Signal to Noise Ratio of Monolithic Scintillator Detectors for High Resolution PET

被引:0
作者
Maas, Marnix C. [1 ]
van der Laan, D. J. [1 ]
Schaart, Dennis R. [1 ]
Bruyndonckx, Peter [2 ]
Lemaitre, Cedric [2 ]
van Eijk, Carel W. E. [1 ]
机构
[1] Delft Univ Technol, Mekelweg 15, NL-2629 JB Delft, Netherlands
[2] Vrije Univ Brussel VIB, B-1050 Brussels, Belgium
来源
2006 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOL 1-6 | 2006年
关键词
Monolithic scintillator detectors; positron emission tomography; avalanche photodiode; signal to noise ratio;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Monolithic scintillator detectors, consisting of several cm(3) of scintillating material coupled to one or more Hamamatsu S8550 avalanche photodiode (APD) arrays, have been proposed as detectors for high resolution positron emission tomography (PET). These detectors eliminate the dead space caused by reflective material between crystal pixels, and provide good depth of interaction (DOI) information. The entry point of impinging annihilation photons on the front surface of the detector is derived from the scintillation light distribution on the APD array(s). In this work, the various contributions to the energy and timing resolutions were analysed, as well as the influence of several contributions to the spatial resolution. Energy resolutions of around 11.5% FWHM at 511 keV were measured, the major part of which was found to be due to the intrinsic photon variance of the scintillator. A timing resolution of 1.6 ns FWHM was measured against a BaF2-PMT detector. The timing resolution was found to be strongly gain dependent. Spatial resolutions of around 1.75 mm FWHM were measured, fairly independent of the APD gain. This figure still needs to be corrected for the similar to 0.9 mm FWHM measurement beam.
引用
收藏
页码:3027 / 3031
页数:5