Factors influencing daily quality assurance measurements of magnetic resonance imaging scanners

被引:3
|
作者
Owusu, Nana [1 ,2 ]
Magnotta, Vincent A. [1 ,2 ,3 ,4 ]
机构
[1] Univ Iowa, Dept Radiol, Iowa City, IA 52240 USA
[2] Univ Iowa, Dept Biomed Engn, Iowa City, IA 52240 USA
[3] Univ Iowa, Dept Psychiat, Iowa City, IA 52240 USA
[4] 169 Newton Rd, Iowa City, IA 52242 USA
关键词
Magnetic resonance imaging; Signal-to-noise ratio; Quality assurance; MRI;
D O I
10.1007/s12194-021-00638-y
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Magnetic resonance imaging is commonly used in hospitals and clinics to aid medical diagnoses. Scanner performance should be assessed regularly, including daily, weekly, and yearly evaluations to ensure high-quality and artifact-free images. Of these assessments, the daily quality assurance monitors the image quality of the scanner using a manufacturer-provided protocol. In this study, we sought to determine the factors that introduced variability in daily quality assurance data. A phantom was scanned using a head coil in two schemes: with varied phantom placement daily, and with a single phantom placement, and evaluated over approximately 1 month. Minor placement and localization changes accounted for approximately 50% of the variability in the signal-to-noise ratios observed in these measures, driven by changes in the measured signal, while the noise remained constant. The changes in the signal-to-noise ratios were small over the 2-month study period.
引用
收藏
页码:396 / 401
页数:6
相关论文
共 50 条
  • [21] Magnetic resonance imaging-based classification of the myodural bridge complex and its influencing factors
    Feng, Xiao
    Liu, Cong
    Hu, Dong-Mei
    Zhang, Jian-Fei
    Zheng, Nan
    Chi, Yan-Yan
    Yu, Sheng-Bo
    Sui, Hong-Jin
    Xu, Qiang
    SURGICAL AND RADIOLOGIC ANATOMY, 2024, 46 (02) : 125 - 135
  • [22] Magnetic Resonance Imaging of the Brain Using Compressed Sensing - Quality Assessment in Daily Clinical Routine
    Moench, Sebastian
    Sollmann, Nico
    Hock, Andreas
    Zimmer, Claus
    Kirschke, Jan S.
    Hedderich, Dennis M.
    CLINICAL NEURORADIOLOGY, 2020, 30 (02) : 279 - 286
  • [23] Longitudinal assessment of quality assurance measurements in a 1.5 T MR-linac: Part II-Magnetic resonance imaging
    Subashi, Ergys
    Dresner, Alex
    Tyagi, Neelam
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2022, 23 (06):
  • [24] Electrodynamic headphones and woofers for application in magnetic resonance imaging scanners
    Baumgart, F
    Kaulisch, T
    Tempelmann, C
    Gaschler-Markefski, B
    Tegeler, C
    Schindler, F
    Stiller, D
    Scheich, H
    MEDICAL PHYSICS, 1998, 25 (10) : 2068 - 2070
  • [25] Anterior cruciate ligament reconstruction: Magnetic resonance imaging and factors influencing outcome
    Galal, Amr
    Abdul-Maksoud, Safia
    Al-Kandary, Salwa
    Abdul-Salam, Samir
    Awad, Atef
    EGYPTIAN JOURNAL OF RADIOLOGY AND NUCLEAR MEDICINE, 2011, 42 (02): : 193 - 200
  • [26] Factors influencing medial sesamoid arthritis in patients with hallux valgus deformity: Magnetic resonance imaging evaluation
    Choi, Jun Young
    Ahn, Hee Chan
    Baek, Hyeon Seok
    Park, Jong Hun
    Suh, Jin Soo
    FOOT AND ANKLE SURGERY, 2019, 25 (05) : 612 - 617
  • [27] Troubleshooting and Quality Assurance in Hyperpolarized Xenon Magnetic Resonance Imaging Tools for High-Quality Image Acquisition
    Shammi, Ummul Afia
    Delgado, Gabriela Maria Garcia
    Thomen, Robert
    JOVE-JOURNAL OF VISUALIZED EXPERIMENTS, 2024, (203):
  • [28] Cone beam computed tomography for dose calculation quality assurance for magnetic resonance-only radiotherapy
    Wyatt, Jonathan J.
    Pearson, Rachel A.
    Walker, Christopher P.
    Brooks, Rachel L.
    Pilling, Karen
    McCallum, Hazel M.
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2021, 17 : 71 - 76
  • [29] Influence of imaging quality on magnetic resonance-based pressure gradient measurements
    Delles, Michael
    Rengier, Fabian
    Ley, Sebastian
    von Tengg-Kobligk, Hendrik
    Kauczor, Hans-Ulrich
    Unterhinninghofen, Roland
    Dillmann, Ruediger
    MEDICAL IMAGING 2010: BIOMEDICAL APPLICATIONS IN MOLECULAR, STRUCTURAL, AND FUNCTIONAL IMAGING, 2010, 7626
  • [30] Determination of acceptance criteria for geometric accuracy of magnetic resonance imaging scanners used in radiotherapy planning
    Kavaluus, Henna
    Nousiainen, Katri
    Kaijaluoto, Sampsa
    Seppala, Tiina
    Saarilahti, Kauko
    Tenhunen, Mikko
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2021, 17 : 58 - 64