Erythrocyte sedimentation rate measurements using MIX-RATE® X20 and VISION A automated analyzers: Method validation and comparison study

被引:4
作者
Prompetchara, Eakachai [1 ]
Nowaratsopon, Sunudda [2 ]
Wongkamchai, Suppakorn [2 ]
Srieakpanit, Jittra [1 ]
Ketloy, Chutitorn [1 ]
机构
[1] Chulalongkorn Univ, Dept Lab Med, Fac Med, Bangkok, Thailand
[2] King Chulalongkorn Mem Hosp, Div Lab Med, Bangkok, Thailand
关键词
automated; erythrocyte sedimentation rate; MIX-RATE (R) X20; VISION A; Westergren method; MATIC CUBE 200; DIAGNOSIS;
D O I
10.1111/ijlh.13914
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: The Westergren method for erythrocyte sedimentation rate (ESR) measurement has a few drawbacks such as being a time-consuming process, poses a risk of biohazard exposure and requires high sample volume. Recent alternative methods and analyzers were developed to overcome those limitations. In this study, we validated two automated ESR analyzers, MIX-RATE (R) X20 and VISION A, and assessed their analytical performance against the Westergren method. Methods: The analyzers were validated for inter-run and intra-run precision. Hemolysis interference and sensitivity to fibrinogen were also analysed. Analytical performance was performed using 177 patient samples spanning low (<40 mm/h), medium (40-80 mm/h), and high (>80 mm/h) ESR ranges. Method agreement and bias against the Westergren method were calculated. Results: The highest intra-run imprecision was seen in the low ESR range for both analyzers. They showed very high agreement with the Westergren method assessed by Spearmen rank correlation coefficient analysis, r = 1.000, p < .0001 for both analyzers. Bland-Altman analysis yielded overall insignificant mean biases for all comparisons. However, systematic positive and negative bias were observed at medium and high ESR levels analysed by MIX-RATE (R) X20 while negative bias was evidenced in the high ESR level measured by VISION A. Conclusions: Overall, results from both automated ESR analyzers showed comparable analytical performance with the Westergren method especially for low ESR levels. However, both positive and negative systematic bias were documented in the high levels. Thus, for clinical use, it must be assessed whether these biases could affect the cut-off for significant clinical values.
引用
收藏
页码:846 / 853
页数:8
相关论文
共 22 条
[1]   Diagnosis and Management of Rheumatoid Arthritis A Review [J].
Aletaha, Daniel ;
Smolen, Josef S. .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2018, 320 (13) :1360-1372
[2]   General Assembly, Diagnosis, Laboratory Test: Proceedings of International Consensus on Orthopedic Infections [J].
Barrack, Robert ;
Bhimani, Samrath ;
Blevins, Jason L. ;
Blevins, Kier ;
Demetres, Michelle ;
Figgie, Mark ;
Fillingham, Yale ;
Goodman, Susan ;
Huddleston, James ;
Kahlenberg, Cynthia ;
Lautenbach, Charles ;
Lin, Jianhao ;
Lonner, Jess ;
Lynch, Maureen ;
Malkani, Arthur ;
Martin, Lily ;
Mirza, Serene ;
Najjad, Muhammad Kazim Rahim ;
Penna, Sreeram ;
Richardson, Shawn ;
Sculco, Peter ;
Shahi, AliSina ;
Szymonifka, Jackie ;
Wang, Qiaojie .
JOURNAL OF ARTHROPLASTY, 2019, 34 (02) :S187-S195
[3]   Comparison of iSED and Ves-Matic Cube 200 Erythrocyte Sedimentation Rate Measurements With Westergren Method [J].
Bogdaycioglu, Nihal ;
Yilmaz, Fatma Meric ;
Sezer, Sevilay ;
Oguz, Esra .
JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2015, 29 (05) :397-404
[4]  
Brigden ML, 1999, AM FAM PHYSICIAN, V60, P1443
[5]   2012 provisional classification criteria for polymyalgia rheumatica: a European League Against Rheumatism/American College of Rheumatology collaborative initiative [J].
Dasgupta, Bhaskar ;
Cimmino, Marco A. ;
Maradit-Kremers, Hilal ;
Schmidt, Wolfgang A. ;
Schirmer, Michael ;
Salvarani, Carlo ;
Bachta, Artur ;
Dejaco, Christian ;
Duftner, Christina ;
Jensen, Hanne Slott ;
Duhaut, Pierre ;
Poor, Gyula ;
Kaposi, Novak Pal ;
Mandl, Peter ;
Balint, Peter V. ;
Schmidt, Zsuzsa ;
Iagnocco, Annamaria ;
Nannini, Carlotta ;
Cantini, Fabrizio ;
Macchioni, Pierluigi ;
Pipitone, Nicolo ;
Del Amo, Montserrat ;
Espigol-Frigole, Georgina ;
Cid, Maria C. ;
Martinez-Taboada, Victor M. ;
Nordborg, Elisabeth ;
Direskeneli, Haner ;
Aydin, Sibel Zehra ;
Ahmed, Khalid ;
Hazleman, Brian ;
Silverman, Barbara ;
Pease, Colin ;
Wakefield, Richard J. ;
Luqmani, Raashid ;
Abril, Andy ;
Michet, Clement J. ;
Marcus, Ralph ;
Gonter, Neil J. ;
Maz, Mehrdad ;
Carter, Rickey E. ;
Crowson, Cynthia S. ;
Matteson, Eric L. .
ANNALS OF THE RHEUMATIC DISEASES, 2012, 71 (04) :484-492
[6]   Hodgkin lymphoma: ESMO Clinical Practice Guidelines for diagnosis, treatment and follow-up [J].
Eichenauer, D. A. ;
Aleman, B. M. P. ;
Andre, M. ;
Federico, M. ;
Hutchings, M. ;
Illidge, T. ;
Engert, A. ;
Ladetto, M. .
ANNALS OF ONCOLOGY, 2018, 29 :19-29
[7]   A short history of the discovery of the erythrocyte sedimentation rate [J].
Grzybowski, A. ;
Sak, J. .
INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2012, 34 (04) :442-444
[8]   The Top 25 Laboratory Tests by Volume and Revenue in Five Different Countries [J].
Horton, Susan ;
Fleming, Kenneth A. ;
Kuti, Modupe ;
Looi, Lai-Meng ;
Pai, Sanjay A. ;
Sayed, Shahin ;
Wilson, Michael L. .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2019, 151 (05) :446-451
[9]   ICSH review of the measurement of the erythocyte sedimentation rate [J].
Jou, J. M. ;
Lewis, S. M. ;
Briggs, C. ;
Lee, S. -H. ;
De la Salle, B. ;
McFadden, S. .
INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2011, 33 (02) :125-132
[10]  
Koepke, 2000, APPROVED STANDARD, V20, P1