An adaptive finite element simulation of fretting wear damage at the head neck taper junction of total hip replacement: The role of taper angle mismatch

被引:39
|
作者
Fallahnezhad, Khosro [1 ]
Oskouei, Reza H. [1 ]
Badnava, Hojjat [2 ]
Taylor, Mark [1 ]
机构
[1] Flinders Univ S Australia, Med Device Res Inst, Adelaide, SA, Australia
[2] Behbahan Univ Technol, Discipline Mech Engn, Behbahan, Iran
关键词
Fretting wear; Material loss; Taper junction; Hip implants; Finite element simulation; METAL; CORROSION; ARTHROPLASTY; CONTACT; PSEUDOTUMOR; PROSTHESIS; COMPONENTS; RETRIEVAL; EVOLUTION; STRENGTH;
D O I
10.1016/j.jmbbm.2017.07.003
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
An adaptive finite element simulation was developed to predict fretting Wear in a head-neck taper junction of hip joint implant through a two dimensional (2D) model and based on the Archard wear equation. This model represents the most critical section of the head-neck junction which was identified from a 3D model of the junction subjected to one cycle of level gait loading. The 2D model was then used to investigate the effect of angular mismatch between the head and neck components on the material loss and fretting wear process over 4 million gait cycles of walking. Generally, junctions with distal angular mismatches showed a better resistance to fretting wear. The largest area loss in the neck after 4 million cycles of loading was 1.86E-02 mm(2) which was found in the junction with a proximal mismatch angle of 0.124. While, the minimum lost area (4.30E-03 mm(2)) was found in the junction with a distal angular mismatch of 0.024. Contact stress, amplitude of sliding and contact length were found as the key parameters that can influence the amount of material loss and the process of fretting wear damage. These parameters vary over the fretting wear cycles and are highly dependent on the type and magnitude of the taper angle mismatch. This study also showed that lost area does not have a linear relationship with the mismatch angle of taper junctions.
引用
收藏
页码:58 / 67
页数:10
相关论文
共 21 条
  • [1] A review on the finite element simulation of fretting wear and corrosion in the taper junction of hip replacement implants
    Feyzi, Mohsen
    Fallahnezhad, Khosro
    Taylor, Mark
    Hashemi, Reza
    COMPUTERS IN BIOLOGY AND MEDICINE, 2021, 130
  • [2] Adaptive finite element simulation of fretting wear and fatigue in a taper junction of modular hip prosthesis
    Messellek, Ali Cherif
    Ouali, Mohand Ould
    Amrouche, Abdelwaheb
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2020, 111
  • [3] Influence of the taper on the fretting wear of the femoral stem-femoral head taper junction in total hip prosthesis
    Capitanu, Lucian
    Badita, Liliana-Laura
    Tiganesteanu, Constantin
    Florescu, Virgil
    INDUSTRIAL LUBRICATION AND TRIBOLOGY, 2022, 74 (04) : 385 - 391
  • [4] A combined experimental and finite element approach to analyse the fretting mechanism of the head-stem taper junction in total hip replacement
    Bitter, Thom
    Khan, Imran
    Marriott, Tim
    Lovelady, Elaine
    Verdonschot, Nico
    Janssen, Dennis
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2017, 231 (09) : 862 - 870
  • [5] A large taper mismatch is one of the key factors behind high wear rates and failure at the taper junction of total hip replacements: A finite element wear analysis
    Ashkanfar, Ariyan
    Langton, David J.
    Joyce, Thomas J.
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2017, 69 : 257 - 266
  • [6] A computational approach to fretting wear prediction at the head-stem taper junction of total hip replacements
    English, Russell
    Ashkanfar, Ariyan
    Rothwell, Glynn
    WEAR, 2015, 338 : 210 - 220
  • [7] The effect of different assembly loads on taper junction fretting wear in total hip replacements
    English, Russell
    Ashkanfar, Ariyan
    Rothwell, Glynn
    TRIBOLOGY INTERNATIONAL, 2016, 95 : 199 - 210
  • [8] Understanding the role of head size and neck length in micromotion generation at the taper junction in total hip arthroplasty
    Bologna, Federico A.
    Putame, Giovanni
    Audenino, Alberto L.
    Terzini, Mara
    SCIENTIFIC REPORTS, 2024, 14 (01)
  • [9] The wear of the taper junction of the stem trunnion with the femoral head of a total modular hip prosthesis
    Capitanu, Lucian
    Badita, Liliana-Laura
    Tiganesteanu, Constantin
    Florescu, Virgil
    JURNAL TRIBOLOGI, 2019, 23 : 61 - 75
  • [10] Finite element analysis of the head-neck taper interface of modular hip prostheses
    Dyrkacz, R. M. R.
    Brandt, J. M.
    Morrison, J. B.
    O' Brien, S. T.
    Ojo, O. A.
    Turgeon, T. R.
    Wyss, U. P.