Design and Validation of a Semi-Active Variable Stiffness Foot Prosthesis

被引:70
作者
Glanzer, Evan M. [1 ]
Adamczyk, Peter G. [1 ,2 ]
机构
[1] Univ Wisconsin, Dept Mech Engn, Madison, WI 53706 USA
[2] Intelligent Prosthet Syst LLC, Madison, WI 53711 USA
基金
美国国家卫生研究院;
关键词
Prosthetics; lower limb; variable stiffness; TRANSTIBIAL AMPUTEE GAIT; STAIR AMBULATION; HUMAN WALKING; MECHANICS; BACKPACK; FEET;
D O I
10.1109/TNSRE.2018.2877962
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper presents the design and validation of a novel lower limb prosthesis called the variable stiffness foot (VSF), designed to vary its forefoot stiffness in response to user activity. The VSF is designed as a semi-active device that adjusts its stiffness once per stride during swing phases, in order to minimize size, mass, and power consumption. The forefoot keel is designed as an overhung composite beam, whose stiffness is varied by moving a support fulcrum to change the length of the overhang. Stiffness modulation is programmed in response to the gait characteristics detected through foot trajectory reconstruction based on an embedded inertial sensor. The prototype VSF has a mass of only 649 g including the battery, and a build height of 87 mm. Mechanical testing demonstrated a forefoot stiffness range of 10-32 N/mm for the prototype, a threefold range of stiffness variation. The stiffness range can be altered by changing the keel material or geometry. Actuation testing showed that the VSF can make a full-scale stiffness adjustment within three strides, and tracks moderate speed-driven variations within one swing phase. Human subjects testing demonstrated greater energy storage and return with lower stiffness settings. This capability may be useful for the modulating prosthesis energy return to better mimic human ankle function. Subjective feedback indicated clear perception by the subjects of contrasts among the stiffness settings, including interpretation of scenarios for which different settings may be beneficial. Future applications of the VSF include adapting stiffness to optimize stairs, ramps, turns, and standing.
引用
收藏
页码:2351 / 2359
页数:9
相关论文
共 50 条
[1]  
Adamczyk P. G., 2008, THESIS
[2]   Novel Method to Evaluate Angular Stiffness of Prosthetic Feet From Linear Compression Tests [J].
Adamczyk, Peter G. ;
Roland, Michelle ;
Hahn, Michael E. .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2013, 135 (10)
[3]   Sensitivity of biomechanical outcomes to independent variations of hindfoot and forefoot stiffness in foot prostheses [J].
Adamczyk, Peter Gabriel ;
Roland, Michelle ;
Hahn, Michael E. .
HUMAN MOVEMENT SCIENCE, 2017, 54 :154-171
[4]   Comparison of four different categories of prosthetic feet during ramp ambulation in unilateral transtibial amputees [J].
Agrawal, Vibhor ;
Gailey, Robert S. ;
Gaunaurd, Ignacio A. ;
O'Toole, Christopher ;
Finnieston, Adam ;
Tolchin, Ronald .
PROSTHETICS AND ORTHOTICS INTERNATIONAL, 2015, 39 (05) :380-389
[5]   Kinematics and kinetics with an adaptive ankle foot system during stair ambulation of transtibial amputees [J].
Alimusaj, Merkur ;
Fradet, Laetitia ;
Braatz, Frank ;
Gerner, Hans J. ;
Wolf, Sebastian I. .
GAIT & POSTURE, 2009, 30 (03) :356-363
[6]  
[Anonymous], 2017, BIONX EMPOWER FAQ
[7]  
[Anonymous], 2018, REN REN LP
[8]  
[Anonymous], 2017, ABILITY DYNAMICS RUS
[9]  
[Anonymous], 2018, KINN
[10]  
Au SK, 2008, IEEE ROBOT AUTOM MAG, V15, P52, DOI 10.1109/MRA.2008.927697