Design of a superconducting volume coil for magnetic resonance microscopy of the mouse brain

被引:18
作者
Nouls, John C. [1 ]
Izenson, Michael G. [2 ]
Greeley, Harold P. [2 ]
Johnson, G. Allan [1 ]
机构
[1] Duke Univ, Med Ctr, Ctr In Vivo Microscopy, Durham, NC 27710 USA
[2] Creare Res & Dev Inc, Hanover, NH USA
关键词
superconducting coil; microscopy; mouse brain; finite-element radiofrequency model; Helmholtz pair;
D O I
10.1016/j.jmr.2007.12.018
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
We present the design process of a superconducting volume coil for magnetic resonance microscopy of the mouse brain at 9.4 T. The yttrium barium copper oxide coil has been designed through an iterative process of three-dimensional finite-element simulations and validation against room temperature copper coils. Compared to previous designs, the Helmholtz pair provides substantially higher B, homogeneity over an extended volume of interest sufficiently large to image biologically relevant specimens. A custom-built cryogenic cooling system maintains the superconducting probe at 60 +/- 0.1 K. Specimen loading and probe retuning can be carried out interactively with the coil at operating temperature, enabling much higher through-put. The operation of the probe is a routine, consistent procedure. Signal-to-noise ratio in a mouse brain increased by a factor ranging from 1.1 to 2.9 as compared to a room-temperature solenoid coil optimized for mouse brain microscopy. We demonstrate images encoded at 10 x 10 x 20 mu m for an entire mouse brain specimen with signal-to-noise ratio of 18 and a total acquisition time of 16.5 h, revealing neuroanatomy unseen at lower resolution. Phantom measurements show an effective spatial resolution better than 20 mu m. (c) 2008 Elsevier Inc. All rights reserved.
引用
收藏
页码:231 / 238
页数:8
相关论文
共 30 条
  • [1] A HIGH-TEMPERATURE SUPERCONDUCTING RECEIVER FOR NUCLEAR-MAGNETIC-RESONANCE MICROSCOPY
    BLACK, RD
    EARLY, TA
    ROEMER, PB
    MUELLER, OM
    MOGROCAMPERO, A
    TURNER, LG
    JOHNSON, GA
    [J]. SCIENCE, 1993, 259 (5096) : 793 - 795
  • [2] PERFORMANCE OF A HIGH-TEMPERATURE SUPERCONDUCTING RESONATOR FOR HIGH-FIELD IMAGING
    BLACK, RD
    EARLY, TA
    JOHNSON, GA
    [J]. JOURNAL OF MAGNETIC RESONANCE SERIES A, 1995, 113 (01) : 74 - 80
  • [3] DIFFUSION-LIMITED RESOLUTION IN NUCLEAR MAGNETIC-RESONANCE MICROSCOPY
    CALLAGHAN, PT
    ECCLES, CD
    [J]. JOURNAL OF MAGNETIC RESONANCE, 1988, 78 (01): : 1 - 8
  • [4] RECENT PROGRESS IN NMR MICROSCOPY TOWARDS CELLULAR IMAGING
    CHO, ZH
    AHN, CB
    JUH, SC
    JO, JM
    FRIEDENBERG, RM
    FRASER, SE
    JACOBS, RE
    [J]. PHILOSOPHICAL TRANSACTIONS OF THE ROYAL SOCIETY OF LONDON SERIES A-MATHEMATICAL PHYSICAL AND ENGINEERING SCIENCES, 1990, 333 (1632): : 469 - 475
  • [5] NUCLEAR MAGNETIC-RESONANCE MICROSCOPY WITH 4-MU-M RESOLUTION - THEORETICAL-STUDY AND EXPERIMENTAL RESULTS
    CHO, ZH
    AHN, CB
    JUH, SC
    LEE, HK
    JACOBS, RE
    LEE, S
    YI, JH
    JO, JM
    [J]. MEDICAL PHYSICS, 1988, 15 (06) : 815 - 824
  • [6] 3D MR microscopy with resolution 3.7 μm by 3.3 μm by 3.3 μm
    Ciobanu, L
    Seeber, DA
    Pennington, CH
    [J]. JOURNAL OF MAGNETIC RESONANCE, 2002, 158 (1-2) : 178 - 182
  • [7] Perspectives with cryogenic RF probes in biomedical MRI
    Darrasse, L
    Ginefri, JC
    [J]. BIOCHIMIE, 2003, 85 (09) : 915 - 937
  • [8] QUICK MEASUREMENT OF NMR-COIL SENSITIVITY WITH A DUAL-LOOP PROBE
    DARRASSE, L
    KASSAB, G
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1993, 64 (07) : 1841 - 1844
  • [9] Design of superconducting MRI surface coil by using method of moment
    Fang, J
    Chow, MS
    Chan, KC
    Wong, KK
    Shen, GX
    Gao, E
    Yang, ES
    Ma, QY
    [J]. IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2002, 12 (02) : 1823 - 1827
  • [10] A refined circuit model of high temperature superconducting spiral coils for MRI
    Gao, EZ
    Ma, QY
    [J]. IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2001, 11 (01) : 403 - 406