Influence of rotator cuff integrity on loading and kinematics before and after reverse shoulder arthroplasty

被引:4
作者
Pean, Fabien [1 ]
Favre, Philippe [2 ]
Goksel, Orcun [1 ]
机构
[1] Swiss Fed Inst Technol, Comp Assisted Applicat Med CAiM, CH-8092 Zurich, Switzerland
[2] Zimmer Biomet, Winterthur, Switzerland
关键词
Musculoskeletal simulation; Shoulder; Upper extremity; Rotator cuff tear; Reverse arthroplasty; Joint reaction force; GLENOHUMERAL JOINT; MUSCLE FORCES; UPPER EXTREMITY; MODEL; DESIGN; REVISION; COMPLICATIONS; SUPRASPINATUS; REPLACEMENT; STABILITY;
D O I
10.1016/j.jbiomech.2021.110778
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Reverse Shoulder Arthroplasty has become a very common procedure for shoulder joint replacement, even for scenarios where an anatomical reconstruction would traditionally be used. Our hypothesis is that implanting a reverse prosthesis with a functional rotator cuff may lead to higher joint reaction force (JRF) and have a negative impact on the prosthesis. Available motion capture data during anterior flexion was input to a finite-element musculoskeletal shoulder model, and muscle activations were computed using inverse dynamics. Simulations were carried out for the intact joint as well as for various types of rotator cuff tears: superior (supraspinatus), superior-anterior (supraspinatus and subscapularis), and superior-posterior (supraspinatus, infraspinatus and teres minor). Each rotator cuff tear condition was repeated after shifting the humerus and the glenohumeral joint center of rotation to represent the effect of a reverse prosthesis. Changes in compressive, shear, and total JRF were analyzed. The model compared favorably to in vivo JRF measurements, and existing clinical and biomechanical knowledge. Implanting a reverse prosthesis with a functional rotator cuff or with an isolated supraspinatus tear led to more than 2 times higher compressive JRF than with massive rotator cuff tears (superior anterior or superior-posterior), while the shear force remained comparable. The total JRF increased more than 1.5 times. While a lower shear to compressive ratio may reduce the risk of glenosphere loosening, higher JRF might increase the risk for other failure modes such as fracture or polyethylene wear of the reverse prosthesis.
引用
收藏
页数:7
相关论文
共 59 条
  • [11] Complications and revision of reverse total shoulder arthroplasty
    Boileau, P.
    [J]. ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2016, 102 (01) : S33 - S43
  • [12] Modelling clavicular and scapular kinematics: from measurement to simulation
    Bolsterlee, Bart
    Veeger, H. E. J.
    van der Helm, F. C. T.
    [J]. MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2014, 52 (03) : 283 - 291
  • [13] Adjustments to Zatsiorsky-Seluyanov's segment inertia parameters
    de Leva, P
    [J]. JOURNAL OF BIOMECHANICS, 1996, 29 (09) : 1223 - 1230
  • [14] The Supraspinatus and the Deltoid - Not just two arm elevators
    de Witte, P. B.
    Werner, S.
    ter Braak, L. M.
    Veeger, H. E. J.
    Nelissen, R. G. H. H.
    de Groot, J. H.
    [J]. HUMAN MOVEMENT SCIENCE, 2014, 33 : 273 - 283
  • [15] Dickerson Clark R., 2007, Computer Methods in Biomechanics and Biomedical Engineering, V10, P389, DOI 10.1080/10255840701592727
  • [16] Indications for Reverse Total Shoulder Arthroplasty in Rotator Cuff Disease
    Drake, Gregory N.
    O'Connor, Daniel P.
    Edwards, T. Bradley
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2010, 468 (06) : 1526 - 1533
  • [17] Reverse total shoulder arthroplasty
    Familiari, Filippo
    Rojas, Jorge
    Doral, Mahmut Nedim
    Huri, Gazi
    McFarland, Edward G.
    [J]. EFORT OPEN REVIEWS, 2018, 3 (02): : 58 - 69
  • [18] Reverse Prostheses in Arthropathies With Cuff Tear: Are Survivorship and Function Maintained Over Time?
    Favard, Luc
    Levigne, Christophe
    Nerot, Cecile
    Gerber, Christian
    De Wilde, Lieven
    Mole, Daniel
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2011, 469 (09) : 2469 - 2475
  • [19] An algorithm for estimation of shoulder muscle forces for clinical use
    Favre, P
    Sheikh, R
    Fucentese, SF
    Jacob, HAC
    [J]. CLINICAL BIOMECHANICS, 2005, 20 (08) : 822 - 833
  • [20] Favre P., 1895, PHILOS T SER MATH PH, V367, P2095