Ion Funnel: A New Ion Funnel Interface for Structures for Loss less Ion Manipulations

被引:22
作者
Chen, Tsung-Chi [1 ]
Webb, Ian K. [1 ]
Prost, Spencer A. [1 ]
Harrer, Marques B. [1 ]
Norheim, Randolph V. [1 ]
Tang, Keqi [1 ]
Ibrahim, Yehia M. [1 ]
Smith, Richard D. [1 ]
机构
[1] Pacific NW Natl Lab, Div Biol Sci, Richland, WA 99352 USA
基金
美国国家卫生研究院;
关键词
IONIZATION-MASS-SPECTROMETRY; ELECTROSPRAY-IONIZATION; MOBILITY SPECTROMETRY; EFFICIENCY; MS; BIOMOLECULES; PRESSURE;
D O I
10.1021/ac503564c
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Structures for lossless ion manipulations (SLIM) have recently demonstrated the ability for near lossless ion focusing, transfer, and trapping in subatmospheric pressure regions. While lossless ion manipulations are advantageously applied to the applications of ion mobility separations and gas phase reactions, ion introduction through ring electrode ion funnels or more conventional ion optics to SLIM can involve discontinuities in electric fields or other perturbations that result in ion losses. In this work, we developed and investigated a new funnel design that aims to seamlessly couple to SLIM at the funnel exit. This rectangular ion funnel (RIF) was initially evaluated by ion simulations, fabricated utilizing printed circuit board technology, and tested experimentally. The RIF was integrated to a SLIM-time of flight (TOF) MS system, and the operating parameters, including RF, DC bias of the RIF electrodes, and electric fields for effectively interfacing with a SLIM, were characterized. The RIF provided a 2-fold sensitivity increase without significant discrimination over a wide m/z range and well matched to that of SLIM, along with greatly improved SLIM operational stability.
引用
收藏
页码:716 / 722
页数:7
相关论文
共 32 条
[1]   A simple electrospray interface based on a DC ion carpet [J].
Anthony, Staci N. ;
Shinholt, Deven L. ;
Jarrold, Martin F. .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2014, 371 :1-7
[2]   Ion mobility spectrometry-mass spectrometry performance using electrodynamic ion funnels and elevated drift gas pressures [J].
Baker, Erin Shammel ;
Clowers, Brian H. ;
Li, Fumin ;
Tang, Keqi ;
Tolmachev, Aleksey V. ;
Prior, David C. ;
Belov, Mikhail E. ;
Smith, Richard D. .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2007, 18 (07) :1176-1187
[3]  
Belov ME, 2000, J AM SOC MASS SPECTR, V11, P19, DOI 10.1016/S1044-0305(99)00121-X
[4]   MASS-SPECTROMETRY WITH ION SOURCES OPERATING AT ATMOSPHERIC-PRESSURE [J].
BRUINS, AP .
MASS SPECTROMETRY REVIEWS, 1991, 10 (01) :53-77
[5]   Enhanced ion utilization efficiency using an electrodynamic ion funnel trap as an injection mechanism for ion mobility spectrometry [J].
Clowers, Brian H. ;
Ibrahim, Yehia M. ;
Prior, David C. ;
Danielson, William F., III ;
Belov, Mikhail E. ;
Smith, Richard D. .
ANALYTICAL CHEMISTRY, 2008, 80 (03) :612-623
[6]   Some thoughts on electrospray ionization mechanisms [J].
Crotti, Sara ;
Seraglia, Roberta ;
Traldi, Pietro .
EUROPEAN JOURNAL OF MASS SPECTROMETRY, 2011, 17 (02) :85-99
[7]   COLLISIONAL FOCUSING EFFECTS IN RADIO-FREQUENCY QUADRUPOLES [J].
DOUGLAS, DJ ;
FRENCH, JB .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1992, 3 (04) :398-408
[8]   ELECTROSPRAY IONIZATION-PRINCIPLES AND PRACTICE [J].
FENN, JB ;
MANN, M ;
MENG, CK ;
WONG, SF ;
WHITEHOUSE, CM .
MASS SPECTROMETRY REVIEWS, 1990, 9 (01) :37-70
[9]   ELECTROSPRAY IONIZATION FOR MASS-SPECTROMETRY OF LARGE BIOMOLECULES [J].
FENN, JB ;
MANN, M ;
MENG, CK ;
WONG, SF ;
WHITEHOUSE, CM .
SCIENCE, 1989, 246 (4926) :64-71
[10]   Simulation of Electric Potentials and Ion Motion in Planar Electrode Structures for Lossless Ion Manipulations (SLIM) [J].
Garimella, Sandilya V. B. ;
Ibrahim, Yehia M. ;
Webb, Ian K. ;
Tolmachev, Aleksey V. ;
Zhang, Xinyu ;
Prost, Spencer A. ;
Anderson, Gordon A. ;
Smith, Richard D. .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2014, 25 (11) :1890-1896