Ultrasensitive CO2 laser photoacoustic system

被引:53
作者
Dumitras, D. C. [1 ]
Banita, S. [1 ]
Bratu, A. M. [1 ]
Cernat, R. [1 ]
Dutu, D. C. A. [1 ]
Matei, C. [1 ]
Patachia, M. [1 ]
Petrus, M. [1 ]
Popa, C. [1 ]
机构
[1] Natl Inst Laser Plasma & Radiat Phys, Dept Lasers, Bucharest 077125, Romania
关键词
Laser photoacoustic spectroscopy; Photoacoustic signal; Noises; Extracavity arrangement; Minimum detectable concentration; Trace gas measurements; WATER-VAPOR; SPECTROSCOPY; SPECTROMETER; AMMONIA; CELL;
D O I
10.1016/j.infrared.2010.05.001
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The present paper describes an extremely sensitive apparatus based upon laser photoacoustic spectroscopy (LPAS) methods which can be used for the detection and measurement of trace gases at very low concentrations (parts per trillion by volume - pptV). Two experimental set-ups were designed and characterized with the photoacoustic (PA) cell in an external configuration: the first one with a low power CO2 laser where the saturation effects are negligible, and a second one with a high power CO2 laser where the saturation effects are important and have to be taken into consideration. In the first case, the minimum detectable concentration was 0.9 ppbV (parts per billion by volume), while in the second case this parameter was improved to 0.29 ppbV. Comparing with the best results published previously in the literature, our minimum detectable concentration is better by a factor of 4.2 in the first case and by a factor of 13.1 in the second case. All measurements were done in nitrogen and ethylene with the 10P(14) line of a continuous wave CO2 laser. This technology can dramatically impact detection in numerous areas. (C) 2010 Elsevier B.V. All rights reserved.
引用
收藏
页码:308 / 314
页数:7
相关论文
共 19 条
[1]  
[Anonymous], [No title captured]
[2]   Geometrical optimization of a longitudinal resonant photoacoustic cell for sensitive and fast trace gas detection [J].
Bijnen, FGC ;
Reuss, J ;
Harren, FJM .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1996, 67 (08) :2914-2923
[3]  
Dumitras DC, 2007, J OPTOELECTRON ADV M, V9, P3655
[4]   An improved CO2 laser intracavity photoacoustic spectrometer for trace gas analysis [J].
Fink, T ;
Buscher, S ;
Gabler, R ;
Yu, Q ;
Dax, A ;
Urban, W .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1996, 67 (11) :4000-4004
[5]  
Harren F.J. M., 2000, Encyclopedia of Analytical Chemistry, V3, P2203
[6]   SENSITIVE INTRACAVITY PHOTOACOUSTIC MEASUREMENTS WITH A CO2 WAVE-GUIDE LASER [J].
HARREN, FJM ;
BIJNEN, FGC ;
REUSS, J ;
VOESENEK, LACJ ;
BLOM, CWPM .
APPLIED PHYSICS B-PHOTOPHYSICS AND LASER CHEMISTRY, 1990, 50 (02) :137-144
[7]   PHOTOACOUSTIC MEASUREMENTS OF AGRICULTURALLY INTERESTING GASES AND DETECTION OF C2H4 BELOW THE PPB LEVEL [J].
HARREN, FJM ;
REUSS, J ;
WOLTERING, EJ ;
BICANIC, DD .
APPLIED SPECTROSCOPY, 1990, 44 (08) :1360-1368
[8]   OPTOACOUSTIC DETECTION USING STARK MODULATION [J].
KAVAYA, MJ ;
MARGOLIS, JS ;
SHUMATE, MS .
APPLIED OPTICS, 1979, 18 (15) :2602-2606
[9]   HARMONIC SATURATED SPECTROSCOPY APPLIED TO MOLECULAR PHOTOACOUSTIC DETECTION [J].
KLIMCAK, CM ;
GELBWACHS, JA .
APPLIED OPTICS, 1985, 24 (02) :247-253
[10]   ATMOSPHERIC-POLLUTION MONITORING USING CO2-LASER PHOTOACOUSTIC-SPECTROSCOPY AND OTHER TECHNIQUES [J].
MEYER, PL ;
SIGRIST, MW .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1990, 61 (07) :1779-1807