Virtual non-contrast in second-generation, dual-energy computed tomography: Reliability of attenuation values

被引:145
作者
Toepker, Michael [1 ]
Moritz, Thomas [1 ]
Krauss, Bernhard [2 ]
Weber, Michael [1 ]
Euller, Gordon [1 ]
Mang, Thomas [1 ]
Wolf, Florian [1 ]
Herold, Christian J. [1 ]
Ringl, Helmut [1 ]
机构
[1] Med Univ Vienna, Vienna Gen Hosp, Dept Radiol, Vienna, Austria
[2] Siemens AG, Healthcare Sect, D-91301 Forchheim, Germany
关键词
Dual-energy; Computed tomography; Virtual non-contrast; Tissue attenuation measurements; Image reconstruction; INITIAL-EXPERIENCE; IMAGE QUALITY; RENAL MASSES; SOURCE CT; ABDOMEN;
D O I
10.1016/j.ejrad.2011.12.011
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To evaluate the reliability of attenuation values in virtual non-contrast images (VNC) reconstructed from contrast-enhanced, dual-energy scans performed on a second-generation dual-energy CT scanner, compared to single-energy, non-contrast images (TNC). Materials and methods: Sixteen phantoms containing a mixture of contrast agent and water at different attenuations (0-1400 HU) were investigated on a Definition Flash-CT scanner using a single-energy scan at 120kV and a DE-CT protocol (100 kV/SN140kV). For clinical assessment, 86 patients who received a dual-phase CT, containing an unenhanced single-energy scan at 120kV and a contrast enhanced (110 ml Iomeron 400 mg/ml; 4 ml/s) DE-CT (100 kV/SN140kV) in an arterial (n = 43) or a venous phase, were retrospectively analyzed. Mean attenuation was measured within regions of interest of the phantoms and in different tissue types of the patients within the corresponding VNC and TNC images. Paired t-tests and Pearson correlation were used for statistical analysis. Results: For all phantoms, mean attenuation in VNC was 5.3 +/- 18.4 HU, with respect to water. In 86 patients overall, 2637 regions were measured in TNC and VNC images, with a mean difference between TNC and VNC of -3.6 +/- 8.3 HU. In 91.5% (n = 2412) of all cases, absolute differences between TNC and VNC were under 15 HU, and, in 75.3% (n = 1986), differences were under 10 HU. Conclusions: Second-generation dual-energy CT based VNC images provide attenuation values close to those of TNC. To avoid possible outliers multiple measurements are recommended especially for measurements in the spleen, the mesenteric fat, and the aorta. (C) 2012 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:E398 / E405
页数:8
相关论文
共 21 条
[1]   Image Fusion in Dual Energy Computed Tomography Effect on Contrast Enhancement, Signal-to-Noise Ratio and Image Quality in Computed Tomography Angiography [J].
Behrendt, Florian F. ;
Schmidt, Bernhard ;
Plumhans, Cedric ;
Keil, Sebastian ;
Woodruff, Seth G. ;
Ackermann, Diana ;
Muehlenbruch, Georg ;
Flohr, Thomas ;
Guenther, Rolf W. ;
Mahnken, Andreas H. .
INVESTIGATIVE RADIOLOGY, 2009, 44 (01) :1-6
[2]   Multi-detector row CT attenuation measurements: Assessment of intra- and interscanner variability with an anthropomorphic body CT phantom [J].
Birnbaum, Bernard A. ;
Hindman, Nicole ;
Lee, Julie ;
Babb, James S. .
RADIOLOGY, 2007, 242 (01) :109-119
[3]   Diffuse Liver Disease: Strategies for Hepatic CT and MR Imaging [J].
Boll, Daniel T. ;
Merkle, Elmar M. .
RADIOGRAPHICS, 2009, 29 (06) :1591-U84
[4]   Problems in the detection and characterization of small renal masses [J].
Bosniak, MA ;
Rofsky, NM .
RADIOLOGY, 1996, 200 (01) :286-287
[5]   Estimating Effective Dose for CT Using Dose-Length Product Compared With Using Organ Doses: Consequences of Adopting International Commission on Radiological Protection Publication 103 or Dual-Energy Scanning [J].
Christner, Jodie A. ;
Kofler, James M. ;
McCollough, Cynthia H. .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2010, 194 (04) :881-889
[6]   Dual-Energy Multidetector CT: How Does It Work, What Can It Tell Us, and When Can We Use It in Abdominopelvic Imaging? [J].
Coursey, Courtney A. ;
Nelson, Rendon C. ;
Boll, Daniel T. ;
Paulson, Erik K. ;
Ho, Lisa M. ;
Neville, Amy M. ;
Marin, Daniele ;
Gupta, Rajan T. ;
Schindera, Sebastian T. .
RADIOGRAPHICS, 2010, 30 (04) :1037-1055
[7]   Dual energy CT (DECT) of the liver: conventional versus virtual unenhanced images [J].
De Cecco, Carlo Nicola ;
Buffa, Vitaliano ;
Fedeli, Stefano ;
Luzietti, Monica ;
Vallone, Andrea ;
Ruopoli, Roberto ;
Miele, Vittorio ;
Rengo, Marco ;
Paolantonio, Pasquale ;
Enrici, Michelangelo Maurizi ;
Laghi, Andrea ;
David, Vincenzo .
EUROPEAN RADIOLOGY, 2010, 20 (12) :2870-2875
[8]   Dual-Energy and Dual-Source CT: Is There a Role in the Abdomen and Pelvis? [J].
Fletcher, Joel G. ;
Takahashi, Naoki ;
Hartman, Robert ;
Guimaraes, Luis ;
Huprich, James E. ;
Hough, David M. ;
Yu, Lifeng ;
McCollough, Cynthia H. .
RADIOLOGIC CLINICS OF NORTH AMERICA, 2009, 47 (01) :41-+
[9]   First performance evaluation of a dual-source CT (DSCT) system [J].
Flohr, TG ;
McCollough, CH ;
Bruder, H ;
Petersilka, M ;
Gruber, K ;
Süss, C ;
Grasruck, M ;
Stierstorfer, K ;
Krauss, B ;
Raupach, R ;
Primak, AN ;
Küttner, A ;
Achenbach, S ;
Becker, C ;
Kopp, A ;
Ohnesorge, BM .
EUROPEAN RADIOLOGY, 2006, 16 (02) :256-268
[10]   Dual-Energy CT in Patients Suspected of Having Renal Masses: Can Virtual Nonenhanced Images Replace True Nonenhanced Images? [J].
Graser, Anno ;
Johnson, Thorsten R. C. ;
Hecht, Elizabeth M. ;
Becker, Christoph R. ;
Leidecker, Christianne ;
Staehler, Michael ;
Stief, Christian G. ;
Hildebrandt, Henriette ;
Godoy, Myrna C. B. ;
Finn, Myra E. ;
Stepansky, Flora ;
Reiser, Maximilian F. ;
Macari, Michael .
RADIOLOGY, 2009, 252 (02) :433-440