An adaptive finite element simulation of fretting wear damage at the head neck taper junction of total hip replacement: The role of taper angle mismatch
被引:39
|
作者:
论文数: 引用数:
h-index:
机构:
Fallahnezhad, Khosro
[1
]
Oskouei, Reza H.
论文数: 0引用数: 0
h-index: 0
机构:
Flinders Univ S Australia, Med Device Res Inst, Adelaide, SA, AustraliaFlinders Univ S Australia, Med Device Res Inst, Adelaide, SA, Australia
Oskouei, Reza H.
[1
]
Badnava, Hojjat
论文数: 0引用数: 0
h-index: 0
机构:
Behbahan Univ Technol, Discipline Mech Engn, Behbahan, IranFlinders Univ S Australia, Med Device Res Inst, Adelaide, SA, Australia
Badnava, Hojjat
[2
]
论文数: 引用数:
h-index:
机构:
Taylor, Mark
[1
]
机构:
[1] Flinders Univ S Australia, Med Device Res Inst, Adelaide, SA, Australia
Fretting wear;
Material loss;
Taper junction;
Hip implants;
Finite element simulation;
METAL;
CORROSION;
ARTHROPLASTY;
CONTACT;
PSEUDOTUMOR;
PROSTHESIS;
COMPONENTS;
RETRIEVAL;
EVOLUTION;
STRENGTH;
D O I:
10.1016/j.jmbbm.2017.07.003
中图分类号:
R318 [生物医学工程];
学科分类号:
0831 ;
摘要:
An adaptive finite element simulation was developed to predict fretting Wear in a head-neck taper junction of hip joint implant through a two dimensional (2D) model and based on the Archard wear equation. This model represents the most critical section of the head-neck junction which was identified from a 3D model of the junction subjected to one cycle of level gait loading. The 2D model was then used to investigate the effect of angular mismatch between the head and neck components on the material loss and fretting wear process over 4 million gait cycles of walking. Generally, junctions with distal angular mismatches showed a better resistance to fretting wear. The largest area loss in the neck after 4 million cycles of loading was 1.86E-02 mm(2) which was found in the junction with a proximal mismatch angle of 0.124. While, the minimum lost area (4.30E-03 mm(2)) was found in the junction with a distal angular mismatch of 0.024. Contact stress, amplitude of sliding and contact length were found as the key parameters that can influence the amount of material loss and the process of fretting wear damage. These parameters vary over the fretting wear cycles and are highly dependent on the type and magnitude of the taper angle mismatch. This study also showed that lost area does not have a linear relationship with the mismatch angle of taper junctions.
机构:
Flinders Univ S Australia, Coll Sci & Engn, Med Device Res Inst, 1284 South Rd, Clovelly Pk, SA 5042, AustraliaFlinders Univ S Australia, Coll Sci & Engn, Med Device Res Inst, 1284 South Rd, Clovelly Pk, SA 5042, Australia
Feyzi, Mohsen
Hashemi, Reza
论文数: 0引用数: 0
h-index: 0
机构:
Flinders Univ S Australia, Coll Sci & Engn, Med Device Res Inst, 1284 South Rd, Clovelly Pk, SA 5042, AustraliaFlinders Univ S Australia, Coll Sci & Engn, Med Device Res Inst, 1284 South Rd, Clovelly Pk, SA 5042, Australia