Evaluation of Manganese(II) and Manganese(VII) Speciation in Water Samples by Ion Pair High-performance Liquid Chromatography-inductively Coupled Plasma Mass Spectrometry

被引:12
作者
Grygo-Szymanko, Emilia [1 ]
Tobiasz, Anna [1 ]
Miliszkiewicz, Natalia [1 ]
Dudek-Adamska, Danuta [1 ]
Walas, Stanislaw [1 ]
机构
[1] Jagiellonian Univ, Dept Analyt Chem, Fac Chem, Ingardena 3, PL-30060 Krakow, Poland
关键词
High-performance liquid chromatography-inductively coupled plasma mass spectrometry (HPLC ICP-MS); ion pair chromatography; manganese speciation; water analysis; SPECTROPHOTOMETRIC DETERMINATION; INJECTION-ANALYSIS; HYDROGEN-PEROXIDE; AQUEOUS-SOLUTIONS; MN(II); TRANSPORT; CHROMIUM; ZINC(II); MN(VII); TIRON;
D O I
10.1080/00032719.2016.1267185
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The speciation of Mn(II) and Mn(VII) is reported by ion pair chromatography. To optimize the separation, sample pH, ion pair reagent, Mn(II) complexing agent, and composition of mobile phase were characterized. The separation of Mn(II) and Mn(VII) was performed using ethylenediamine tetraacetic acid to complex Mn(II), tetrabutylammonium hydroxide as an ion pair reagent, and a C8 column. The separation of the manganese species was demonstrated by high-performance liquid chromatography-inductively coupled plasma mass spectrometry (HPLC ICP-MS). The conversion of Mn(VII) to Mn(II) occurred during the separation and influenced the quantification; hence, the factors affecting this process including the storage time, manganese species ratio, and sample matrix composition were evaluated and suitable calibration was demonstrated. The method was validated by characterization of the selectivity, specificity, linearity, limits of detection and quantification, repeatability, and intermediate precision. The detection limit for Mn(II) was 0.22 mu g center dot L-1, while for Mn(VII), the value was 1.55 mu g center dot L-1.
引用
收藏
页码:2147 / 2160
页数:14
相关论文
共 35 条
[1]   A nano-structured material for reliable speciation of chromium and manganese in drinking waters, surface waters and industrial wastewater effluents [J].
Abdolmohammad-Zadeh, H. ;
Sadeghi, G. H. .
TALANTA, 2012, 94 :201-208
[2]   SPECTROPHOTOMETRIC DETERMINATION OF MANGANESE(II) AND ZINC(II) WITH 4-(2-PYRIDYLAZO)RESORCINOL (PAR) [J].
AHRLAND, S ;
HERMAN, RG .
ANALYTICAL CHEMISTRY, 1975, 47 (14) :2422-2426
[3]   Ion-pairing and reversed phase liquid chromatography for the determination of three different quinolones: Enrofloxacin, lomefloxacin and ofloxacin [J].
Amin, Alaa S. ;
Dessouki, Hassan A. ;
Agwa, Ibrahim A. .
ARABIAN JOURNAL OF CHEMISTRY, 2011, 4 (03) :249-257
[4]  
[Anonymous], 2004, World Health Organization Guidelines for Drinking Water Quality Third Edition, V1
[5]   Manganese: Recent advances in understanding its transport and neurotoxicity [J].
Aschner, Michael ;
Guilarte, Tomas R. ;
Schneider, Jay S. ;
Zheng, Wei .
TOXICOLOGY AND APPLIED PHARMACOLOGY, 2007, 221 (02) :131-147
[6]   Manganese transport in eukaryotes: The role of DMT1 [J].
Au, Catherine ;
Benedetto, Alexandre ;
Aschner, Michael .
NEUROTOXICOLOGY, 2008, 29 (04) :569-576
[7]   Speciation analysis of chromium in drinking water samples by ion-pair reversed-phase HPLC-ICP-MS: validation of the analytical method and evaluation of the uncertainty budget [J].
Baralkiewicz, Danuta ;
Pikosz, Barbara ;
Belter, Magdalena ;
Marcinkowska, Monika .
ACCREDITATION AND QUALITY ASSURANCE, 2013, 18 (05) :391-401
[8]   PROPOSAL FOR A METAL SPECIATION STUDY IN SEDIMENTS [J].
CAMPANELLA, L ;
DORAZIO, D ;
PETRONIO, BM ;
PIETRANTONIO, E .
ANALYTICA CHIMICA ACTA, 1995, 309 (1-3) :387-393
[9]  
Chauhan J. S., 2013, INT J ENG SCI INVENT, V2, P36
[10]   Speciation of metal-EDTA complexes by flow injection analysis with electrospray ionization mass spectrometry and ion chromatography with inductively coupled plasma mass spectrometry [J].
Chen, ZuLiang ;
Sun, Qian ;
Xi, YunFei ;
Owens, Gary .
JOURNAL OF SEPARATION SCIENCE, 2008, 31 (21) :3796-3802