Safety of Multi-Line Transmit Beam Forming for Fast Cardiac Imaging - A Simulation Study

被引:3
作者
Santos, Pedro [1 ,2 ]
Tong, Ling [1 ]
Ortega, Alejandra [1 ]
Lovstakken, Lasse [2 ,3 ]
Samset, Eigil [2 ,4 ,5 ]
D'hooge, Jan [1 ,3 ]
机构
[1] Katholieke Univ Leuven, Dept Cardiovasc Sci, Leuven, Belgium
[2] GE Vingmed Ultrasound AS, Horten, Norway
[3] Norwegian Univ Sci & Technol, MI Lab, N-7034 Trondheim, Norway
[4] Univ Oslo, Oslo, Norway
[5] Ctr Cardiol Innovat, Oslo, Norway
来源
2014 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS) | 2014年
关键词
Multi-Line Transmit; beam forming; acoustic safety; mechanical index; thermal index; acoustic pressure; acoustic intensity;
D O I
10.1109/ULTSYM.2014.0295
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Achieving higher frame rate is a hot topic in cardiac ultrasound. An attractive way to accomplish this is through Multi-Line Transmit (MLT) beam forming. However, transmitting into multiple directions simultaneously may pose additional safety concerns due to acoustic superposition of the MLT beams. In this study, acoustic MLT fields were compared against conventional Single-Line Transmit fields. MLT beam forming showed more than a two-fold increase in peak-compressional acoustic pressure (from 1.8 MPa to 4.6 MPa) and in the mechanical index (MI) (from 1.0 to 2.2), as well as a six-fold increase in averaged intensity. As this might bring safety issues, two modifications of standard MLT beam forming were proposed: i) transmitting the MLT beams with different pulse phases; and ii) reorganising the delay lines of the MLT transmit event. Results showed a decrease in acoustic parameters in the near field using either modification, with the latter leading to peak pressures of 2.3 MPa and an MI of 1.2. In conclusion, straightforward implementation of MLT does bring additional hazard, but small modifications of the transmit beam forming can make it safe for clinical use.
引用
收藏
页码:1199 / 1202
页数:4
相关论文
共 12 条
  • [1] AIUM, 1988, J ULTRAS MED, V7, pS13
  • [2] AIUM/NEMA, 2009, Revision 2
  • [3] [Anonymous], 2010, 62359 IEC
  • [4] Implementation of Parallel Transmit Beamforming Using Orthogonal Frequency Division Multiplexing-Achievable Resolution and Interbeam Interference
    Demi, Libertario
    Viti, Jacopo
    Kusters, Lieneke
    Guidi, Francesco
    Tortoli, Piero
    Mischi, Massimo
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2013, 60 (11) : 2310 - 2320
  • [5] Multi-Line Transmission in 3-D With Reduced Crosstalk Artifacts: A Proof of Concept Study
    Denarie, Bastien
    Bjastad, Tore
    Torp, Hans
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2013, 60 (08) : 1708 - 1718
  • [6] The calculation of the transient near and far field of a baffled piston using low sampling frequencies
    Dhooge, J
    Nuyts, J
    Bijnens, B
    DeMan, P
    Suetens, P
    Thoen, J
    Herregods, MC
    VandeWerf, F
    [J]. JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1997, 102 (01) : 78 - 86
  • [7] FDA, 2008, INF MAN SEEK MARK CL, P62
  • [8] Myocardial rapid velocity distribution
    Kanai, H
    Koiwa, Y
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 2001, 27 (04) : 481 - 498
  • [9] Accelerated Focused Ultrasound Imaging
    Madore, Bruno
    White, P. Jason
    Thomenius, Kai
    Clement, Gregory T.
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2009, 56 (12) : 2612 - 2623
  • [10] MALLART R, 1992, P SOC PHOTO-OPT INS, V1733, P120, DOI 10.1117/12.130591