The ultrasonic/shear-force microscope: Integrating ultrasonic sensing into a near-field scanning optical microscope

被引:16
作者
La Rosa, A
Cui, X
McCollum, J
Li, N
Nordstrom, R
机构
[1] Portland State Univ, Dept Phys, Portland, OR 97207 USA
[2] Portland State Univ, Mech & Mat Engn Dept, Portland, OR 97207 USA
关键词
D O I
10.1063/1.2052649
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
An ultrasonic transducer is incorporated into a near-field scanning optical microscope (NSOM) to augment its versatility to characterize the properties of layers adsorbed to a sample's surface. Working under typical NSOM operation conditions, the ultrasonic transducer-attached underneath the sample-demonstrates sufficient sensitivity to monitor the waves generated by the tapered NSOM probe that oscillates in the proximity of, and parallel to, the sample's top surface. This capability makes the newly integrated ultrasonic/shear-force microscope a valuable diagnostic tool in the study of sliding friction and surface phenomena in general. Here, it is used to concurrently and independently monitor the effects that probe-sample interactions exert on the probe (that is attached to a piezoelectric tuning fork) and on the sample (that is attached to the ultrasonic transducer). The signal from the tuning fork (TF) constitutes the so called "shear-force" signal, widely used in NSOM as a feedback to control the probe's vertical position but whose working mechanism is not yet well understood. Tests involving repeated vertical z motion of the probe towards and away from the sample's surface reveal that the TF and ultrasonic (US) signals have distinct z dependence. Additionally, where the TF signal showed abrupt changes during the approach, the US changed accordingly. A shift in the probe's resonance frequency that depends on the probe-sample distance is also observed through both the TF and the US responses. Within the sensitivity of the apparatus, ultrasonic signals were detected only at probe-sample distances where the probe's resonance frequency had shifted significantly. These measured signals are consistent with a probe entering and leaving a viscoelastic fluid-like film above the sample. The film acts as the medium where waves are generated and coupled to the ultrasonic sensor located beneath the sample. To our knowledge, this is the first reported use of ultrasonic detection for detailed monitoring of the distance dependence of probe-sample interactions, and provides direct evidence of sound as an energy dissipation channel in wear-free friction. This newly integrated ultrasonic/shear-force microscope, which can be implemented with any functionalized proximal probe (including aperture and apertureless NSOM), can become a valuable metrology tool in surface science and technology. (c) 2005 American Institute of Physics.
引用
收藏
页数:6
相关论文
共 28 条
[1]  
[Anonymous], 1995, INTERMOLECULAR SURFA
[2]   Theoretical and experimental study of the forces between different SNOM probes and chemically treated AFM cantilevers [J].
Bernal, MP ;
Marquis-Weible, F ;
Boillat, PY ;
Lambelet, P .
PROCEEDINGS OF THE IEEE, 2000, 88 (09) :1460-1470
[3]   COMBINED SHEAR FORCE AND NEAR-FIELD SCANNING OPTICAL MICROSCOPY [J].
BETZIG, E ;
FINN, PL ;
WEINER, JS .
APPLIED PHYSICS LETTERS, 1992, 60 (20) :2484-2486
[4]   Influence of environmental conditions on shear-force distance control in near-field optical microscopy [J].
Brunner, R ;
Marti, O ;
Hollricher, O .
JOURNAL OF APPLIED PHYSICS, 1999, 86 (12) :7100-7106
[5]   Distance control of photothermoelastic probe: Application to multi-acquisition and high resolution microscopy [J].
Cavallier, B ;
Cretin, B .
PHOTOACOUSTIC AND PHOTOTHERMAL PHENOMENA: TENTH INTERNATIONAL CONFERENCE, 1999, 463 :99-101
[6]  
Cui X, UNPUB
[7]   Influence of the water layer on the shear force damping in near-field microscopy [J].
Davy, S ;
Spajer, M ;
Courjon, D .
APPLIED PHYSICS LETTERS, 1998, 73 (18) :2594-2596
[8]   Measurements of stiff-material compliance on the nanoscale using ultrasonic force microscopy [J].
Dinelli, F ;
Biswas, SK ;
Briggs, GAD ;
Kolosov, OV .
PHYSICAL REVIEW B, 2000, 61 (20) :13995-14006
[9]   MOTIONS AND RELAXATIONS OF CONFINED LIQUIDS [J].
GRANICK, S .
SCIENCE, 1991, 253 (5026) :1374-1379
[10]  
Gregor MJ, 1996, APPL PHYS LETT, V68, P307, DOI 10.1063/1.116068